Rhizome

Gait, Stair-Climb, Actigraphy and Timed Function Endpoints in FDA Approvals and Labeling

Chetan Mishra
Chetan Mishra
Published Nov 17, 2026 · Updated Nov 23, 2026

Performance-based functional endpoints such as timed walks, stair climbs, rise-from-floor tests and sensor-derived activity measures are increasingly proposed as primary or key secondary endpoints, particularly in neuromuscular, neurological and rare-disease programs. Whether FDA accepts them as evidence of clinically meaningful benefit, and whether the result reaches the label, depends on how sponsors handle validation, clinical meaningfulness and the fit between the measure and the context of use. Teams planning endpoint strategy can make better decisions by knowing where these measures have already succeeded or failed in FDA review.

The analysis below surveys FDA approvals and labeling decisions in which gait, stair-climb, timed function and actigraphy or wearable endpoints played a role, drawing on approved labeling and review documents. It identifies which measures served as primary, secondary or exploratory endpoints across therapeutic areas. It also looks at how reviewers judged each one, including device validation, the target population, the magnitude of change that counts as meaningful, and whether the evidence supported a labeled claim.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.

Gait, stair-climb, actigraphy and timed function endpoints in FDA approvals: what supported labeling and how reviewers judged it

Performance-based functional endpoints have supported FDA approvals and labeled claims in multiple sclerosis, peripheral artery disease, Duchenne muscular dystrophy (DMD), lysosomal storage disorders, hypophosphatasia and pulmonary arterial hypertension. These include timed walks, stair climbs, rise-from-floor tests, the 6-minute walk test (6MWT) and treadmill distance. In the FDA review record, wearable sensor and actigraphy endpoints mostly appear as secondary or exploratory measures. One prominent attempt to use actigraphy as supportive efficacy evidence was rejected because the device was not validated for the target population. The sections below summarize the key precedents and the recurring reviewer concerns.

Key takeaways

  • A timed walk can carry a labeled claim on its own. Ampyra (dalfampridine) is indicated "to improve walking in adult patients with multiple sclerosis (MS)," and the indication states this "was demonstrated by an increase in walking speed" 755. Pletal (cilostazol) is indicated for symptom reduction in intermittent claudication "as indicated by an increased walking distance" 35.
  • Timed function tests anchored some DMD approvals. Duvyzat (givinostat) used 4-stair climb (4SC) time as its primary endpoint 606. Agamree (vamorolone) used time to stand (TTSTAND) velocity 473. For Elevidys, CBER's Center Director relied on time to rise, 10-meter walk/run and stair ascent, together with other evidence, for the ambulatory traditional approval after the primary NSAA endpoint failed. Primary reviewers disagreed with that decision 736739.
  • Endurance walk tests remain standard in rare and cardiopulmonary disease, but reviewers repeatedly questioned clinical meaningfulness, MCIDs set after unblinding, effort dependence and outliers 12312412590.
  • Sensor-derived endpoints have a thin approval record. Actigraphy served as the primary efficacy variable in reviews of OTC analgesic/sleep-aid combinations 275276. For Hetlioz LQ, by contrast, FDA concluded that actigraphy could not support effectiveness decisions in Smith–Magenis syndrome 677. Wearable measures such as PKG scores (Vyalev), stride velocity 95th centile (Elevidys) and step count (Camzyos) appeared as secondary or exploratory endpoints 780428790.

Regulatory framework reviewers apply

Performance outcomes (PerfOs). FDA's final guidance Patient-Focused Drug Development: Selecting, Developing, or Modifying Fit-for-Purpose Clinical Outcome Assessments (October 2025) defines a PerfO as a clinical outcome assessment based on standardized, patient-performed tasks. Performance is quantified using predefined procedures 859. FDA warns that task performance may say little about functioning outside the test setting. Score changes can also be hard to translate into meaningful daily-life change, so supplementary PRO, ObsRO or ClinRO assessments are often needed 863. An earlier PFDD guidance notes that PerfOs depend on patient cooperation and motivation 864.

Digital health technologies (DHTs). The final guidance Digital Health Technologies for Remote Data Acquisition in Clinical Investigations (December 2023) is FDA's main reference 290. It asks sponsors to verify that the DHT measures the physical parameter accurately and precisely. FDA also recommends that sponsors validate that it appropriately assesses the clinical characteristic in the proposed population, whether or not the DHT is a device 308. Endpoints should be well defined and reliable. FDA recommends stakeholder input on clinical meaningfulness and early discussion with the review division 313. Assessments, timing and data aggregation should be defined precisely 314. The guidance also asks for prespecified endpoints and plans to reduce missing data, such as monitoring, reminders and run-in periods 315317.

Disease-specific expectations. The 2018 DMD guidance lists time to climb four stairs, 10-meter walk/run, NSAA and the 6MWT as functional measures for ambulatory patients 870871. It flags several 6MWT problems: maturation effects in younger boys, a floor effect once ambulation is lost, sensitivity to how zero values are handled, and non-normal data 871. Because functional endpoints can be influenced by effort and coaching, it calls blinding critical. It also recommends standardized encouragement and, where practicable, testers who are unaware of the clinical course 872.

Precedent summary

ProductEndpointRoleKey resultLabeling outcome
Ampyra (dalfampridine)Timed 25-foot walk (T25FW) responderPrimaryResponders 34.8% vs 8.3% (MS-F203); 42.9% vs 9.3% (MS-F204) 350351Indication: improve walking, shown by increased walking speed 755
Pletal (cilostazol)Treadmill maximal walking distancePrimaryPooled effect about 63 m 26Indication tied to increased walking distance 35
Duvyzat (givinostat)4-stair climb timePrimaryDifference −1.78 s (95% CI −3.46 to −0.11), p=0.037 609Section 14 primary endpoint; DMD age 6 years and older 606612
Agamree (vamorolone)TTSTAND velocityPrimary+0.0603 rises/s (95% CI 0.0230 to 0.0976), p=0.0018 476Label figure of TTSTAND velocity 621
Elevidys (delandistrogene moxeparvovec)Time to rise, 10MWR, stair ascentKey secondary (time to rise, 10MWR); stair ascent additionalTTR −0.64 s; 10MWR −0.42 s 737Traditional approval, ambulatory patients 4 years and older 732
Skyclarys (omaveloxolone)mFARS (includes upright stability)Primary−2.41 points, p=0.0138 393Section 14 reports the total mFARS result and lists upright stability as one of four domains 401402
Lamzede (velmanase alfa)3-minute stair climb (3MSCT)Primary+2.7 steps/min, p=0.41 57Section 14 reports 3MSCT 6465
Vimizim (elosulfase alfa)6MWTPrimary+22.5 m, p=0.0174 121Placebo-controlled results table 130
Naglazyme (galsulfase)12-minute walkPrimary efficacy assessment+92 m, p=0.025 573FDA proposed walk-distance label text 583
Pombiliti (cipaglucosidase alfa)6MWDPrimary+5.33 m, p=0.608 (primary analysis) 88Label reports 14 m (95% CI −1, 28) 92
Nexviazyme (avalglucosidase alfa)6MWTKey secondary+30.01 m, nominal p=0.0405 508Label shows nominal, unadjusted p-value 513514
Strensiq (asfotase alfa)6MWT (% predicted)Other efficacyJuvenile-onset: 0/8 normal at baseline, 6/6 at Month 48 243"Gait/Mobility" labeling subsection 243
Winrevair (sotatercept)6MWDPrimaryHodges–Lehmann 40.8 m, p<0.001 533534Label figure of 6MWD subgroups 545
Attruby (acoramidis)6MWD4th tier of hierarchical primary; key secondary40 m at Month 30 659Functional capacity reported separately 663664

Timed walks as the basis of an indication

Ampyra and the T25FW responder endpoint

The pivotal MS-F203 and MS-F204 trials used T25FW responder status rather than mean change. Walks were timed by stopwatch twice per visit and the speeds averaged. A responder walked faster at 3 of 4 on-treatment visits than at the fastest of 5 off-treatment visits 347348349. FDA reviewers called the responder definition unconventional and unvalidated. F203's prespecified validation and maintenance tests succeeded, but small mean differences and a nonsignificant global assessment left clinical meaningfulness unresolved 354355. Mean speed gains were 0.21 vs 0.05 ft/s (F203) and 0.22 vs 0.11 ft/s (F204) 352353. Reviewers also noted that the responder criterion ignored the size of improvement 356. They limited generalizability to ambulatory patients who completed 25 feet in 8 to 45 seconds 357.

Even so, the final label states the indication as improved walking "demonstrated by an increase in walking speed" 755. Section 14 describes the T25FW responder method 361756. The label notes an association with MSWS-12 improvement but no established drug–placebo MSWS-12 difference 363. A purely performance-based measure carried the claim, with the patient-reported measure kept in a supporting role.

Pletal and treadmill walking distance

Cilostazol's program used treadmill-measured absolute claudication distance as the main measure 23. It included eight randomized, double-blind, placebo-controlled trials in 2,274 patients 24. Reviewers found 100 mg twice daily superior to placebo in six of eight studies, with a weighted pooled effect of about 63 m 2526. They noted that baseline distance and treadmill incline limited cross-study comparisons 29. LOCF might underestimate delayed benefit, but worst-outcome sensitivity analyses did not change conclusions 3233. Agreement with physical quality-of-life results supported meaningfulness 28.

Neuromuscular disease: timed function tests in DMD and ataxia

Duvyzat: 4-stair climb as primary endpoint

Study 48 enrolled ambulant boys who had two screening 4SC assessments within one second of each other, averaging 8 seconds or less 199. In the primary population (81 givinostat, 39 placebo), 4SC time worsened 1.25 vs 3.03 seconds over 18 months 200. The label presents this as "statistically significant less decline," not as improvement from baseline 607.

The statistical review was critical. Single imputation understated uncertainty and could bias results toward givinostat 202. An observed-data MMRM (p=0.083) and a rank MMRM (p=0.147) were not significant 203. One placebo patient imputed as unable to perform the test strongly influenced the variance 205. In its regulatory action, FDA still judged the −1.78-second difference clinically meaningful because it reflected less deterioration in a strength-dependent functional outcome 702. NSAA, rise from floor and 6MWT favored givinostat but did not reach multiplicity-adjusted significance 703. FDA concluded that Study 48 plus confirmatory evidence established effectiveness 709.

Agamree: time to stand velocity

VBP15-004 used change in TTSTAND velocity (rises/second) at Week 24 as the primary endpoint 473. Velocity improved by +0.0483 rises/s on 6 mg/kg vs −0.0120 on placebo (p=0.0018) 476. Reviewers described the changes as small, a slight improvement against placebo decline 480. They raised a measurement point specific to velocity transforms: near-normal function, reaction-time effects could magnify velocity changes. Sensitivity analyses set a floor of 2.5 seconds on times 483. Missing-data, protocol-deviation and influential-observation analyses supported robustness 481. Key secondary timed tests also favored treatment: 6MWT +41.6 m (p=0.0033) and 10-m run/walk +0.244 m/s (p=0.0018) 627628.

Elevidys: secondary timed tests after a failed primary

In Study 301 Part 1, the primary NSAA endpoint was not significant (difference 0.65; p=0.244) 425426. Time to rise (−0.64 s) and 10-meter walk/run (−0.42 s) favored treatment 427. The stair-ascent difference was −0.36 s (95% CI −0.71, −0.01) 434. The Center Director relied on time to rise, 10MWR and stair ascent, together with other evidence, to grant traditional approval for ambulatory patients aged 4 and older 736732. Primary reviewers disagreed. The statistical reviewer noted that 10MWR effects pointed in opposite directions in Studies 301 and 102. Without prespecification or type I error control, the analyses were exploratory 738. The clinical reviewer recommended a Complete Response 739. Clinical reviewers also stressed effort and encouragement dependence and scoring variability 431. Elevidys shows that timed tests can support approval through a decision-maker's judgment even when the primary review team does not accept the inference.

Exondys 51: when the 6MWT was not accepted

For eteplirsen, FDA reviewers criticized a sponsor analysis that left out two rapidly declining patients who lost ambulation and pooled dose groups without prespecification 451. Reviewers said a historical-control difference of that size would be clinically important if seen in a double-blind, placebo-controlled trial. However, they judged the 6MWT especially bias-prone and unreliable in the open-label study 452. They noted strong effects of effort and motivation near loss of ambulation 448. Other problems included unequal testing opportunities for treated patients and controls 455 and scoring that assigned zero to incomplete tests in historical controls 454.

Skyclarys: a rating scale with a performance component

Omaveloxolone's mFARS is a neurological-exam-based rating scale whose items include upright stability 391392. The primary analysis showed −2.41 points vs placebo (p=0.0138) 393. The upright-stability subscore alone was not significant (p=0.1365) 394. FDA accepted mFARS but noted uncertainty about the meaningfulness of some items 396. Section 14 reports the total mFARS result and lists upright stability as one of four domains 401402. The separate subscore result comes from the FDA review.

Lysosomal and metabolic disease: walk and stair-climb tests

  • Vimizim (MPS IVA). The primary endpoint was 6MWT change at Week 24. Tests were run twice on separate days and averaged, with walking aids held constant 118119. The weekly regimen gained 22.5 m over placebo (p=0.0174); every-other-week dosing gained 0.5 m (p=0.9542) 121122. The medical officer called the improvement modest and its clinical meaningfulness unclear 123. No MCID was prespecified before unblinding 124. Reviewers cited effort dependence, motivation and developmental influences 125. An exploratory subgroup with baseline 6MWT of 200 m or less appeared to drive efficacy but could not support a subgroup claim 127.
  • Naglazyme (MPS VI). The 12-minute walk improved by 92 m (p=0.025) 573. Baseline imbalance raised regression-to-the-mean concerns, which were partly eased by later placebo-to-active crossover gains 575577. Eight subjects had paired tests differing by 200 m or more 578. Including one test taken during illness lowered the estimate to 72 m (p=0.052 to 0.059) 579. A belt-worn ActiTrac activity monitor used as a secondary endpoint was dropped as unreliable. Subjects did not wear it consistently, and shaking could register as activity 789.
  • Lumizyme (late-onset Pompe). In LOTS, the week-78 6MWT difference was about 28 m (p=0.06) 813. Reviewers noted that no guideline defined a clinically meaningful change in Pompe disease and that responder analyses were hard to interpret 816. Model assumption violations led to ANCOVA with re-randomization, which moved p from 0.035 to 0.06 819.
  • Nexviazyme. The 6MWT key secondary showed +30.0 m vs alglucosidase alfa. Formal superiority testing was not possible because the primary endpoint failed superiority 508509. The label carries the result with a nominal, unadjusted p-value 514.
  • Pombiliti. The primary 6MWD analysis was not significant (5.33 m; p=0.608) 88. One comparator patient improved by 355 m, which reviewers considered likely implausible. Including that patient skewed the data and violated MMRM normality, and FDA considered excluding the patient reasonable given questions about effort 90. The label reports a 14 m difference (95% CI −1, 28) 92. It reports 17 m (nominal p=0.047) in ERT-experienced patients using nonparametric ANCOVA 9394.
  • Lamzede (alpha-mannosidosis). The primary 3MSCT endpoint was not significant (+2.7 steps/min; p=0.41) 57. Reviewers placed this against a 7 stairs/min threshold from another disease, since no threshold was established for alpha-mannosidosis 59. They excluded the applicant's responder analyses because the thresholds lacked Agency agreement 61. They still judged the three clinical endpoints collectively meaningful 60. Section 14 reports the 3MSCT results 6465.
  • Strensiq (juvenile-onset HPP). The 6MWT was a supportive, not primary, measure 235. Reviewers judged sustained normal or near-normal walking performance "more than minimally clinically important." Open concerns included measurement error, MCID and differences between European and US reference norms 240. They also cited effort dependence and children's understanding as limitations 241.

Cardiopulmonary indications

Winrevair (PAH). Week-24 6MWD was the STELLAR primary endpoint, with a Hodges–Lehmann difference of 40.8 m (95% CI 27.5 to 54.1) 533534. Missing tests were multiply imputed under a missing-at-random assumption 538. Death and worsening imputations and a control-based pattern-mixture analysis (39.9 m) reproduced the result 539541. Eligibility required two screening 6MWDs of 150 to 500 m 543.

Attruby (ATTR-CM). 6MWD was the fourth tier of a hierarchical win-ratio primary endpoint 653. Reviewers warned that such composites can be driven by small effects on the least important component 661. They concluded that mortality and hospitalization mainly drove the result 662. The final label presents 6MWD separately as functional capacity, with a 40 m difference at Month 30 659663.

Sensor-derived and actigraphy endpoints

Actigraphy as a primary variable: OTC sleep-aid combinations

The Advil PM review includes an "Analysis of Primary Efficacy Variable: Total Sleep Time derived from Actigraphy" 275. The Aleve PM review lists actigraphic wake after sleep onset and sleep latency as primary efficacy variables, and total sleep time and sleep efficiency as secondary 276. These reviews do not include an explicit FDA judgment on actigraphy validity 265268. They are the clearest cases found of an actigraphy-derived variable serving as the main efficacy measure in an approved application.

Hetlioz LQ: a documented rejection

In the tasimelteon review for sleep disturbances in Smith–Magenis syndrome, actigraphy from a Philips watch was a secondary objective intended as supportive evidence 161164. FDA's consultant concluded that the data could not support effectiveness decisions 677. The integrated review found that the algorithm classified true wakefulness poorly 678. The deficiencies follow the DHT guidance framework closely:

  • Validation in the target population. Accuracy and reliability had not been established in SMS. Published case-control comparisons did not validate the measurements 679680.
  • Algorithm transparency. The proprietary algorithm was not submitted. Reported sensitivity of 0.96, specificity of 0.38 and accuracy of 0.77 came from adults with sleep disorders 682683.
  • Wear compliance. Five of 25 ITT patients had no actigraphy data, and there was no wear-compliance requirement for analysis 684685.
  • Meaningful change. There was no meaningful-change threshold and no evidence the device could measure such a change precisely 687.

The consultant left room for future use of actigraphy as a secondary endpoint if these gaps were addressed 688.

Wearables as secondary or exploratory endpoints

  • Vyalev (foscarbidopa/foslevodopa). PKG wearable bradykinesia and dyskinesia scores were secondary endpoints in the gatekeeping hierarchy. Median bradykinesia changed by +1.72 (p=0.0184), and dyskinesia measures were not significant 780782784. FDA noted that none of the other secondary endpoints measured clinically meaningful change in patient function 786.
  • Elevidys. SV95C, measured with ankle-worn devices over 3-week windows, was a Study 301 secondary endpoint 589590. The review documents cited here report no SV95C result or FDA assessment of it 592.
  • Camzyos (mavacamten). Daily step count and other accelerometer parameters were exploratory endpoints 790791. The primary response endpoint used CPET pVO2 combined with NYHA class 879.
  • Adasuve (loxapine). Wrist Actiwatch activity counts were a secondary outcome 274.

Recurring reviewer concerns

  1. Effort, motivation and blinding. Effort dependence appears in the Vimizim, Strensiq, Pombiliti, Exondys 51 and Elevidys reviews 12524190448431. This matches the DMD guidance emphasis on blinding and standardized encouragement 872. Reviewers gave open-label and externally controlled walk data little weight 452.
  2. Clinical meaningfulness without an agreed threshold. Reviewers rejected MCIDs and responder thresholds set after unblinding or without agreement 12461. They noted when no disease-specific threshold existed 81659. Approvals succeeded when reviewers could link the change to everyday function or slower decline 702239.
  3. Missing data and inability to perform. How "unable to perform" is imputed can determine significance, as with Duvyzat 202205. Reviewers look for prespecified missing-data handling and sensitivity analyses, such as the multiple imputation and control-based pattern-mixture analysis in the Winrevair review 539541.
  4. Variability, outliers and distribution. Paired-test discrepancies 578, implausible improvers 90 and non-normal data 482 repeatedly prompted nonparametric or log-transformed analyses 30.
  5. Population and range limits. Walking-speed windows 357, screening 6MWD limits 543 and subgroups driving efficacy 127 shaped how far reviewers would extend conclusions.
  6. DHT-specific validation. For sensor endpoints, reviewers expect population-specific validation, a disclosed algorithm, wear-time rules and a defined meaningful change, as in the Hetlioz LQ review 679682685687. They also expect the device to produce data subjects will actually generate, as the Naglazyme ActiTrac experience shows 789.

Implications for endpoint strategy

The precedents point to a practical hierarchy. Standardized, blinded, in-clinic timed tests (T25FW, 4SC, time to rise, TTSTAND, 6MWT) have the longest record of supporting primary efficacy and labeled claims. That record is strongest when the endpoint is prespecified with a defensible approach to missing values and inability to perform, and is backed by a meaningfulness argument agreed before unblinding 702755. Wearable and actigraphy measures have mostly reached secondary or exploratory status in the review record. Sponsors who want them to carry efficacy should address the verification, validation, usability and missing-data expectations in FDA's 2023 DHT guidance early with the review division 308313317. The Hetlioz LQ review shows in detail which deficiencies lead reviewers to set DHT data aside 677687.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.