Children are often left out of medical device development. Pediatric populations are small and varied, trials are hard to run, and commercial returns are limited, so many devices reach the market with adult-only labeling. For regulatory and clinical teams, knowing how FDA defines the pediatric population, when adult clinical data can support a pediatric indication, and which incentives apply shapes decisions about indication scope, evidence generation and submission pathway.
The analysis below covers FDA's statutory and regulatory definition of pediatric patients, the section 515A disclosure requirement for PMA, HDE and PDP submissions, and the agency's final guidance on extrapolating existing clinical data to pediatric uses. It also reviews the incentives available to pediatric device sponsors, including the HDE profit exemption, and looks at how these policies play out in actual PMA, De Novo, 510(k) and HDE decisions.
FDA pediatric indications for medical devices: extrapolation, incentives and precedents
FDA's pediatric device policy rests on four pieces. The first is a statutory and regulatory definition of the pediatric population. The second is a disclosure requirement (section 515A) for PMA, HDE and PDP submissions. The third is a framework for extrapolating existing adult clinical data to pediatric uses in PMA, HDE and De Novo submissions. The fourth is a small set of incentives, most notably the HDE profit exemption for pediatric humanitarian use devices. This article summarizes each one and then shows how they appear in actual PMA, De Novo, 510(k) and HDE decisions.
Key takeaways
- "Pediatric" runs from birth through age 21. Under 21 CFR 814.3(s), pediatric patients are those aged 21 or younger at diagnosis or treatment, up to but not including the 22nd birthday 288. Because of this definition, many "adult" device indications (18 years and older) also cover ages 18 to 21, and FDA reviews address that group explicitly 144145.
- Extrapolation is a formal, final-guidance framework. FDA's 2016 final guidance Leveraging Existing Clinical Data for Extrapolation to Pediatric Uses of Medical Devices covers PMA, De Novo and HDE submissions, but not 510(k)s 248261262.
- Safety and effectiveness are extrapolated separately. FDA expects safety extrapolation, and full extrapolation in particular, to be less common than effectiveness extrapolation 265266.
- Section 515A asks for information, not new studies. PMAs, PMA supplements, HDEs and PDPs must describe affected pediatric subpopulations and patient numbers from readily available information. That information alone does not support a pediatric indication 808189.
- The main financial incentive is the HDE profit exemption. Humanitarian use devices (HUDs) indicated for pediatric patients may be sold for profit up to an annual distribution number (ADN) 332335336.
How FDA defines the pediatric population
FDA's Premarket Assessment of Pediatric Medical Devices guidance (final, March 24, 2014) defines pediatric patients as individuals from birth through age 21 316317. It gives these approximate subgroups 318319:
| Subgroup | Approximate age range |
|---|---|
| Newborn / neonate | Birth to 1 month |
| Infant | Greater than 1 month to 2 years |
| Child | Greater than 2 to 12 years |
| Adolescent | Greater than 12 through 21 years |
The guidance also flags low-birth-weight newborns (under 2.5 kg), very-low-birth-weight newborns (under 1.5 kg) and preadolescents (typically 11 to 13 years) as groups that clinical studies and labeling may need to address. It cautions that size and developmental stage can matter more than chronological age for a given device 320. The regulation follows the same structure: 21 CFR 814.100(c) names neonates, infants, children and adolescents as the pediatric subpopulations 289.
Premarket expectations for pediatric devices
The 2014 premarket assessment guidance sets out several expectations for sponsors seeking pediatric indications:
- Clinical evidence and subject protection. Clinical evaluation may be needed to support pediatric marketing. Those studies should follow IDE requirements and be designed to protect a vulnerable population, taking account of growth, development and differences from adults 321322.
- Design and bench testing. Sponsors should consider whether an adult device needs redesign for pediatric height, weight, anatomy, physiology, activity level, maturity, immune status and growth. Relevant testing may include biocompatibility, sterility, environmental exposure and device-specific bench or animal work 324.
- Labeling and risk. Labeling should be pediatric-specific where needed. Pediatric risks should be handled through study and device design, not by assuming that adult use translates directly 325322308.
Section 515A pediatric use information
Section 515A applies to new HDEs, PMAs, PMA supplements and PDPs submitted on or after April 10, 2014. These submissions must describe, from readily available information, the pediatric subpopulations affected by the disease or condition and the number of affected pediatric patients 8081. FDA's Providing Information about Pediatric Uses of Medical Devices guidance (final, May 1, 2014) recommends the following 79:
- Put the material in a separate section titled "515A Pediatric Device Use Information" and document a reasonable search effort 84.
- Draw on public-domain biomedical, regulatory and product-research sources. A new clinical study is not required 8283.
- Present a table of indication, pediatric incidence and prevalence, subpopulation and age range, applicable device or component, and data source 8586.
- Update the information with each PMA supplement, or state that nothing new is available and refer back to the last complete submission 8788.
The codified requirements are 21 CFR 814.20(b)(13) for PMAs, 814.104(b)(6) for original HDEs and 814.39(c)(2) for PMA supplements. FDA can also request the information in an amendment or as a condition in an approvable letter 290291292293294295. The guidance states that 515A information on its own does not establish safety and effectiveness for a new pediatric indication 89.
Extrapolation of adult data to pediatric uses
Scope and statutory basis
The 2016 extrapolation guidance traces its authority to the Pediatric Medical Device Safety and Improvement Act (PMDSIA), Title III of the 2007 FDA Amendments Act. PMDSIA permits adult effectiveness data to support pediatric effectiveness when the course of the disease and the device's effects are sufficiently similar. Where appropriate, it also permits extrapolation between pediatric subpopulations 337. The guidance applies to pediatric indications sought through PMA, De Novo or HDE, and not to 510(k) submissions 261262323.
When extrapolation is appropriate
FDA asks four questions 263264:
- Are the disease course and the device's effects sufficiently similar in adults and children?
- Are the adult endpoints relevant to pediatric patients?
- Are the existing data high-quality valid scientific evidence?
- Taken together, do the data support a fair and responsible conclusion of reasonable assurance of safety and effectiveness, or of probable benefit for an HDE?
The decision pathway
- Threshold questions. Does the condition occur in the intended pediatric population, and do the existing data include an endpoint relevant to the pediatric device effect? If the answer to either is no, extrapolating adult data is not appropriate 267.
- Differences assessment. FDA looks for clinically meaningful differences in implantation or contact, device characteristics, pediatric patient characteristics, disease characteristics and other factors 268. The device effect should point in the same direction in adults and children, with a similar but not necessarily identical magnitude of benefit 269.
- Full extrapolation. If no meaningful differences are expected and the adult data are good enough, adult data can replace prospective pediatric data for the endpoint 270271.
- Partial extrapolation. If differences can be accounted for and the adult data are of adequate quality, adult and pediatric data are combined statistically 270272. If the data cannot support a sound conclusion, they should not be extrapolated 273.
Statistical borrowing
For partial extrapolation, FDA describes Bayesian methods that combine adult prior information with pediatric results. How much information is borrowed depends partly on how similar the data turn out to be 274275. Hierarchical models borrow more when study results agree. They do not assume adult and pediatric patients can simply be pooled, and observed differences limit the borrowing 276. Models can adjust for measured age-related covariates such as size or growth 277. For a single adult study, the guidance discusses discounting approaches such as power priors and their limitations, and it notes that non-Bayesian borrowing methods are also acceptable 278279.
Illustrative scenarios from the guidance
- Full extrapolation for effectiveness only. For an adult-approved pleural air-leak sealant, adult effectiveness data could replace a pediatric effectiveness study. FDA still recommended a pediatric safety study because of concerns about adhesions and possible reoperation 356.
- Partial extrapolation of a reference standard. For a diagnostic proposed down to age 2, the adult reference test needed too much blood to use safely in young children. Calibrated adult reference-standard data could be borrowed, with body size modeled 357.
- Partial extrapolation across follow-up periods. Eight-month adult data were modeled with prospectively collected adolescent data to predict a 12-month response, which reduced the adolescent sample size 358.
The guidance also says extrapolation need not follow statutory age-group boundaries. Extrapolating from adults to adolescents may be easier than extrapolating to younger children 359.
510(k) devices: the X-ray imaging example
The extrapolation guidance does not cover 510(k)s. FDA does, however, address pediatric evidence for some cleared device types through device-specific guidance. The Pediatric Information for X-ray Imaging Device Premarket Notifications guidance (final, November 28, 2017) covers devices indicated for pediatric use and general-use X-ray devices expected to see considerable pediatric use 113371. Its recommendations include:
- Pediatric risk-mitigation features, such as size-appropriate automatic exposure control and pediatric protocols 373374375.
- Laboratory image-quality and dose testing across the indicated range of patient sizes, using size-representative phantoms 376377.
- Pediatric-specific labeling, including the caution "Use special care when imaging patients outside the typical adult size range" 379380381.
FDA favors laboratory testing in this guidance. It will request pediatric clinical images only if laboratory testing and extrapolation from adult images are not enough to show substantial equivalence 378.
Pediatric device incentives and support programs
| Mechanism | What it provides |
|---|---|
| HDE profit exemption | The 2007 FDA Amendments Act allowed HUDs indicated for pediatric patients or a pediatric subpopulation to be sold for profit, with restrictions. FDASIA later widened the set of eligible devices 332333. FDA decides eligibility and sets the ADN when it approves the HDE or the applicable supplement 334. |
| Annual distribution number (ADN) | An eligible HUD may be sold for profit each calendar year until sales exceed the ADN 335. The ADN is the number of devices reasonably needed to treat, diagnose or cure 8,000 people in the US per year. For example, a device needed twice per patient per year has an ADN of 16,000 336. |
| Statutory extrapolation authority (PMDSIA) | Lets sponsors use adult effectiveness data in place of some pediatric studies, which lowers the cost of pursuing a pediatric indication 337. |
| Pediatric Device Consortia grants | In September 2018, FDA announced five grants to consortia that advise and support developers of children's medical devices 456. |
| Breakthrough Devices and TAP | Some pediatric devices have received Breakthrough Device designation, including Harmony TPV 473474. FDA described the Autus Size-Adjustable Valve (approved October 2026) as the first pediatric approval from the TAP program 479. |
The 2019 HDE guidance repeats that new HDE applications must include pediatric subpopulation information. FDA treats this as a disclosure requirement, not an incentive 339.
Examples of devices authorized for pediatric populations
PMA approvals with pediatric indications
| Device (PMA, applicant) | Decision | Pediatric indication | Evidence basis |
|---|---|---|---|
| EXCOR Pediatric VAD (P160035, Berlin Heart) 40 | June 6, 2017 41 | Bridge to transplant in pediatric patients, with no numeric age range in the indication 42 | Pediatric IDE trial 44. Originally approved under HDE H100004 (December 16, 2011) 435. |
| Harmony TPV System (P200046, Medtronic) | March 26, 2021 427 | Pediatric and adult patients with severe pulmonary regurgitation 472 | Pediatric data supported the indication, so extrapolation was not needed 427. |
| Dexcom G4 Platinum (Pediatric) CGM (P120005/S002) 25 | February 3, 2014 25 | Ages 2 to 17 48 | Pediatric supplement 48 |
| BreathTek UBT with pediatric urea hydrolysis rate calculation (P100025, Otsuka) 3233 | February 22, 2012 35 | Ages 3 to 17 36 | Pediatric clinical studies 3637 |
| MED-EL COMBI 40+ cochlear implant (P000025/S017) 23 | July 10, 2008 23 | Children 12 months and older 46 | Pediatric clinical trial data 47 |
| OmniaSecure MRI SureScan Lead 3930M (P240036, Medtronic) | April 22, 2025 51 | Adolescents 12 to under 22, subject to weight and anatomy criteria 51 | Partial extrapolation using adult performance, real-world evidence and existing clinical data 51 |
| Impella RP System (P170011, Abiomed) | September 20, 2017 415 | Pediatric patients with body surface area of at least 1.5 m² 415 | Leveraged existing data 415 |
| MI Cancer Seek (P240010, Caris) | November 5, 2024 422 | Ages 1 to 22 discussed 422 | Leveraged clinical data plus pediatric samples 422 |
| LIAISON XL MUREX Anti-HBs (P180039, DiaSorin) 9 | February 21, 2020 9 | Ages 2 to 21 38 | Pediatric samples tested, with no adult-data leveraging 3839 |
Summaries of safety and effectiveness data (SSEDs) for PMAs now usually include a "Pediatric Extrapolation" section. Many state that no clinical data were leveraged, for example Symplicity Spyral, WiSE CRT and SAPIEN M3 410411413. Several companion diagnostics take a narrower route, extending adult data to ages 18 to 21. Examples are OncoMate MSI Dx and Signatera CDx 4950412.
De Novo authorizations
| Device (DEN, company, date) | Pediatric ages | How pediatric use was supported |
|---|---|---|
| Pediatric Vision Scanner (DEN130051, Rebiscan, June 8, 2016) 135 | 2 to 8 150 | Prospective multicenter diagnostic-accuracy trial, with 333 children enrolled 151152 |
| ImPACT / ImPACT Pediatric (DEN150037, August 22, 2016) 128 | 5 to 12 in the pediatric module 146 | Age-stratified normative database built from 915 children, plus a test-retest study 147148 |
| ThermoNeuroModulation Device (DEN170023, Scion Neurostim, March 26, 2018) 123 | 12 to 21 149 | Extrapolation from a trial whose youngest subjects were 18, supported by comparison of adolescent and adult migraine profiles 149 |
| EndeavorRx (DEN200026, Akili, June 15, 2020) 141 | 8 to 12 155470 | Studies conducted in children 156 |
| Control-IQ Technology (DEN190034, Tandem, December 13, 2019) 122 | 14 and older 153 | Limited interim safety data from a study in ages 6 to 12 153 |
| ARC-EX System (DEN240014, Onward Medical, December 19, 2024) 124 | 18 to 21 144 | Extrapolation from data in ages 22 to 74 144 |
FDA press announcements also record De Novo authorizations with pediatric indications 463464465467468469:
- Monarch eTNS for ADHD, ages 7 to 12.
- IB-Stim for IBS-related functional abdominal pain, ages 11 to 18.
- CARPEDIEM for critically ill pediatric patients weighing 2.5 to 10 kg.
510(k) clearances
Cleared pediatric devices often define their population by weight or size rather than age, and they usually rely on a pediatric predicate rather than new pediatric clinical data:
- ANNE Pediatric (K221530, Sibel, November 22, 2022). Indicated for neonates and infants up to 2 years 173205. Clinical testing in 137 neonates included a heart-rate comparison against a reference monitor 206.
- Quantum Blood and Fluid Warming System (K192325, Life Warmer, October 18, 2019). Pediatric patients older than 28 days were added without new performance testing. The submission relied on prior testing, an updated pediatric risk analysis and pediatric-specific labeling 181224226.
- 3880 MRI Patient Monitoring System (K182900, Iradimed, December 14, 2018). Indications use FDA's infant, child and adolescent age bands 184230.
- F&P Optiflow Junior 2/2+ (K222197, April 6, 2023) and Comfort Flo CubCannula (K241778, December 19, 2024). Both are neonatal and pediatric respiratory interfaces cleared against pediatric predicates 163194209.
- Embrace Neonatal MRI System (July 2017). Cleared for neonates weighing 1 to 4.5 kg 459460.
Humanitarian device exemptions
- Melody Transcatheter Pulmonary Valve (H080002, January 25, 2010). Approved for pediatric and adult patients with dysfunctional RVOT conduits 436439.
- Flourish Pediatric Esophageal Atresia Device (2017). Approved under an HDE for infants up to one year old 457458.
Practical considerations for sponsors
- Address ages 18 to 21 explicitly. Because the pediatric definition extends through age 21, an "adult" indication starting at 18 already includes pediatric patients. Recent De Novo and PMA decisions handle this with a short extrapolation rationale stating that patients aged 18 to 21 have no additional differences or risks compared with the adults studied 14415415749.
- Plan extrapolation before the adult pivotal study. Partial extrapolation depends on relevant shared endpoints and on data quality 267270. Adult protocols designed with pediatric borrowing in mind make later hierarchical modeling more feasible 276277.
- Handle safety separately. As the sealant example shows, a full effectiveness extrapolation can still come with a dedicated pediatric safety study 356265.
- Weigh HDE early for rare pediatric conditions. The pediatric profit exemption and ADN can make a HUD commercially viable 332336. Some pediatric devices first approved under HDE, such as EXCOR, later moved to PMA 435.
- For 510(k)s, use device-specific guidance and predicates. The extrapolation framework does not apply, so pediatric support usually comes from bench testing across patient sizes, risk analysis and pediatric labeling, as in the X-ray imaging guidance and K192325 376224226.