Pulse oximeters are known to overestimate oxygen saturation in patients with darker skin, and FDA has moved to address this through its premarket expectations. For device sponsors, regulatory teams and clinical study designers, the agency's position on subject diversity, skin tone measurement and accuracy acceptance criteria directly shapes desaturation study design, sample size and the content of a 510(k) submission.
The analysis below traces FDA's requirements from the 2013 final 510(k) guidance to the January 2025 draft guidance on pulse oximeters for medical purposes. It covers participant numbers, skin tone stratification and objective measurement methods, proposed success criteria, scope across prescription and over-the-counter devices, the draft's current status, and what recent 510(k) summaries for cleared pulse oximeters show about clearance practice.
FDA pulse oximeter accuracy across skin pigmentation: from the 2013 guidance to the 2025 draft
Key points
- Current final guidance (2013): FDA's final 510(k) guidance for pulse oximeters, issued March 2013, recommends a controlled desaturation study with at least 10 healthy subjects and at least 200 paired SpO₂/SaO₂ observations under ISO 80601-2-61:2011 113. For skin pigmentation, it asks for a range of pigmentations that includes at least two darkly pigmented subjects or 15% of the subject pool, whichever is larger 16.
- Proposed approach (January 2025 draft): The draft guidance Pulse Oximeters for Medical Purposes recommends at least 150 participants 78. Participants would be stratified by Monk Skin Tone (MST), with skin tone also measured objectively by Individual Typology Angle (ITA) 3234. Three co-primary success criteria would apply, two of which limit differences in SpO₂ bias across pigmentation 106107108. The draft covers both prescription and over-the-counter (OTC) devices intended for medical purposes 134.
- Status: The January 7, 2025 document is a draft for comment and not for implementation 133. When finalized, it will supersede the 2013 guidance 140. Among the FDA guidance records reviewed, it is still listed as a draft 26. No final pulse oximeter guidance turned up in FDA press announcements after February 2025 347.
- Clearance practice: 510(k) summaries for product code DQA cleared from 2024 through 2026 still largely follow the 2013 design: about 12 subjects, about 3 dark-skinned subjects (usually Fitzpatrick V–VI), and a pooled ARMS result 183184177193. A few also report Monk scores 171178. None of the summaries reviewed reports an ITA measurement.
The 2013 baseline: what FDA has required in 510(k)s
The final guidance Pulse Oximeters - Premarket Notification Submissions [510(k)s] is dated March 3, 2013 125. For clinical accuracy, it points sponsors to Clause 201.12.1.101.2 and Annex EE.2 of ISO 80601-2-61:2011. It recommends at least 10 healthy subjects of varied age and sex and at least 200 paired pulse oximeter/co-oximeter observations 13. Sponsors are asked to document the desaturation profile, the target saturation plateaus, and the criteria used to judge whether reference SaO₂ is stable at the sensor site 14. The ARMS specification should conform to Clause 201.12.1.101.1 of the same standard 15.
On skin pigmentation, the 2013 guidance asks for subjects with a range of pigmentations. This includes at least two darkly pigmented subjects or 15% of the subject pool, whichever is larger 16. FDA later told its advisory panel that this threshold came from U.S. racial census data, not from the population's distribution of skin tones 314.
FDA recognizes ISO 80601-2-61, and pulse oximeter 510(k) submissions generally reference it 344.
Why FDA revisited the framework
November 1, 2022 advisory panel
FDA convened the Anesthesiology and Respiratory Therapy Devices Panel of the Medical Devices Advisory Committee on November 1, 2022 (docket FDA-2022-N-2110) 142. The question was whether pulse oximeters are less accurate in people with darker skin pigmentation 309. FDA's executive summary described studies suggesting missed hypoxemia ("occult hypoxemia"), delayed decisions on treatment eligibility, and worse outcomes in people with darker skin pigmentation 311. The summary made three cautions:
- Race or ethnicity is an inaccurate proxy for skin pigmentation 312.
- Real-world comparisons can be confounded by differences in patients, devices, and the timing of SpO₂ and SaO₂ measurements 313.
- The Fitzpatrick scale was designed to assess susceptibility to sunburn. It does not directly measure skin color or melanin 315.
The panel's 24-hour summary reports that the panel judged the real-world evidence to show reduced accuracy in people with darker skin pigmentation 319. It recommended:
- adequately sized studies across the full pigmentation spectrum 319
- performance measures beyond pooled ARMS 319
- prescription labeling that reports accuracy across pigmentation levels 320
- clear warnings that currently available OTC products are not for medical decision-making 320
FDA's background section in the 2025 draft also notes that its 2021 safety communication listed skin pigmentation among several factors that may affect readings 363.
November 2023 discussion paper and February 2, 2024 panel
In November 2023, FDA published a discussion paper for public comment on improving pulse oximeter performance studies. The paper took into account skin pigmentation and patient-reported race and ethnicity 2023. The panel reconvened on February 2, 2024 (docket FDA-2023-N-4916) 141.
FDA's proposal at that meeting included the following:
- Study size: A desaturation study with at least 24 participants and 480 paired datapoints, replacing the 2013 figures of 10 participants and 200 pairs 267342.
- Scope: The approach would apply to both prescription and OTC devices intended for medical purposes 268269.
- Skin tone assessment: Visual forehead assessment on the 10-point MST scale. At least 25% of participants would come from each of three cohorts (MST 1–4, 5–7 and 8–10). Each MST value would need at least one participant or 15% of its cohort, whichever is greater 270.
- Objective measurement: ITA measured at the emitter site 271.
- Non-disparate performance: The maximum absolute difference in mean SpO₂ bias across ITA and MST levels would have to be below 1.5 percentage points when SaO₂ is above 85%. Between 70% and 85%, it would have to be below 3.5 points 272.
- Overall accuracy: ARMS below 3%, demonstrated with statistical significance 273.
The panel agreed that the 2013 guidance should be revised 274. It supported using MST, with participants spanning at least scores 2 through 9, and objective ITA measurement 274. It was generally in agreement with FDA's definition of non-disparate performance 274. Members questioned whether 24 participants would represent darker skin tones adequately. They also questioned whether measuring near, rather than in, the sensor's light path would capture the relevant pigmentation 275276.
FDA also distinguished two groups of OTC devices at this meeting 345:
- General wellness, sports or aviation oximeters: These generally do not undergo premarket review.
- OTC oximeters intended for medical purposes: These require premarket authorization.
The January 2025 draft guidance
FDA announced the draft on January 6, 2025. It proposes stronger collection and validation of clinical data across skin pigmentations for pulse oximeters intended for medical purposes 1724. The full title is Pulse Oximeters for Medical Purposes - Non-Clinical and Clinical Performance Testing, Labeling, and Premarket Submission Recommendations, dated January 7, 2025 26133.
Scope
The draft covers pulse oximeters intended for medical purposes, whether prescription or OTC, and whether standalone or part of a multi-parameter device 134135. It does not cover the sporting, aviation or wellness oximeter product codes it lists 136137.
Skin tone assessment and enrollment
- Forehead MST: Assess pigmentation visually on the 10-level MST scale. Use colorimetry to measure forehead L* and b* and calculate ITA 32. An appendix covers calibrated printing of MST charts and verifying them with a spectrophotometer 33.
- Sensor site ITA: Measure ITA on the skin surface that directly contacts the sensor emitter 35.
- Cohort allocation: At least 25% of participants in each of MST 1–4, 5–7 and 8–10. At least 50% of the MST 8–10 group should have a forehead ITA of −50° or lower. Each MST group should be at least 40% male and at least 40% female 34.
Sample size
The draft recommends at least 150 participants, sized for at least 80% power on each success criterion 78. Each participant should contribute an average of 20 to 24 paired SpO₂/SaO₂ points, with a minimum of 17 and a maximum of 30 78. At least 30% of pairs should fall in each of the three SaO₂ decades 78.
Success criteria
The draft proposes three co-primary criteria, each to be demonstrated with statistical significance (a one-sided test at 2.5%, or a two-sided 95% confidence interval) 108109:
- Overall accuracy: ARMS below 3% across all participants 103104105.
- Bias across forehead MST groups: In pairwise comparisons of MST 1–4, 5–7 and 8–10, the largest difference in SpO₂ bias must be below 3.5% at SaO₂ 70% to 85%. It must be below 1.5% at SaO₂ above 85% up to 100% 106.
- Bias across sensor site ITA: For a 100-point change in emitter-site ITA, the difference in SpO₂ bias must stay below the same 3.5% and 1.5% limits 107.
The subgroup criteria limit differences in bias. They are not per-subgroup ARMS limits 106.
Labeling
- Prescription devices: Report ARMS overall and by SaO₂ decade (70–80%, 80–90%, 90–100%) 110111. Report bias by forehead MST group and by sensor-site ITA 2.
- OTC devices: Use plain-language directions that address normal ranges, when to seek medical attention, and how to interpret fluctuating readings 115. Summarize the desaturation study, including participant demographics, overall ARMS and accuracy by SaO₂ range 115. FDA also recommends usability testing, such as labeling comprehension testing 127.
- Devices that demonstrate non-disparate performance: Include a prominent statement that the device was evaluated to perform comparably across a wide variety of skin tones, supported by clinical data 128360.
Previously cleared devices and modifications
For pulse oximeters already cleared through a 510(k), the draft recommends that manufacturers gather clinical data to evaluate non-disparate performance, if they have not already done so 358359. These data may come from a controlled desaturation study or from real-world data. Manufacturers would submit the data in a new 510(k), with revised labeling and the 510(k) summary 358359360.
The draft also gives examples of changes likely to require a new 510(k) 357:
- labeling based on non-disparate-performance data
- significant changes to the sensor
- significant changes to the SpO₂ algorithm or SpO₂ software inputs
How the expectations compare
| Element | 2013 final guidance | Feb 2024 panel proposal | Jan 2025 draft guidance |
|---|---|---|---|
| Participants / paired points | ≥10 subjects, ≥200 pairs 13 | ≥24 participants, 480 pairs 267 | ≥150 participants; 20–24 pairs each on average (17–30) 78 |
| Pigmentation enrollment | ≥2 darkly pigmented subjects or 15%, whichever is larger 16 | ≥25% in each of MST 1–4, 5–7, 8–10; each MST value represented 270 | ≥25% per MST group; ≥50% of MST 8–10 with ITA ≤ −50° 34 |
| Skin tone measurement | Not specified beyond "range of skin pigmentations" 16 | Forehead MST plus emitter-site ITA 270271 | Forehead MST with colorimetric ITA, plus ITA at the emitter contact site 3235 |
| Accuracy criterion | ARMS per ISO 80601-2-61 15 | ARMS <3% with statistical significance 273 | ARMS <3%, upper 95% CI bound below 3% 103104 |
| Pigmentation criterion | None stated beyond enrollment 16 | Bias difference <1.5 pts (SaO₂ >85%) and <3.5 pts (70–85%) 272 | Same limits applied across MST groups and per 100-point ITA change 106107 |
| OTC medical-purpose devices | Not addressed in the passages reviewed | Same study design as prescription devices 346 | In scope; added OTC labeling and usability recommendations 134115127 |
What recent 510(k) clearances show
As long as the 2013 guidance remains the final guidance, recent DQA 510(k) summaries mostly reflect its structure, though some sponsors have started to add Monk data.
| Decision date | 510(k) | What the summary reports |
|---|---|---|
| Jan 4, 2024 | K230277 | 4 Fitzpatrick V–VI subjects among 12; overall ARMS 1.98; names the FDA pulse oximeter guidance 193194195. |
| Aug 13 to Dec 1, 2025 | K250931, K251691, K251693, K251695, K251696 (Unimed reusable sensors) | 3 dark subjects (Fitzpatrick V–VI) among 12; each lists the 2013 Pulse Oximeters—Premarket Notification Submissions guidance 183184185186187188. |
| Sept 25, 2025 | K250837 | 6 dark subjects, Fitzpatrick V–VI / Monk 8–9; overall ARMS 2.08 over 70–100%; clinical passage cites ISO 80601-2-61 178179. |
| Dec 19, 2025 | K251751 | 3 dark subjects among 13 (Fitzpatrick II–V), included per FDA's pulse oximeter guidance 180182. |
| Feb 17, 2026 | K252805 | 3 dark subjects among 12; ARMS within ±2% over 70–100% 177. |
| Feb 27, 2026 | K252448 | Light, medium and dark participants; overall ARMS 1.89%; treats ≤3.0% as the FDA-recommended ARMS criterion 173174175176. |
| May 11, 2026 | K252655 | 3 dark subjects, Fitzpatrick V–VI / Monk 8–9; reports a dark-subgroup ARMS of about 2.05–2.10% alongside overall ARMS of 2.10% 171172. |
Older summaries show how pigmentation reporting has developed. In 2016 and 2018, Beijing Choice's MD300-series summaries counted subjects by pigmentation group, for example three subjects with "dark pigmentation" among 12 6769. Other pre-2020 summaries described subjects only as "light to dark," with no count of darkly pigmented participants 747677.
Across the recent summaries, the subgroup sizes are small (three to six darkly pigmented subjects), and pooled ARMS remains the headline metric. That is the pattern the 2022 panel criticized when it called for measures beyond pooled ARMS 319. Results are still reported mostly against the Fitzpatrick scale. Under the 2025 draft, the evidence would instead rest on MST- and ITA-based bias-difference testing in a much larger population 78106107.
Practical implications for sponsors
- New prescription and OTC medical-purpose oximeters: The 2013 guidance remains the final reference 25. The 2025 draft shows where FDA expects clinical evidence to go: about 150 or more participants, MST/ITA stratification and statistical tests of non-disparate bias 7834106107. Sponsors planning new desaturation studies have a reason to design against the draft criteria and to discuss the protocol with FDA.
- Previously cleared devices: The draft anticipates new 510(k)s that carry non-disparate-performance data and the related labeling claims 358360. Changes to the sensor or the SpO₂ algorithm are flagged as likely to require a new 510(k) 357.
- Wellness, sports and aviation oximeters: These products remain outside the draft's scope 136137. FDA distinguishes them from medical-purpose OTC devices by intended use 345.
- Consumer messaging: FDA continues to publish consumer-facing material. In March 2025 it issued a Pulse Oximeter Basics Consumer Update on safe and accurate use, including at home 348.
Open questions for further research include how FDA will resolve public comments on the 150-participant design, how real-world data could support non-disparate-performance claims for legacy devices, and whether any 510(k) has yet been cleared with an MST/ITA-based non-disparate-performance labeling statement.