Rhizome

Potency Assays in CAR-T and Cell Therapy BLAs: CBER Precedent and Review Deficiencies

Chetan Mishra
Chetan Mishra
Oct 13, 2026

Potency is one of the hardest CMC elements to get right in a cell or gene therapy BLA. The mechanism of action is often complex, lot sizes are small, and assays have to be validated well before late-stage clinical data can confirm a link to clinical effect. For regulatory and CMC teams planning a submission, CBER's past decisions on which assays it accepted and why it pushed back are the most practical guide to what a defensible potency strategy looks like.

The analysis below reviews the potency testing approaches in approved CAR-T, TCR-T, gene-modified hematopoietic stem cell, and other cell therapy products, drawing on BLA review memoranda, release specifications, and FDA guidance. It also groups the potency-related deficiencies CBER reviewers raised during these reviews into recurring categories, with links to the source documents.

Rhizome wrote the following article. Let it answer your questions for free by signing up or via Claude or ChatGPT.

Potency assays in CAR-T and cell therapy BLAs: what CBER accepted and where reviewers pushed back

Key takeaways

  • None of the licensed CAR-T products reviewed here relies on CAR expression alone for potency. Release specifications pair CAR expression and viability with an antigen-specific functional readout. Examples are IFN-γ release against CD19-expressing target cells for KYMRIAH 919920, a "CD19-specific" redacted assay for AUCATZYL 773, and flow-cytometric cytotoxicity for the TCR-T product TECELRA 424. FDA's CAR T guidance treats CAR expression as an identity measure and asks for potency testing that measures biological activity 799800.
  • Most products use a potency matrix rather than a single assay. Examples include RYONCIL (two complementary assays plus viability and cell concentration) 660663, LANTIDRA (glucose stimulation index, islet yield, viability) 754756757, CASGEVY (on-target editing frequency plus two redacted assays) 697, and the lentiviral HSC products 605569623.
  • In CBER reviews, potency-related deficiencies fall into seven recurring groups:
    • a single assay judged insufficient to control biological activity (CARVYKTI) 719720
    • a surrogate assay that lacked a demonstrated link to mechanism or clinical outcome (RYONCIL) 664667
    • acceptance criteria not anchored to clinical lots (TECELRA, TREGZI, CASGEVY) 982596699
    • validation and robustness gaps (KRESLADI, LYFGENIA) 626708
    • stability-indicating potency questions (YESCARTA, CARVYKTI) 850723
    • sampling and assay-execution problems (LANTIDRA) 759763
    • release assays not yet linked to an orthogonal functional assay (TREGZI) 593
  • A late fix to the potency assay can push back the action date. CARVYKTI's replacement potency assay arrived late enough to trigger a major amendment, which moved the action due date from November 29, 2021 to February 28, 2022 509994995.

The regulatory framework CBER applies

2011 final guidance, Potency Tests for Cellular and Gene Therapy Products.

  • Each lot must be tested for potency before release, and potency limits should help ensure lots are biologically active and consistently manufactured 832833.
  • The preferred approach is a quantitative bioassay that reflects relevant biological properties and, ideally, the mechanism of action 834835.
  • A matrix of complementary assays is acceptable when a single assay is insufficient. Qualitative assays used for release, stability, or comparability should be accompanied by at least one quantitative assay 837.
  • A non-biological surrogate assay needs data showing that it correlates with biological activity and can distinguish active from inactive or degraded product 838839.
  • The BLA must describe a validated potency assay or matrix. Acceptance criteria should reflect the potency of the lots used in pivotal studies 843.

2023 draft guidance, Potency Assurance for Cellular and Gene Therapy Products. This draft moves from a "release test" model to a potency assurance strategy. That strategy combines material controls, process-parameter controls, in-process testing, and lot-release assays 960.

  • For most CGT products, release testing should include at least one bioassay related to the intended therapeutic effect. The bioassay does not need to reproduce the full mechanism of action 969820970.
  • Acceptance criteria should have a quantitative lower limit that rejects sub-potent lots, and an upper limit where excess activity is a safety risk. "Report only" criteria are not recommended for release assays 818.
  • Lot-release and in-process assays must be validated before a BLA 961830.
  • Release criteria for a licensed product should keep commercial lots consistent with the lots that supplied the primary evidence of effectiveness 972.

2024 final guidance, Considerations for the Development of CAR T Cell Products.

  • CAR expression is listed as an identity measure. Potency testing may use orthogonal methods such as cell killing and cytokine secretion 799800.
  • Bispecific CARs need activity assessed against each target antigen 801.
  • Lot-release and stability assays should be qualified before effectiveness-supporting studies and validated to support a BLA 802803.

2026 draft guidance, Potency Assessment of Active Immunotherapy Products.

  • Quantitative bioassays are the most direct assessment, and complementary assays should be used when one test cannot cover all potency-related CQAs 951953.
  • For live cell-based products, the draft asks for a measure of relevant cellular activity, with cell viability included in the overall assessment 954.

Potency assays in licensed CAR-T and TCR-T products

Product (BLA)Potency-related release testing described in CBER reviewNotable review observations
KYMRIAH (125646)IFN-γ release in response to CD19-expressing target cells; CAR expression by flow cytometry; transduction efficiency by CAR-qPCR 919920IR on flow cytometry compensation, fluorescence-minus-one controls, and whether analyst adjustments changed release outcomes 924925; 2022 CMC commitment to revise the % viable T cell criterion 927
YESCARTA (125643)Cell viability, anti-CD19 CAR expression, and a redacted potency test 848849Large drop in a redacted potency result on stability compared with release; CBER said a drop of that size would exceed normal assay variation 850
TECARTUS (125703)Cell viability, anti-CD19 CAR expression, and a third redacted potency test 933934Methods and validations found adequate; no potency-specific deficiency in the reviewed pages 935
BREYANZI (125714)Potency tested separately on the CD4 and CD8 components; release confirms CAR identity and CD19-specific activation 938939Potency test and criteria redacted 376
ABECMA (125736)Potency test on final formulated product (method redacted) 555556; CD137 activation listed among release attributes 473Exploratory analysis found higher antigen-specific in vitro IFN-γ among attributes possibly associated with higher Cmax and AUC0-28d, with no strong overall relationship 474
CARVYKTI (125746)CAR expression from viable T cells (potency/identity) plus a separate redacted potency assay that replaced the original commercial assay 718509Original approach judged insufficient (see deficiencies below) 719720
AUCATZYL (125813)One potency test, "CD19-specific (b)(4)"; CAR expression is listed separately for identity and quantity 773Potency criterion set by tolerance interval because potency data were limited 774
TECELRA (125789)Flow-cytometric cytotoxicity: MAGE-A4 TCR+ T cells killing T2 targets presenting MAGE-A4 peptide, compared against a reference 402424425Lower limit revised twice in response to IRs 982983
AMTAGVI (125773, TIL)Seven potency-related entries, six redacted, plus dose (total viable cells ≥7.5 × 10⁹) 678Methods found adequate; formal meeting required before adding quality attributes or making major manufacturing changes 680681

Nonclinical functional work in these files generally pairs co-culture cytotoxicity with cytokine release. For CARVYKTI, this meant BCMA-expressing targets at different effector-to-target ratios with IFN-γ and TNF-α readouts 464465. For AUCATZYL, it meant target-specific killing, cytokine secretion, and antigen-specific proliferation 470471. The specific release assay was typically selected from this characterization package, which matches the guidance recommendation 800.

Potency assays in other licensed cell therapies

ProductPotency approachSource
PROVENGE (sipuleucel-T)CD54 upregulation ratio, a marker of APC activation; product at or above the median ratio (23.3) was associated with better survival in a simple analysis (HR 0.79)9991000
RYONCIL (remestemcel-L)Final matrix of two complementary assays (one redacted, plus an optimized IL-2Rα inhibition bioassay), with viability and cell concentration (≥6.68 × 10⁶ cells/mL) under potency660661663
LANTIDRA (donislecel)Glucose stimulation index (insulin ELISA, high/low glucose ratio ≥1), islet yield (≥5,000 EIN/kg initial), viability (≥70%)754755756757
TREGZI (Orca Biosystems)Redacted potency test plus viability, % viable Treg, and viable Treg dose; a traditional Treg suppression assay was judged impractical for release592593594
ZEVASKYN (prademagene zamikeracel)Four redacted potency entries, sampled during manufacturing because of the short shelf life565566
ENCELTO (revakinagene taroretcel)Redacted potency test; rhCNTF expression and viability also listed among release tests612613
Cord blood (DUCORD, Bloodworks, MD Anderson)TNC, viability, viable CD34+ cells (>1.25 × 10⁶/unit); CFU assay for MD Anderson888889891892
CASGEVY (exa-cel)On-target editing frequency plus two redacted potency assays697
ZYNTEGLO / LYFGENIA / SKYSONAEach uses VCN and % LVV+ cells, plus a product-specific readout: transgene protein expression and CFC (ZYNTEGLO), transgene protein expression (LYFGENIA), %ALDP+ cells (SKYSONA)605707686
LENMELDY / KRESLADIViability, VCN, transduction efficiency, ARSA activity (LENMELDY); neutrophil adhesion assay (KRESLADI)569623

Potency-related deficiencies in CBER reviews

1. A single assay that does not control biological activity

CARVYKTI is the clearest CAR-T example.

  • In IR #11 (July 9, 2021), CBER said the then-current calculation of lot-release potency did not sufficiently control biological activity. It also said the proposed commercial assay was unacceptable as a standalone potency assay (IRs #11 and #20) 719720.
  • FDA tentatively considered an alternative assay. Before deciding, it asked for the method description, validation, specifications, and batch results 720993.
  • Janssen replaced the original assay. Validation at the commercial testing site and updated batch analyses arrived in Amendment 56, which was designated a major amendment and extended the review clock 509994995.
  • CBER ultimately considered its potency concerns resolved. The new assay is validated at two sites, and any additional site needs validation data before it can test 509.

For CASGEVY, CBER had asked before Phase 3 for biologically relevant cellular assays in addition to editing-efficiency assays. At BLA submission, Vertex proposed a single potency measure. The reviewer instead accepted a combination of assays for the commercial specification 541542698.

2. Surrogate assays without a demonstrated link to mechanism or outcome

RYONCIL's review history is the most detailed potency case in the CBER record.

  • The 2020 application relied on a non-biological assay of TNFR1 content 657. CBER found the TNFR1-survival analyses confounded by differences in populations and manufacturing processes. Knockdown data showed that inhibition of T cell proliferation was independent of TNFR1 664665.
  • The complete response asked Mesoblast to:
    • identify every assay it considered a potency test
    • justify how those tests ensure each lot's activity
    • show a statistically meaningful relationship to clinical outcomes, surrogate markers, or a relevant in vitro activity 667668
  • Reviewers also questioned assay variability. They noted that a manufacturing-related loss of potency had not been ruled out 669670.
  • The 2024 resubmission resolved these deficiencies with two complementary assays. The reviewer still noted that additional quantitative potency assays would be needed in the future 674.

The absence of clinical correlation is not always disqualifying. For CASGEVY, CBER noted that correlations with clinical outcomes were not performed, yet it accepted the potency specification 698.

3. Acceptance criteria not anchored to clinical lots

  • TECELRA. After IR #7, FDA noted that the revised lower potency limit relied on a lot from a patient who did not respond. After IR #8, the limit was reset to the lowest value associated with a reasonable expectation of benefit, adjusted for inter-assay variability 982983.
  • TREGZI. CBER objected to widening dose, purity, and potency criteria beyond clinically demonstrated ranges and set its own criteria in CMC IR #32. The applicant accepted them 596597.
  • CASGEVY. CMC IRs #8 to #10 tightened on-target editing frequency, viable cell count, and viability. The editing-frequency criterion was justified by the lowest value in a clinical lot with acceptable safety and efficacy 699700.
  • RYONCIL. CBER asked for both a minimum and a maximum TNFR1 limit consistent with pivotal lots, and for a higher minimum on the IL-2Rα assay 672.
  • ZYNTEGLO. VCN, protein expression, and CFC limits were tied to values seen in patients who achieved transfusion independence 606.

4. Validation, robustness, and site-specific limits

  • KRESLADI. Rocket committed to prospectively revalidate the drug product potency assay, with a final report due March 31, 2027. CBER found the release methods adequate only together with the assay commitments 624626.
  • LYFGENIA. A redacted drug product assay had inadequate robustness validation, and robustness studies are due as a PMC 708.
  • RYONCIL. CBER asked for additional robustness data for potency assay validation 671.
  • KYMRIAH. A flow cytometry IR covered compensation and gating controls for CAR+ percentages 924925. Another questioned an MOI calculation that assumed linearity despite validation data showing a nonlinear relationship 926.
  • DUCORD. Issues with the viable CD34+ assay were raised at mid-cycle and resolved through validation 890.

5. Stability-indicating potency

  • YESCARTA. CBER could not tell whether a large potency drop after freeze-thaw was a one-time effect or a trend. Kite committed to additional stability studies 850852.
  • CARVYKTI. Replacing the potency assay left ongoing stability lots with no time-zero data. Janssen agreed to place three more lots on long-term stability 723996.
  • TREGZI. Potency fell below the proposed commercial specification at 72 hours. The reviewer accepted 72-hour storage based on stable viability and clinical experience 598599.
  • CASGEVY. A shipping-validation PMC was required, partly because potency-related attributes had not been evaluated 704.

6. Sampling representativeness and assay execution

  • LANTIDRA. Glucose stimulation index samples came only from the top, pre-culture islet fraction. As a result, CBER could not assess final product potency 759760. FDA said testing in the final container was still important even if results arrived after transplant 762.
  • LANTIDRA, CRL. A complete response deficiency cited inadequate QC operator training on the full GSI procedure 763.
  • ZEVASKYN. Potency samples are taken during manufacturing because of testing limits and the short shelf life 566.

7. Orthogonal functional assays

For TREGZI, CBER asked the applicant to correlate the release assay with a traditional Treg function assay. That function assay was moved into process validation, stability, and comparability work, and it remained unqualified 593.

Implications for sponsors

  • Plan for a functional readout in the release specification. CAR expression, viability, and VCN are accepted as parts of a matrix. Reviews and guidance consistently expect a measure of biological activity tied to mechanism as well 800969719.
  • Set potency limits from clinical lots, especially responders. IRs on TECELRA, TREGZI, and CASGEVY all pushed acceptance criteria back toward clinical experience 982596699.
  • Lock and validate the commercial potency assay before filing. CARVYKTI shows that replacing an assay during review can cost a major amendment and new stability commitments 994723.
  • Build stability data with the commercial potency assay from the start. Assay changes and freeze-thaw effects led to post-approval stability commitments for CARVYKTI and YESCARTA 723852.
  • Expect redactions to hide most specifics. For most licensed CAR-T products, the method, target cell, and numerical acceptance criterion of the functional potency assay are redacted (b)(4) in the public review 848933773555. Precedent analysis therefore depends on review structure, IR history, and PMC language more than on assay parameters.
Rhizome wrote the article above. Let it answer your questions for free by signing up or via Claude or ChatGPT.