Over-the-counter continuous glucose monitors moved glucose sensing out of prescriber-managed diabetes care and into consumer hands. For device sponsors, that shift raises practical questions: which regulatory pathway applies, which predicates are acceptable, and how labeling, intended population, and interoperability claims must change when no prescriber is involved. Clinical and regulatory teams planning a consumer glucose product, or extending a prescription sensor to OTC use, need to understand how FDA has handled these decisions so far.

The analysis below traces the regulatory basis for OTC CGMs back to the integrated CGM classification regulation and its originating De Novo. It reviews the first OTC 510(k) clearances and their predicates, and compares OTC and prescription iCGM product codes by indication, intended user population, and insulin-delivery connectivity. It also examines how the codified special controls apply across both categories.

Rhizome did this research. Let it help you with your most critical decisions by creating a free account.

Over-the-counter CGMs in the US: how FDA cleared them and how they differ from prescription iCGMs

Key points

  • FDA did not create a new device classification for over-the-counter (OTC) continuous glucose monitors. OTC CGMs have been cleared through 510(k) under the existing integrated continuous glucose monitoring system (iCGM) regulation, 21 CFR 862.1355, a Class II device type with special controls 155218180292.
  • That regulation traces to the 2018 De Novo for the Dexcom G6 (DEN170088), which was authorized for prescription use only 74242.
  • The first OTC CGM was the Dexcom Stelo Glucose Biosensor System (K234070), cleared March 5, 2024, with the prescription Dexcom G7 (K231081) as predicate 4156255. Abbott's Lingo (K233655, May 29, 2024) and Libre Rio (K233861, June 7, 2024) followed, both using the prescription FreeStyle Libre 2 (K222447) as predicate 32191181.
  • FDA separates OTC iCGMs from prescription iCGMs mainly through dedicated OTC product codes (SAF and SBH). Both codes are limited to people who are not on insulin and who do not have a significant risk of problematic hypoglycemia, and both exclude connection to insulin delivery controllers 9295. The prescription code QBJ explicitly includes automated insulin dosing systems 93.
  • The codified special controls in 21 CFR 862.1355 apply to OTC and prescription iCGMs alike. The retrieved regulation text has no separate OTC-specific special controls 290. OTC clearances met the same clinical performance special controls 194223348. OTC-specific risks were handled through narrower indications, restricted display ranges, removal of glucose alarms, and lay-user self-selection testing 157220182195277.

Regulatory foundation: one regulation, multiple product codes

21 CFR 862.1355 and the G6 De Novo

FDA classified the iCGM device type through the De Novo for the Dexcom G6 Continuous Glucose Monitoring System (DEN170088), decided March 27, 2018 74. The regulation defines an iCGM as a device that automatically measures glucose in bodily fluids continuously or frequently over a specified period and reliably and securely transmits the data to digitally connected devices, "including automated insulin dosing systems," for use alone or with those devices to manage a condition related to glycemic control 291. It is Class II (special controls) 292.

The G6 De Novo indication was diabetes management in people age 2 years and older. It covered replacing fingerstick testing for treatment decisions and autonomous communication with automated insulin dosing systems, and it carried the condition "This device is for prescription use only" 241242.

OTC product codes under the same regulation

FDA later created OTC product codes that sit under 21 CFR 862.1355 and are classified as Class II 9295:

Product codeFDA classification nameKey definitional limits
QBJIntegrated Continuous Glucose Monitoring System, Factory CalibratedTransmits data to digitally connected devices "including automated insulin dosing systems"; purpose is managing a disease or condition related to glycemic control 93
SAFIntegrated Continuous Glucose Monitor For Non-Intensive Glucose Monitoring, Over-The-CounterFor persons "not on insulin" without "a significant risk of problematic hypoglycemia"; connected devices exclude "devices intended to control the delivery of insulin"; device output "should not be used to make treatment decisions without consulting with a qualified healthcare professional" 92
SBHIntegrated Continuous Glucose Monitor For Non-Intensive Glucose Management, Over-The-CounterSame not-on-insulin and problematic-hypoglycemia limits and the same insulin-delivery exclusion. Detects glucose levels "across the range spanning clinically significant hypoglycemia and hyperglycemia levels." The definition has no HCP-consultation sentence 95

The difference between SAF ("monitoring") and SBH ("management") matters for labeling. SAF's definition carries the statement that output should not drive treatment decisions without an HCP. SBH's definition does not, and it refers to the full clinically significant hypo/hyperglycemic range 9295.

The OTC clearances

Device (510(k))SponsorDecision dateProduct codePredicate
Stelo Glucose Biosensor System (K234070)Dexcom2024-03-05SAFDexcom G7 CGM System (K231081) 4155156
Lingo Glucose System (K233655)Abbott2024-05-29SAFFreeStyle Libre 2 Flash GMS (K222447) 3218219
Libre Rio CGM System (K233861)Abbott Diabetes Care2024-06-07SBHFreeStyle Libre 2 Flash GMS (K222447) 1180181
Signos Glucose Monitoring System (K250106)Signos2025-03-21SAFStelo (K234070) 2107109
Stelo Glucose Biosensor System (K260935)Dexcom2026-06-12SBHStelo (K234070) 5132133

Two features of this pathway stand out:

  1. Prescription predicates for OTC devices. Stelo and Lingo/Libre Rio were found substantially equivalent to prescription iCGMs (G7 and FreeStyle Libre 2), and the change in prescription status was listed as a difference in the predicate comparison tables 157220182. Stelo uses the same hardware design as G7. Its firmware and app were changed to support a longer wear period, restrict connections, and tailor the interface to Stelo users 158.
  2. Software-only OTC entrants. Signos is an OTC mobile application that receives data from an iCGM sensor. It was cleared under SAF with Stelo as predicate, and its indication adds that the information "may be useful in helping users to maintain a healthy weight" 106107109.

FDA's March 5, 2024 announcement described Stelo as the first OTC CGM. It was intended for adults 18 and older who do not use insulin, including people with diabetes on oral medication and people without diabetes who want to understand how diet and exercise affect glucose. FDA cited expanded access without HCP involvement as the rationale 255260261. The announcement stated that Stelo is not for people with problematic hypoglycemia because it does not alert users to that condition, and that users should consult a health care provider before making medical decisions based on its output 260262.

Indications and intended populations: OTC vs prescription

AttributeOTC iCGM (Stelo K234070, Lingo, Libre Rio)Prescription iCGM (Dexcom G7, G6)
Type of useOTC 154217Prescription 169242
PopulationAdults 18+ not on insulin; Libre Rio: "non-insulin using persons age 18 and older" 153216178"Management of diabetes in persons 2 years and older" 166
Diabetes statusNot uniform. FDA's Stelo announcement covers adults with diabetes who do not use insulin and adults without diabetes 260; the Lingo and Libre Rio indications do not specify diabetes status 216178Persons with diabetes 166
Clinical claimDetect euglycemic and dysglycemic levels; help users understand how lifestyle, diet and exercise affect glucose (Stelo, Lingo) 153216. Libre Rio: detect trends and patterns and aid detection of euglycemia, hyperglycemia and hypoglycemia 178Replace fingerstick BG testing for diabetes treatment decisions; aid detection of hyper/hypoglycemia for acute and long-term therapy adjustment 167
Medical actionUser not intended to take medical action without consulting a qualified HCP (Stelo K234070, Lingo) 154216Supports treatment decisions directly 167
ConnectivityNot interoperable with automated insulin delivery (Libre Rio); OTC codes exclude insulin delivery controllers 1829295Autonomous communication with digitally connected devices, including AID systems 168
Glucose alarmsNone (Stelo, Lingo, Libre Rio) 157281221182Present on the predicates; for example, Libre 2 has optional low/high alarms and a mandatory urgent-low alarm 182221
Real-time display rangeStelo 70 to 250 mg/dL; Lingo 55 to 200 mg/dL 157220G7 40 to 400 mg/dL; Libre 2 40 to 400 mg/dL 157220
Sensor wearStelo 15 days vs G7 10 days (each with a 12-hour grace period); Lingo 14 days vs Libre 2 15 days 157221G7: 10 days plus 12-hour grace period; Libre 2: 15 days 157221

Stelo's labeling also states that it should not be used by people with problematic hypoglycemia, because it was not designed for that population. It also should not be used on dialysis, because performance was not evaluated there 279. For Lingo and Libre Rio, the retrieved review text did not state a problematic-hypoglycemia warning. The not-on-insulin limit and the absence of alarms were the clearest stated limitations 286310. The SAF/SBH code definitions, however, both exclude persons with a significant risk of problematic hypoglycemia 9295.

Special controls: identical text, different implementation

What 21 CFR 862.1355 requires

The special controls apply to every device under the regulation, OTC or prescription 292:

  1. Clinical accuracy. Data must come from the intended-use population and cover the full measuring range and wear period, compared against a traceable, FDA-accepted laboratory method 293. Adult performance must meet lower one-sided 95% confidence bounds 294:

    • Below 70 mg/dL: more than 85% within ±15 mg/dL and more than 98% within ±40 mg/dL 295296
    • 70 to 180 mg/dL: more than 70% within ±15% and more than 99% within ±40% 297298
    • Above 180 mg/dL: more than 80% within ±15% and more than 99% within ±40% 299300
    • Across the full measuring range: more than 87% within ±20% 301

    Readings must never cross between below 70 and above 180 mg/dL 302. Rate-of-change limits apply 303304. Pediatric performance must be similar to adult performance unless FDA accepts a justification 305.

  2. Secure, reliable data transmission at clinically meaningful intervals, with no clinically significant data gaps that would prevent connected devices from achieving their intended use 270269.

  3. Manufacturing and release controls that ensure released devices meet the performance specifications 306.

  4. Interference testing 307.

  5. Wear-limit enforcement that prevents sensors from being used beyond their claimed wear period 308.

  6. Usability study showing intended users can use the device safely and obtain the expected accuracy 309.

  7. Labeling of clinical performance for each intended-use population and insertion site. This covers accuracy by glucose band (<54, 54 to <70, 70 to 180, >180 to 250, >250 mg/dL), rate-of-change accuracy, data gaps, alert performance (if applicable), and observed sensor life 268.

The retrieved text of 21 CFR 862.1355 does not contain separate special controls for OTC devices or non-insulin users, and it does not itself prescribe a prescription-use statement 290.

How OTC sponsors met the same controls

  • Clinical performance. Stelo's decision summary states that the device met the iCGM clinical-performance special controls 194. Against YSI venous plasma over about 15 days of wear, it reported MARD of 7.3% to 9.1% and 92.1% to 96.0% of readings within 20 mg/dL or 20% 190191. Readings outside the 70 to 250 mg/dL display window were evaluated separately, and the full 40 to 400 mg/dL sensor range was available retrospectively 192273. Lingo used the predicate's clinical studies with a new analysis for the changed age range, measuring range and wear period, and met the special-control requirements 222223224. Libre Rio referenced the 862.1355 special controls and relied on analytical performance established in K222447 183184.
  • Usability (special control 6) adapted to OTC. Stelo's human factors program tested whether lay users could correctly identify themselves as appropriate users from pre-purchase labeling and then use the system safely 195196. Sixty participants evaluated four e-commerce self-selection platforms, and the 15 who self-identified correctly went on to simulated-use testing 277. Lingo's app includes onboarding with a self-selection questionnaire that users must consent to before use 287. The Libre Rio review text cited human factors support for substantial equivalence but did not describe OTC-specific self-selection methods 312.
  • Alerts and labeling (special control 7). The labeling control requires alert performance data only "if applicable" 268. All three initial OTC devices removed glucose alerts and alarms. For Stelo, this included urgent-low, low, high and rapid-change alerts 281221182.
  • Interoperability (special control 2). Prescription iCGMs are cleared to feed AID systems 168. The OTC product codes exclude insulin delivery controllers 9295. Libre Rio's comparison table lists it as not interoperable with AID 182, and Stelo's firmware restricts connections 158.

How the OTC category has evolved

The OTC category has moved toward claims closer to prescription iCGMs while keeping the insulin exclusion:

  • SBH "management" code. Libre Rio was cleared under SBH in June 2024 1180. In June 2026, Stelo K260935 moved from SAF to SBH 5132133.
  • Pediatric expansion. Stelo K260935 lowered the age threshold from 18 to 2 years and older, still limited to people not on insulin 130134. Pediatric support came from reanalyzing a prior G7 pediatric study (164 children) with the Stelo algorithm, not from a new Stelo pediatric study. Accuracy and sensor survival beyond about day 10.5 were not directly studied in children 352353354. FDA concluded the device met the iCGM clinical-performance special controls, with that limitation noted 348.
  • Broader display range and removal of the HCP statement. The display range widened from 70 to 250 mg/dL to 55 to 350 mg/dL, and a Direct to Watch feature was added 135136. The statement that users should not take medical action without consulting an HCP was removed from the indication 134. This tracks the SBH code definition, which omits that sentence 95.

Related category: glucose range monitoring (21 CFR 862.1359)

A separate, non-iCGM classification was created by the De Novo for the Biolinq Shine Autonomous Time-in-Range Microsensor (DEN240080), decided September 23, 2025, under product code SFU 263. The 862.1359 "glucose range monitoring system" provides qualitative or semi-quantitative glucose information, not quantitative concentrations, and is not intended for acute decisions such as insulin dosing 91351. Biolinq Shine is indicated for people aged 22 or older who are not on insulin. Its users are not intended to take medical action without HCP consultation, and FDA designated it a prescription device 349350. Its special controls mirror the iCGM structure: clinical validation, data availability, interference, wear limits, usability and population-specific labeling 315. Not-on-insulin populations can therefore fall under either regulation, and OTC status depends on the product code and clearance, not only on the user population.

Practical implications for regulatory strategy

  • Pathway. An OTC iCGM is a 510(k) under 21 CFR 862.1355, with SAF or SBH as product code. Prescription iCGMs (QBJ) have served as predicates. The predicate comparison must address the shift to OTC, the restricted population, and the loss of AID interoperability and alarms 157220182.
  • Evidence. The quantitative accuracy special controls do not relax for OTC. Sponsors have reused the predicate's clinical datasets with new analyses for the modified algorithm, range and wear period 222352. Expect FDA to review out-of-display-range performance separately 192356.
  • OTC-specific human factors. Self-selection from pre-purchase labeling (including e-commerce platforms) and safe lay use are evaluated under the usability control 195277.
  • Labeling boundaries. The not-on-insulin limit, exclusion of problematic hypoglycemia, and any HCP-consultation language should match the target product code's definition 9295279.