Rhizome

How FDA Uses MDSAP Audit Reports in Place of Routine Inspections

Chetan Mishra
Chetan Mishra
Published Nov 17, 2026 · Updated Nov 23, 2026

For device manufacturers weighing whether to join the Medical Device Single Audit Program, the practical question is how much an MDSAP audit actually changes their exposure to FDA. One audit can satisfy several regulators, but FDA defines exactly which of its inspections that audit replaces and which it does not. Quality and regulatory teams need that boundary to plan inspection readiness, budget audit cycles and judge how audit findings may reach FDA compliance staff.

The analysis below draws on FDA's device inspection compliance program, MDSAP program materials and FDA warning letters. It sets out which inspection types MDSAP audits replace and which still apply to participants, and how FDA reviews and classifies MDSAP reports that cover the United States. It then looks at how those results feed into follow-up inspections and enforcement, and what the warning-letter record shows about MDSAP.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.

How FDA uses MDSAP audit reports in place of routine inspections, and what the record shows about outcomes

FDA accepts Medical Device Single Audit Program (MDSAP) audit reports only in place of routine surveillance inspections of device manufacturers that volunteer for the program. Every other FDA inspection type still applies to MDSAP participants. FDA reviews and classifies each report that covers the United States, and treats the results as quality data in compliance decisions. Those results can lead to for-cause or compliance follow-up inspections. FDA warning letters, however, rarely mention MDSAP. In the letters reviewed for this article, none was based on an MDSAP audit report.

Key takeaways

  • FDA's device inspection compliance program, CP 7382.850, says FDA "utilizes MDSAP audit reports as a substitute for agency surveillance inspections" for manufacturers that volunteer to participate 243. It also says that "surveillance inspections are not conducted at manufacturer sites actively enrolled in MDSAP" 253.
  • "All manufacturers continue to be subject to FDA non-surveillance inspections" 136. MDSAP participants can still receive for-cause 248, compliance follow-up 251, Specific Product Risk Assignment (SPRA) 252, PMA preapproval or postmarket 245250 and radiation-control (EPRC) inspections 136.
  • FDA reviews and classifies every MDSAP report that names the United States as a jurisdiction 136. A final MDSAP classification of NAI or VAI is treated in the same way as an FDA inspection classification for planning purposes 244. An MDSAP audit "that resulted in regulatory action" is a basis for a compliance follow-up inspection 245.
  • The 2024 QMSR final rule separates MDSAP audit reports, which FDA "may accept instead of routine surveillance inspections" 277, from ISO 13485 certificates. FDA will not accept a certificate as a substitute for an FDA inspection 280, and certificate holders "are not exempt from FDA inspections" 284.

From pilot to operational program

FDA's November 2013 Federal Register notice announced its participation in the MDSAP International Coalition Pilot. It stated that FDA "will be participating in the MDSAP and will accept the resulting audit reports as a substitute for routine Agency inspections" 168. The pilot had stated limits. FDA's "For Cause" and "Compliance Followup" inspections were not affected. The pilot also did not apply to necessary PMA preapproval or postapproval inspections, or to decisions under section 513(f)(5) 169.

A December 2015 notice ended the earlier ISO 13485:2003 Voluntary Audit Report Submission Pilot and announced the MDSAP operational phase. Under that phase, recognized third parties could submit MDSAP audit reports that FDA would use for routine inspections 170. FDA said MDSAP gave it better assurance than the earlier pilot because the MDSAP audit model covers FDA's requirements under 21 CFR 820 and the other FDA regulations typically covered during FDA inspections 172. FDA stopped accepting submissions under the earlier program after March 31, 2016 173. Operational-phase participation with Australia, Brazil, Canada and Japan was announced to begin January 1, 2017 171. CP 7382.850 gives the same history: a 2014 to 2016 pilot, then the operational phase from January 1, 2017. It also notes that participating regulators committed to using the submitted audit reports 242.

The two programs are often confused. Under the predecessor voluntary audit pilot, low-risk establishments could be removed from FDA's routine inspection work plan for one year. That provision belonged to the ISO 13485:2003 pilot, not to MDSAP 214333.

What an MDSAP report replaces

CP 7382.850 limits the substitution to surveillance inspections and treats "routine" and "surveillance" as the same thing 135137. In practice, MDSAP enrollment affects FDA's inspection planning in two ways:

  • Current enrollment. Sites actively enrolled in MDSAP do not receive FDA surveillance inspections, whether baseline or non-baseline 253254.
  • Audit history. A non-baseline surveillance inspection is generally used for manufacturers whose most recent FDA device inspection or MDSAP audit had a final classification of NAI or VAI 244. The "Inspection Types" table applies this only to manufacturers that are not currently enrolled in MDSAP 245. A baseline inspection fits where there is no FDA inspection or MDSAP audit history, or where risk factors call for a baseline evaluation 254.

Investigators assigned to an MDSAP-participating manufacturer must discuss the assignment with their supervisor before the inspection starts 255138.

The QMSR final rule (89 FR 7496, February 2, 2024) describes the same arrangement. MDSAP gives a single QMS audit based on ISO 13485 plus applicable FDA device requirements. FDA "may accept [it] instead of routine surveillance inspections", and participation is voluntary 277. FDA uses the MDSAP audit reports, not the certificate, as an additional oversight tool 278.

Inspections that still apply to MDSAP participants

Inspection typeTreatment of MDSAP participants in FDA records
For-causeMDSAP participants "may be inspected as part of a for-cause assignment" 248. "MDSAP Regulatory Audit Review" is listed as one source of the information that can prompt a for-cause inspection 247144.
Compliance follow-upMDSAP participants may be inspected under a compliance follow-up assignment 251. An earlier MDSAP audit that resulted in regulatory action is one listed trigger 245.
Specific Product Risk Assignment (SPRA)MDSAP participants "may be inspected as part of a SPRA assignment" 252.
PMA preapproval / PMA postmarketListed as separate inspection types and covered by the statement that all manufacturers remain subject to non-surveillance inspections 245250136. The 2013 pilot notice excluded these inspections expressly 169.
Electronic Product Radiation Control (EPRC)Manufacturers remain subject to inspection of their EPRC activities 136. For diagnostic X-ray makers whose quality system has been evaluated under MDSAP, the inspection may cover EPRC requirements only 146. Investigators should not use such an inspection to look for quality-system deficiencies or cite them on a Form FDA 483 147.
Combination productsIn the QMSR rule, FDA declined to apply the MDSAP model to device-led combination products because they were outside FDA's MDSAP scope 285. CP 7356.000 says third-party audit programs, including MDSAP, may affect its implementation and tells staff to contact the lead center 140141.

FDA's remote regulatory assessment (RRA) Q&A guidance adds that FDA "generally does not intend to conduct RRAs during Medical Device Single Audit Program (MDSAP) audits" 155.

How FDA reviews MDSAP reports and uses them in compliance decisions

CP 7382.850 states that "MDSAP audit reports submitted by MDSAP Auditing Organizations that include the United States as a jurisdiction are reviewed and classified by the FDA" 136. The program does not set out MDSAP-specific classification criteria or name an OAI equivalent for MDSAP reports. What it does set out is how audit results feed FDA's regulatory decisions:

  • Quality data for compliance decisions. Under "Factors Considered by Compliance," MDSAP audit results are among the quality data CDRH may weigh when it evaluates the overall evidence and makes a regulatory decision 139.
  • Compliance actions from report review. The program refers to "compliance actions and communications resulting from Agency evaluation of Regulatory Audit Reports (RARs) submitted by recognized third-party Auditing Organizations" under MDSAP. It points readers to FDA's MDSAP information and to [email protected] rather than listing specific actions 249.
  • Escalation to inspection. Report review can lead to a for-cause inspection 247. A regulatory action based on an MDSAP audit can lead to a compliance follow-up inspection 245.

IMDRF's MDSAP documents describe the same approach across regulators. Each authority uses audit reports in line with its own legislation 90. The reports record the extent to which the manufacturer met the MDSAP audit criteria 89. IMDRF N8 recognizes that regulators may have taken actions affecting marketing authorizations and enforcement on the basis of audit information. If an Auditing Organization later changes a report or a conformity decision, it should notify the authority so that the authority can assess the effect on those actions 93.

MDSAP reports compared with ISO 13485 certificates under the QMSR

The QMSR now incorporates ISO 13485:2016 by reference. FDA notes in the rule that ISO 13485 is the standard used in MDSAP 176177. That does not make an ISO certificate an inspection substitute. In its responses to comments, FDA said:

  • FDA cannot be sure that independent ISO 13485 audits cover other FDA requirements, such as 21 CFR parts 803, 806, 821 and 830. FDA does not oversee non-MDSAP auditing organizations or evaluate reports issued outside MDSAP 279.
  • FDA will not require ISO 13485 certification. It will not accept a certificate as a substitute for FDA oversight, including an inspection or an Establishment Inspection Report 280.
  • The rule does not change FDA's inspection authority. "Manufacturers with a certificate of conformance to ISO 13485 are not exempt from FDA inspections" 283284.
  • In the 2024 rule, FDA said it would review the MDSAP audit approach and Auditing Organization training and revise them as needed to align with the QMSR 276.

FDA's least burdensome guidance presents MDSAP as an example of least burdensome principles. One audit satisfies several jurisdictions, with fewer regulatory audits, fewer redundant requests and less business disruption 261.

What FDA records show about inspection and enforcement outcomes

Inspection volume. The main effect FDA documents is structural: enrolled sites are taken out of the surveillance inspection pool 253. FDA's public program documents do not give a count of inspections that MDSAP replaced. Secondary reporting gives some scale. RAPS reported in November 2018 that CDRH had received and classified nearly 600 MDSAP audits from 2013 to 2017, in addition to FDA's own inspections 213. The same coverage reported growth in both foreign and US device inspections over that period. It did not report any measured reduction caused by MDSAP 221222. Reported participation grew from 157 manufacturers at the end of 2016 to 326 in the second half of 2017 216. By September 2018, 2,629 sites worldwide were participating, more than half of them in the US 218219.

Warning letters. The record shows little direct link between MDSAP and FDA enforcement letters. Of 24 warning letters returned by a full-text search for MDSAP terms, only two mention MDSAP, and neither cites an MDSAP audit report as the basis for the violations:

  • Steiner Biotechnology, LLC (May 19, 2023). Under the quality-audit charge, FDA noted that the firm had provided "MDSAP task audit form templates for 2020, 2021, and 2022," but these covered design controls only and not other quality system requirements 237238.
  • DRG Instruments GmbH (March 31, 2025). MDSAP appears in the firm's corrective-action plan, which committed to reviewing MDSAP guidance documents for US-specific requirements 297296. The letter does not say that FDA reviewed an MDSAP audit report or that MDSAP findings led to the inspection 294.

The remaining letters, including those to Olympus Medical Systems 183, Cardinal Health 258, Royal Philips 264, BYD Precision Manufacture 181 and Jiangsu Caina Medical 360, do not mention MDSAP. Warning letters do not normally state a firm's MDSAP enrollment. The absence of MDSAP from a letter therefore does not show that the firm was not a participant.

Pilot-era enforcement posture. During the pilot, FDA's Kim Trautman said FDA had committed not to issue warning letters while the program was in its learning phase 224. That was a pilot-period commitment. CP 7382.850 now lists MDSAP-based regulatory action and RAR-driven compliance actions as possible outcomes 245249. Separately, FDA indicated it did not intend to require MDSAP participation for devices on the US market 225.

Inspection classification data. In the FDA device inspection records searched, no record mentioned MDSAP 678. FDA's MDSAP report classifications therefore cannot be matched to MDSAP participants from those public records.

Practical implications for device manufacturers

  • MDSAP enrollment takes a site out of FDA's routine surveillance pool 253. It does not prevent for-cause, compliance follow-up, SPRA, PMA or EPRC inspections 136248251252.
  • An MDSAP audit report is a regulatory document that FDA classifies 136. Its findings can feed CDRH compliance decisions 139, prompt for-cause inspections 247 and support regulatory action that leads to compliance follow-up 245.
  • After an FDA or MDSAP outcome of NAI or VAI, a manufacturer that leaves MDSAP is generally placed in non-baseline surveillance 244245.
  • Under the QMSR, conformance to ISO 13485 does not substitute for FDA oversight. MDSAP audit reports are the only third-party audit output FDA uses in place of routine surveillance inspections 278280.
  • Combination products are outside FDA's MDSAP scope 285. Diagnostic X-ray manufacturers whose quality system is covered by MDSAP may receive EPRC-only inspections, and other radiation-emitting product makers remain subject to EPRC inspection 146147.

Open questions in the public record

The FDA documents reviewed do not publish MDSAP-specific classification criteria, the share of MDSAP reports classified as requiring action, or counts of for-cause inspections prompted by MDSAP report review. Questions for further research include how FDA's QMSR-era inspection approach changes MDSAP audit criteria 276, and how individual firms' MDSAP histories compare with their FDA inspection and warning letter records.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.