The Q-Submission program is the main formal channel a device sponsor has to test its testing strategy, clinical data plan and regulatory pathway with FDA before filing a 510(k) or De Novo request. FDA does not publish Pre-Sub written feedback or meeting minutes, so regulatory teams that are planning their own Pre-Sub questions have little public precedent to go on when they try to predict what FDA will challenge.

The analysis below draws together the public record that does exist: 510(k) summaries and FDA decision summaries that cite Q-numbers, FDA's Q-Submission and De Novo guidances, and Pre-Sub outcomes that companies disclosed in SEC filings. It sorts the recurring feedback themes by device area and traces how each one showed up in the final submission and in FDA's decision.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.

FDA Pre-Submission feedback before 510(k) and De Novo submissions: common themes by device type and how they appear in final submissions

FDA does not publish Pre-Submission (Pre-Sub) written feedback or meeting minutes. The public record of what FDA told sponsors comes from three places. First, 510(k) summaries and FDA decision summaries sometimes cite a Q-number and describe what was agreed. Second, FDA's own Q-Submission and De Novo guidances say which topics FDA expects to be raised. Third, public companies sometimes disclose Pre-Sub outcomes in their SEC filings. Read together, these sources show a consistent pattern. Across device types, FDA's Pre-Sub feedback focuses on test protocol design and acceptance criteria, whether clinical data are needed, the scope of the indications for use, and, for novel devices, the regulatory pathway and predicate availability. The specific technical focus depends on the device area.

What a Pre-Sub is and how its feedback is supposed to carry forward

FDA describes a Pre-Sub as a way to get feedback on specific questions that will guide product development or the preparation of a planned submission, including a 510(k) or De Novo request. It is not a pre-review of the intended submission or its data 113. The topics FDA lists include non-clinical and clinical testing protocols 114. They also include benchtop performance testing, biocompatibility, animal studies, PCCPs, software and SaMD questions, sterility and shelf life, clinical study endpoints, and statistical analysis plans. FDA recommends no more than 3 to 4 substantial topics per Pre-Sub 115.

Key procedural points that determine how feedback reaches the final submission:

  • Written feedback is the official record when no meeting is held. FDA intends to provide it within 70 calendar days of the review clock start 123. If a meeting is held, the written feedback plus sponsor-drafted, FDA-reviewed minutes form the official record 124125.
  • Feedback is nonbinding but intended to be stable. FDA describes it as its "best advice" based on the information available at the time. FDA intends not to change it if the later submission is consistent with the Pre-Sub and no new information raises important safety or effectiveness issues 131132. If more than a year has passed since feedback on significant study-design topics and the study has not started, FDA recommends confirming that the advice still applies 134.
  • Silence is not agreement. If FDA did not comment on a topic outside the Pre-Sub questions, that does not mean no further information will be needed on that topic later 135.
  • The 510(k) or De Novo request should close the loop. FDA recommends identifying related Q-Subs in the cover letter. It also recommends a dedicated section that references prior FDA communications and explains how the feedback was addressed, including when the sponsor used an alternative method 137138.

For De Novo requests, the bar is higher. The De Novo acceptance review guidance tells sponsors to identify prior submissions for the same device, including Pre-Subs, by number. Sponsors should provide or cross-reference FDA's feedback and explain where the request addresses it 12131214. If FDA previously identified deficiencies and the sponsor neither responds nor explains why not, FDA may refuse to accept the request 1215. The De Novo eSTAR has a dedicated "Pre-Submission Correspondence & Previous Regulator Interaction" section 1216.

FDA also links Pre-Subs to review efficiency in 510(k)s. When data requirements cannot be determined from guidance or posted 510(k) summaries, for example because the device has a new technology or feature, FDA's position is that the applicant should have used the Pre-Submission Program to get recommendations 1168.

Feedback themes by device type

The table below summarizes recurring Pre-Sub feedback themes that sponsors documented in cleared 510(k)s and, for novel devices, in their own public filings. The sections that follow give representative cases.

Device areaRecurring Pre-Sub feedback themes in the public recordRepresentative examples
In vitro diagnosticsAnalytical study design tied to CLSI protocols; which studies can be omitted for minor changes; strain/isolate selection; reproducibility; pediatric claims; cleaning/disinfection robustnessBinding Site IgA/IgM kits 8851046, JBAIDS kits 682660, Verigene RP Flex 760, Mindray BC-5390 756
Radiology imaging and imaging softwareWhether clinical data are needed; AI algorithm validation; narrowing indications (for example, removing "screening")PerkinElmer XRPad 743, Philips Radiography 7300 C 86, TeleScan 1120
Orthopedic implantsMechanical test protocols and methods; biocompatibility route; MR safetyBodycad UKS 853, AmorChem screw 746, Medacta 826830, KLS Martin 265
Cardiovascular and patient monitoringAdditional testing for expanded use environments and populations; software or alarm changesCardiacLinx 809, Current Health System 11021071
General and plastic surgeryNeed for clinical data versus bench and animal data; biocompatibility after process changes; shelf lifePre-B Seal lung biopsy plug 1077, Echelon SLR 1110, TYRX envelopes 1023
OTC consumer devicesSelf-selection and labeling-comprehension thresholdsiGrow 740, Capillus 860865
Gastroenterology/urology and renalPredicate strategy; labeling warningsLiberty Select Cycler 990, Fresenius acid concentrates 922977
DentalBiological evaluation scopeCerasmart Forte 1264
Novel devices headed to De NovoPathway (De Novo vs 510(k)); whether a predicate exists; need for human datagammaCore 1245, Vicarious Surgical 1252, Titan Enos 12481249

In vitro diagnostics: analytical study design and right-sizing the data package

IVD summaries contain the most detailed public record of Pre-Sub agreements. This is largely because FDA's IVD decision summaries describe the study design in detail.

  • Agreement on CLSI-based protocols. The Binding Site's Special 510(k)s for IgA and IgM kits on the Optilite and SPAplus platforms all relied on one Pre-Sub meeting (Q171503). Precision studies followed CLSI EP5-A3 "as was agreed in pre-submission meeting Q171503" 88510461016. Real-time stability followed EP25-A 888292295. LoQ validation followed CLSI EP17-A2 889293296, and the method comparisons were also run in line with the meeting 3121047298. The submissions reported the results against those protocols. Examples include LoQ claims validated with four samples across two reagent lots against an 8% or 10% CV criterion 293893, and method comparisons on serum and plasma samples using Bland-Altman and Passing-Bablok analyses 1047312. Real-time stability was still ongoing at the time of submission, as the Pre-Sub agreement allowed 292302.
  • Feedback that removes work. For the JBAIDS Plague and Anthrax detection kits, FDA's decision summaries record that no analytical reactivity, interfering species, or stability study "was recommended during the presubmission" because primers, probes and stability claims were unchanged 682683684660661659. Both submissions cross-referenced the prior clearances instead of providing new data 685662. The Plague kit decision summary concluded that the information supported a substantial equivalence decision 687.
  • Strain and isolate selection. Nanosphere selected strains for the Verigene RP Flex analytical reactivity study using FDA recommendations from the pre-submission review, which led to wet testing of 136 strains in total 760. For MicroScan ciprofloxacin panels, Beckman Coulter agreed in Q160981 to try to collect freshly recovered clinical isolates. When no fresh Salmonella Typhi isolates could be obtained, the submission used 74 challenge isolates at three external sites 1008.
  • Reproducibility and special populations. Thermo Fisher added two reproducibility studies to the Sensititre submission "as recommended by the pre-submission (Q141181)" 813. Mindray included pediatric subjects in the method comparison and validated literature reference intervals to support a pediatric claim, "according to FDA's recommendation in the pre-sub" 756.
  • Protocols agreed up front for complex analyzers. CellaVision's DC-1 conclusion states that the performance studies followed "study protocols agreed by FDA in pre-submission (Q171293)" covering all intended cell types and morphologies 880. The clinical and analytical evaluations were reported as successful 882.
  • Leveraging prior data. Athelas used studies from the cleared Athelas One to support Athelas Home "as per Agency guidance in presubmission (Q190305)." A humidity flex test was on the final flex-test list agreed with the Agency 905908.
  • Reprocessing and disinfection. For iBGStar, FDA recommended disinfection and robustness testing of new components in Q120016 774. The 510(k) reported 260 cleaning/disinfection cycles simulating five years of use 776. Abbott designed the i-STAT Alinity hematocrit cleaning robustness study based on Pre-Sub Q130270 843. That submission reported that performance through 10,950 cycles was not supported for all sample types tested 844. Agreeing a protocol does not guarantee a favorable result.
  • Cybersecurity. Stago's coagulation analyzer 510(k) records cybersecurity documentation "agreed with FDA" during Q181507 1279.

Radiology imaging and imaging software: clinical data and indications

  • Whether a clinical study is needed. PerkinElmer's XRPad 4336 MED 510(k) states that no clinical studies were conducted "as agreed upon during PreSubmission discussions with the Agency for the predicate device," and that a clinical concurrence study was unnecessary 743. Kubtec's DIGIVIEW 395 later used the same rationale, citing Pre-Sub discussions for its own predicate 797. Later sponsors have cited an earlier device's Pre-Sub agreement on clinical data as a rationale, though FDA's advice holds only while the device, intended use and labeling stay consistent with the original Pre-Sub.
  • AI algorithm validation. Philips's Radiography 7300 C 510(k) has a section titled "Pre-submission Q222058 related Testing." It provides information "as requested by FDA," including the Smart Collimation Thorax algorithm validation plan and report, plus the AI training and validation data. Statistical testing of all performance endpoints showed superiority or non-inferiority against a clinical user 86.
  • Indications scoping. TeleScan's predicate comparison table states that "screening" was removed from the indications "per FDA pre-submission guidance" 1120. The final indications refer only to diagnostic procedures 1121. In the nView s1 navigation 510(k), indications wording was changed for formatting only, "as discussed in a pre-submission" 1095.

Orthopedic implants: test methods and acceptance criteria

Orthopedic 510(k)s typically cite Pre-Subs as the source of the bench-testing plan.

  • Bodycad's unicompartmental knee 510(k) performed tibial baseplate fatigue testing with a three-point bend method, referencing ASTM WK45235 and ASTM F1800-12, "as recommended by FDA in pre-sub Q150560" 853.
  • Medacta's tibial cones and shoulder system 510(k)s state that testing was based on the risk analysis and a pre-submission "submitted to FDA to review testing protocols," with acceptance criteria based on standards 826830.
  • For biocompatibility, AmorChem used the ISO 10993-1 chemical characterization and risk analysis route "recommended by FDA in pre-submission Q150543" 746.
  • The FINE Osteotomy System's preclinical testing was performed per standards, the literature, and Pre-Sub Q190553 262. The program covered screw and plate mechanics, simulated osteotomy biomechanics, cadaver accuracy, software V&V, biocompatibility, and reprocessing 1056. The CITREFIX suture anchor's performance evaluation was "guided by" Pre-Sub Q170251. It covered pull-out testing at time zero, after soaking, and after 5,000 cycles of simulated loading against a same-size predicate 939941.
  • For MR safety, KLS Martin's 510(k) described RF-induced heating for its thoracic and hand implant portfolio as "the focus of this pre-submission." It used computational modeling with an FDA-qualified MDDT instead of physical ASTM F2182 testing 265266.

Cardiovascular and patient monitoring: expanded environments and software changes

  • CardiacLinx documented that Pre-Sub Q130826 and its approved minutes (Q130826/A001) determined that additional testing could support expanding an ECG monitor's indication to ambulatory and home use. The sponsor ran an ambulatory and home-environment interference test 809.
  • Current Health used a single Q-Submission (Q212340) for successive modifications. These were an alarm-functionality change 1102 and, in a later Special 510(k), an extension of the age range to patients aged 14 and older, which "was also presented in the Q-Submission" 1071. FDA found the later submission substantially equivalent 1074.
  • Several cardiovascular 510(k)s list a completed Pre-Sub only in the submission history without describing its content, for example Abbott's EnSite mapping systems (Q191312) 1028.

General and plastic surgery: clinical data, sterilization changes and shelf life

  • Bench and animal data instead of clinical data. The Pre-B Seal lung biopsy plug 510(k) states that "per pre-submission correspondence with the FDA no clinical data was required." The submission relied on design verification, sterility, biocompatibility, usability, and a comparative GLP animal study 10771078.
  • Biocompatibility after a process change. For an increased sterilization dose on Echelon Endopath staple line reinforcement, Ethicon included a retrospective comparison of predicate and subject implant studies "as agreed upon with the FDA during the Pre-Submission Q231522." It also included a canine GLP study at 40 kGy 11101108.
  • Leveraging prior validation. Integra's Kerrison rongeurs labeling change relied on prior steam sterilization validation, which was "deemed acceptable to support the proposed labeling change per Pre-Submission #Q141112" 794.
  • FDA recommending a Pre-Sub at clearance. In the TYRX absorbable antibacterial envelope clearances, FDA stated that the shelf-life protocol "cannot be considered approved" because the device is a combination product. FDA recommended a Pre-Sub before future device changes 1023972.

OTC consumer devices: quantitative human factors thresholds

Two laser hair-growth 510(k)s show a quantitative Pre-Sub requirement. In both, the Pre-Sub set the OTC success criterion for self-selection and labeling comprehension. The iGrow submission reports that a minimum 80% success rate was required for OTC use 740. For Capillus, a Q-Sub response (Q160385) set the same 80% threshold 860. The submission then reported 94.6% for self-selection and 90.9% for instruction comprehension in 55 adults, described as satisfying FDA's requirements for OTC clearance 864865.

Renal, gastroenterology and dental devices

  • Fresenius reached agreement with FDA on a final warning statement for acid concentrate labeling in June 2018 through a series of Pre-Sub correspondence (Q131379). The 510(k)s for its liquid and dry concentrates cite that agreement 922977.
  • The Liberty Select Cycler 510(k) names its primary predicate "as described in the 8 September 2016 pre-submission (Q161532)" 990.
  • GC America's Cerasmart Forte 510(k) states that its biological evaluation was based on FDA suggestions from Q-Sub Q260748. The evaluation followed ISO 10993-1 and ISO 7405, with cytotoxicity, sensitization and irritation testing 126412671268.

De Novo: pathway, predicate and clinical evidence

FDA calls a Pre-Sub before a De Novo request "optional but strongly recommended," particularly for devices FDA has not previously reviewed. The Pre-Sub gives preliminary feedback on De Novo eligibility, possible predicates, likely controls, and the nonclinical and clinical evidence needed 1199. Recommended content includes a proposed class with a rationale and database searches, a list of risks to health, proposed general and special controls, and planned study protocols 1202120312041205. Suggested questions cover eligibility, additional risks, the adequacy of controls, and clinical study design and control groups 1206. The 2025 Q-Submission guidance discourages asking for a formal classification or De Novo eligibility determination in a Pre-Sub. It directs formal classification questions to a 513(g) request 12101211.

FDA's own De Novo decision summaries rarely describe Pre-Sub content. In the decision summary text searched for this article, none described Pre-Sub feedback that shaped the request 1155. Sponsors' SEC disclosures fill part of that gap:

  • Pathway confirmation, then denial and resubmission. electroCore reported that at a June 2014 pre-submission meeting, FDA confirmed De Novo was an appropriate pathway for gammaCore. The first De Novo request was denied in May 2015. The company resubmitted for narrower cluster headache indications and later provided a re-analysis FDA requested 124512461247. A Pre-Sub pathway agreement did not ensure the first request would be granted.
  • No predicate, so De Novo. Vicarious Surgical disclosed that FDA determined at a February 2022 pre-submission meeting that there was no legally marketed predicate for its surgical system. The company planned a De Novo request and an IDE pivotal study 12521254. It later entered FDA's Safer Technologies Program, which the company noted was not a marketing authorization 1253.
  • Human data and pathway uncertainty. Titan Medical reported confirming through a Pre-Submission that confirmatory human data would be required 12481249. Later FDA communications questioned whether robotically assisted surgical devices would remain eligible for 510(k) or would need De Novo requests 1251.

How Pre-Sub feedback shows up in the final submission

Across the record, Pre-Sub feedback appears in a small number of recognizable forms:

  1. A cited Q-number next to a specific test. This is the strongest form, for example "as agreed in pre-submission meeting Q171503" 885 or "as recommended by FDA in pre-sub Q150560" 853.
  2. A dedicated section. Examples are "Pre-submission Q222058 related Testing" 86, "Prior Formal Correspondence" listing a Pre-Sub number 697, or "Previous correspondence with FDA" noting two Pre-Sub meetings 1113. Some manufacturers include a standing "Pre-Submission, IDE" heading even when the answer is "Not applicable, there is no prior submission" 782805.
  3. A justification for omitting data. Examples are no clinical study 7431077 and no new analytical studies for unchanged assay components 682.
  4. Changed indications or labeling. Examples are removal of "screening" 1120 and an agreed acid concentrate warning statement 922.
  5. Reported results against an agreed threshold. One example is OTC comprehension rates against an 80% criterion 865.

Many 510(k) summaries only list a Q-number in the submission history with no detail, for example iSchemaView RAPID 958, MFI-Cardiac 934 and FlowMet-R 912. The Pre-Sub is visible in these cases, but its content is not. Before the current Q-Sub program, FDA's third-party 510(k) review records included a checklist item rating the "Extent of pre-submission consultation with ODE/OIVD division" (for example, "Adequate" or "Moderate") next to the substantial equivalence decision 768690769.

Practical implications for sponsors

  • Ask about protocols and acceptance criteria, not for a review of results. FDA's feedback that later appears in submissions is almost always about study design. This matches the guidance's statement that a Pre-Sub is not a pre-review of data 113.
  • Use the Pre-Sub to justify what you will not do. Some of the most useful documented outcomes are agreements that a clinical study or a new analytical study is unnecessary 7436821077.
  • Map every feedback item in the submission. FDA expects a section explaining how each item was addressed, including any deviations 138. In a De Novo request, unaddressed deficiencies can lead to refusal to accept 1215.
  • Expect agreements to be tested by the data. An agreed protocol can still produce results that do not fully support a claim 844. An agreed pathway can still end in a denial before a narrower resubmission succeeds 1246.
  • Keep the Pre-Sub current. If significant study-design advice is more than a year old and the study has not started, reconfirm it 134. Present significant changes in the device or indications in a new or supplemental Q-Sub, as several sponsors did for successive modifications 10711102.

Evidence limits

This picture is drawn from sponsor-written 510(k) summaries, FDA decision summaries, FDA guidance and company SEC filings. FDA's Pre-Sub written feedback and minutes are not public, so the record shows what sponsors chose to attribute to Pre-Subs, not everything FDA said. Many summaries name a Q-number without describing its content. FDA's De Novo decision summaries generally do not discuss Pre-Sub interactions. The device-type breakdown above is therefore qualitative and does not measure how often each topic arises across all Pre-Subs.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.