Before approval, drug and biologic sponsors talk about their investigational products constantly: at medical congresses, in peer-reviewed publications, in investor materials, on websites and with payors. Every one of these channels is close to the line FDA draws between legitimate scientific exchange and premature promotion. Regulatory, medical affairs and commercial teams need to know where that line sits, and how FDA has enforced it in practice, to plan pre-launch activity without inviting an untitled or warning letter.
The analysis below traces FDA's position from the regulatory baseline in 21 CFR 312.7 and 812.7 through the guidance documents and enforcement policies on scientific exchange, distribution of publications on unapproved uses, responses to unsolicited requests and payor communications. It then reviews untitled letters and warning letters that cite pre-approval promotion, and identifies the patterns of claims and presentation that FDA has treated as crossing into promotion of an investigational product.
FDA on pre-approval communications: scientific exchange, publications and promotion of investigational products
FDA's position on pre-approval communications rests on one regulatory line. Sponsors may share scientific information about an investigational product. They may not represent, in a promotional context, that the product is safe or effective for the use under investigation, and they may not commercialize it before approval. FDA guidance and enforcement policies address that line: independent education, publications on unapproved uses, responses to unsolicited requests, and payor communications. Untitled letters and warning letters show where FDA thinks companies have crossed it. In the enforcement record, the deciding factor is typically not that a company discussed an investigational product. It is usually that the company drew conclusions about safety, efficacy or superiority that the data could not support, often without making clear that the product was unapproved.
The regulatory baseline: 21 CFR 312.7 and 812.7
Drugs and biologics. Under 21 CFR 312.7(a), a sponsor, investigator, or anyone acting on their behalf "shall not represent in a promotional context that an investigational new drug is safe or effective for the purposes for which it is under investigation or otherwise promote the drug" 2. The same paragraph contains the scientific-exchange carve-out. The provision "is not intended to restrict the full exchange of scientific information concerning the drug, including dissemination of scientific findings in scientific or lay media." Its stated purpose is "to restrict promotional claims of safety or effectiveness of the drug for a use for which it is under investigation and to preclude commercialization of the drug before it is approved for commercial distribution" 2. Related provisions bar a sponsor or investigator from commercially distributing or test marketing an investigational new drug 3. They also bar a sponsor from unduly prolonging an investigation once results appear sufficient to support a marketing application 4.
Devices. Under 21 CFR 812.7, a sponsor, investigator, or anyone acting on their behalf may not "promote or test market an investigational device, until after FDA has approved the device for commercial distribution" 297. They also may not represent "that an investigational device is safe or effective for the purposes for which it is being investigated" 298. The device rule has no counterpart to the drug rule's express scientific-exchange sentence. In practice, FDA's device warning letters focus on the same thing as the drug letters: safety and effectiveness representations.
How enforcement works. In untitled letters, OPDP links a 312.7 violation to misbranding. A company that promotes an investigational drug loses the investigational-use exemption from the "adequate directions for use" requirement. The drug is then misbranded under FD&C Act section 502(f)(1), and introducing or holding it violates section 301(a) or 301(k) 72737980.
Scientific exchange: what FDA treats as non-promotional
The regulation protects scientific exchange but does not define it. FDA's guidance documents describe it mostly by context, not by definition.
- Independent, industry-supported education. The 1997 guidance Industry-Supported Scientific and Educational Activities says FDA does not intend to regulate industry-supported scientific and educational activities under the labeling and advertising provisions when they are independent of the supporting company's influence 25. FDA permits support for the "full exchange of views in scientific and educational discussions, including discussions of unapproved uses" when activities are nonpromotional and independent 34. Independence turns on whether the company can influence content. That influence can be direct (speaker or topic selection) or indirect (a provider's belief that future funding depends on favorable programs) 29. Other factors FDA considers include company-directed redistribution of program materials after the event 30, sales-representative or promotional-exhibit activity in the meeting room 31, and complaints about attempts to influence content 32. A written agreement giving the provider sole control is helpful evidence but is not required 33. Programs subject to substantive company influence may not promote unapproved uses 3536.
- Clinical trial recruitment. FDA's guidance on recruiting study subjects says advertising should make no explicit or implicit claim that an investigational drug, biologic or device is safe or effective, or equivalent or superior to another product. FDA calls such claims misleading and violations of 21 CFR 312.7(a) and 812.7(d). It also says terms like "new treatment" or "new drug" should not be used without explaining that the product is investigational 190.
- Public disclosure of research results. In the emergency research (exception from informed consent) guidance, required public disclosure of study results remains subject to the investigational-promotion rules. Sponsors and investigators must not represent in a promotional context that the investigational product is safe or effective 191192, and the same applies to disclosures before the study ends 193. FDA expects disclosures that let readers understand the study design and both positive and negative results 194.
- Expanded access policies. FDA does not intend to treat a Cures Act expanded-access policy posting as promotion or as evidence of a new intended use. That changes if the posting represents, in a promotional context, that the investigational drug is safe or effective 196197.
- IND-exempt studies. Several IND-exemption guidances make compliance with 312.7 a condition of exemption. They describe it as the investigation not being intended to promote or commercialize the product 198199200. For studies of lawfully marketed oncology products, the investigation must not be used to promote unapproved indications 203204.
Dissemination of publications and scientific information on unapproved uses
The publications framework covers unapproved uses of approved or cleared products. It does not cover products that have no approval at all. Sponsors of investigational products should not assume it gives them a safe harbor.
Evolution of the guidance. FDA's 2009 Good Reprint Practices guidance addressed the distribution of journal articles and reference publications on unapproved new uses 287. In February 2014, FDA proposed to revise it with the draft Distributing Scientific and Medical Publications on Unapproved New Uses: Recommended Practices, which added recommendations on reference texts and clinical practice guidelines 285286. An October 2023 revised draft replaced the 2014 draft, and on January 6, 2025, FDA finalized it as Communications From Firms to Health Care Providers Regarding Scientific Information on Unapproved Uses of Approved/Cleared Medical Products: Questions and Answers (the SIUU guidance) 275. The guidance is nonbinding. It is marked "Not for Current Implementation" because its information-collection provisions in Q2, Q4 and Q5 were under OMB review 288316.
SIUU framework. The 2025 guidance is an enforcement policy for certain firm-initiated communications to health care providers 317. Covered source publications include journal reprints, clinical practice guidelines, reference texts and digital clinical practice resources. These must carry recommended disclosures and may be accompanied by a firm-generated presentation 318290. The core recommendations are:
- Communications should be truthful, non-misleading, and complete enough for HCPs to evaluate the strengths, weaknesses, validity and clinical utility of the evidence 292.
- Disclosures should state that the use is unapproved and its safety and effectiveness have not been established. They should also give the approved uses, limitations, and relevant labeling cautions 319. Firms should provide access to current labeling, contraindications, serious risks, REMS information, and relevant financial relationships with publication contributors 320.
- Firms should supply material study design, methodology and results missing from the source publication, together with material limitations and conflicting findings from other sound studies 321322.
- Disclosures should be clear and prominent. In audiovisual formats, they should appear in audio and text at the same time 323.
- Firm-generated presentations must stay within the scientific content of the source publications, include those publications, and clearly identify which portions the firm produced 325. The policy excludes presentations that use persuasion unrelated to the science or "calls to value" 326.
- Reference-text excerpts should be unaltered and unabridged and include the context needed to interpret them 327. Reference texts must meet independence and peer-review-related criteria 291.
If firms follow these recommendations, FDA does not intend to use the dissemination, standing alone, as evidence of a new intended use. FDA also does not expect submission of the communication when it is first shared with HCPs 294.
Unsolicited requests. The December 2011 draft guidance Responding to Unsolicited Requests for Off-Label Information About Prescription Drugs and Medical Devices remains a draft 305328. The 2025 SIUU guidance still refers to it as a draft 26. A request is unsolicited only if it comes from someone completely independent of the firm. Requests the manufacturer prompts are solicited and can be evidence of intended unapproved use 309225. The draft gives a "Coming Soon, a new use for Product X" exhibit as an example of a prompt that turns resulting requests into solicited ones 224. Responses should be balanced, scientific in tone, and handled by medical or scientific staff independent of sales and marketing. They should not accompany promotional material 305329. Responses to public requests should be moved into private channels 310.
Pre-approval communications to payors
FDA's June 2018 final guidance on payor communications remains the version currently in effect. On June 2, 2026, FDA issued a revised draft of the same guidance, Drug and Device Manufacturer Communications With Payors, Formulary Committees, and Similar Entities: Questions and Answers, marked "Draft — Not for Implementation" 36138. It implements FD&C Act section 502(gg). That provision says products are not misbranded under 502(f)(1) because of certain truthful, non-misleading information given to payors about investigational drugs or devices, or about investigational uses of authorized products 43.
| Permitted "product information" (Q11) | Scope |
|---|---|
| Product description | Drug class, device description and features 370 |
| Indications under investigation | Endpoints and patient population, including enrollment numbers, criteria and demographics 371 |
| Anticipated timeline | Possible timing of FDA approval or clearance 372 |
| Pricing | Medical product pricing information 373 |
| Patient utilization projections | For example, epidemiological incidence and prevalence projections 374 |
| Product-related programs or services | For example, patient support programs 375 |
| Factual study results | Clinical and bench results, without characterizations or conclusions about safety or efficacy 376 |
Required and recommended accompanying information (Q13) includes:
- A clear statement that the product or use is not approved or cleared and that safety or effectiveness has not been established 378.
- The stage of development, including study status and planned or actual submission timing 379.
- Non-selective presentation of results, including positive and negative or null findings, with all material design, methodology and limitation information 380.
- Updates to payors when previously communicated information becomes materially outdated 382.
The draft proposes that FDA would not object under 21 CFR 312.7(a), 812.7(a) or 812.7(d) to communications made in accordance with the guidance, or to treat them, standing alone, as evidence of a new intended use 384.
Enforcement: OPDP untitled letters
OPDP has issued untitled letters headed "Misbranding of an Investigational Drug" that cite 312.7 directly 67818897. The pattern is the same across settings: conference exhibit halls, corporate websites and academic medical center web pages.
| Company | Product | Letter date | Communication | Core objection |
|---|---|---|---|---|
| UCLA | Ga-68 PSMA 8 | Dec 28, 2017 52 | Webpage and linked brochure 98 | "Major advance," greater accuracy, "no common side effects or significant risks" for an unapproved imaging drug 99100 |
| Arog Pharmaceuticals | Crenolanib besylate 7 | Jun 28, 2018 57 | ASH exhibit booth and product webpage 89 | "Combinable with chemotherapy at full doses," "Eradicating Activating Mutations," implied superiority over approved therapies 909293 |
| Phoenix Molecular Imaging Center | Sodium acetate C-11 9 | Feb 15, 2019 49 | Webpage on PET/CT imaging 82 | Claimed better performance than approved PET agents 8384 |
| Nascent Biotech | Pritumumab 6 | Nov 1, 2019 54 | Company website 68 | "Has cured a rare form of brain cancer," survival comparison "demonstrating antibodies are safe and effective" 69 |
| Alar Pharmaceuticals | ALA-3000 (ketamine pamoate) 5 | Aug 21, 2026 51 | APA Annual Meeting exhibit booth and brochure 74 | Claimed no sedation, dissociation or abuse signals, and possible elimination of post-dose monitoring, based on apparently Phase 1 data 7576 |
The letters raise the same issues repeatedly:
- Conclusory claims of safety and efficacy. FDA objected to claims implying established safety and efficacy when data were inadequate and the indications, warnings, adverse reactions and dosing were unknown 7076.
- Implied superiority or differentiation. Comparisons with approved products are a frequent trigger, even when phrased as mechanism or performance claims 8493.
- Missing or inadequate investigational-status disclosure. FDA has noted that a booth lacked investigational-status information and sat near booths for approved products 7894. In the Phoenix letter, a reference to expanded-access trials did not cure repeated safety, efficacy and superiority impressions 85. In the Nascent letter, the website did not clearly disclose the product's investigational status 71.
- Minimizing known class risks. In the Alar letter, FDA pointed to ketamine's Schedule III status and serious risks, including sedation, dissociation, respiratory depression and abuse 77.
FDA did not treat these communications as promotion because they discussed investigational research. It treated them as promotion because they made conclusory clinical-benefit, safety or comparative claims 104.
Enforcement: warning letters
Warning letters on investigational-product promotion come from OPDP, CBER and CDRH and often appear alongside other violations.
- CytoDyn (leronlimab), February 11, 2022, OPDP 112. The letter concerned a video interview with the company's CEO, linked from the corporate website, about an investigational COVID-19 therapy that had neither approval nor emergency use authorization 240241242. It cited claims that results were "really strong," that survival could reach 88% and then 95%, and that a discharge endpoint was "166% better" than placebo 243. FDA set out the line between scientific exchange and promotion under 312.7(a) 244. It found that the statements mischaracterized the data: the larger CD12 trial missed its primary mortality endpoint and every predefined secondary endpoint 245. This letter shows that a CEO interview carried by investor media and linked from the company website falls within 312.7, and that selective or overstated reporting of trial results is not protected as scientific exchange.
- VaxGen (investigational anthrax vaccine), March 24, 2006, CBER 107. FDA found that comparative dosage or effectiveness claims were premature at an early stage of vaccine development 259. It acknowledged the scientific-exchange protection in 312.7(a) and still concluded that the material promoted the vaccine as safe and effective, in violation of 312.6(b) and 312.7(a) 260.
- St. Jude Medical (Epicor LP Cardiac Ablation System), April 23, 2010, CDRH 108. The system was cleared for cardiac tissue ablation, but atrial fibrillation treatment was an investigational indication under an IDE 247. FDA found that website references to Cox Maze III lesions and a "classic box lesion" made "safely, effectively and reproducibly" amounted to promotion of the AF indication. It cited 812.7(a) and 812.7(d) 249250251. The letter shows that FDA reads clinical shorthand by what it means to physicians, not by whether the indication is named.
- Anulex Technologies (Xclose Tissue Repair System), February 11, 2011, CDRH 126. FDA cited promotion of an investigational device and representation that it is safe or effective under 812.7(a) and (d) 254. Annulus fibrosus repair was outside the device's clearance 252253.
- ExThera Medical (Seraph 100 Microbind Affinity Blood Filter), February 6, 2026, CDRH 111. The letter cited a podcast, a conference presentation, a promotional video and a company FAQ. Claims included that the device was "very safe," "as safe as a dialysis procedure," and "approved in the US for COVID-19" 263264265266. FDA found that the statements falsely represented US approval 267. It also said the FAQ implied an established safety profile comparable to approved devices 269. FDA noted that it had earlier asked the company to remove such claims under 812.5(b) and 812.7 268.
Practical lessons from the guidance and enforcement record
- The line is drawn by the claims, not the topic. Section 312.7(a) expressly preserves the full exchange of scientific information, including dissemination of findings in scientific or lay media 2. Conclusory statements that a product "cures," is "safe," avoids known class risks, or beats approved therapy have consistently drawn letters 697599243.
- Exhibit halls are promotional settings. Two of the five untitled letters involved commercial exhibit booths at major medical meetings (ASH and APA), where FDA noted the absence of investigational-status disclosure next to approved products 7894.
- Executive and investor communications are covered. The CytoDyn letter applied 312.7 to a CEO interview linked from the corporate website 242244. In the ExThera letter, FDA treated podcast and conference statements as promotion 263264.
- Accurate reporting of data is part of the line. Both the SIUU guidance and the payor draft call for complete, non-selective presentation of results and their limitations 321380. CytoDyn shows the enforcement risk of selective reporting 245.
- Use the defined channels. Payor communications under section 502(gg) 384, genuinely independent CME 25, and properly handled unsolicited requests 309 have explicit FDA frameworks. Firm-initiated publication dissemination under the SIUU guidance applies to unapproved uses of approved or cleared products, not to unapproved products 275317.
- Check guidance status before relying on it. The unsolicited-requests guidance is still a 2011 draft 305. The SIUU guidance was issued as final but marked "Not for Current Implementation" 316. The 2026 payor guidance is a draft 361.