Section 524B of the FD&C Act made cybersecurity a legal requirement for cyber devices. Before it, cybersecurity was a guidance expectation. Sponsors of 510(k), PMA, De Novo, PDP and HDE submissions now have to show how their device will meet statutory obligations for vulnerability management, secure design, software updates and software bills of materials. Regulatory, quality and engineering teams need to know what FDA expects in a submission, and how FDA treats unsupported or end-of-life software components, so they can avoid refuse-to-accept decisions, deficiency letters and delays to clearance.

The analysis below covers the statutory text of section 524B and FDA's current premarket cybersecurity guidance, including how the guidance changed from its 2024 draft to the 2026 final version. It also covers SBOM content expectations and how FDA treats unsupported software components. It then reviews FDA warning letters and Form 483 observations that cite cybersecurity failures, mapped to the quality system and premarket provisions they invoke.

Rhizome did this research. Let it help you with your most critical decisions by creating a free account.

FDA medical device cybersecurity under section 524B: premarket requirements, SBOMs, unsupported components and enforcement findings

Section 524B of the FD&C Act turned medical device cybersecurity from a guidance expectation into a statutory requirement for "cyber devices." FDA's current interpretation sits in the final guidance Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions (February 3, 2026) 55. That guidance follows the draft Select Updates for the Premarket Cybersecurity Guidance: Section 524B of the FD&C Act (March 13, 2024) 56116. The postmarket framework traces back to Postmarket Management of Cybersecurity in Medical Devices (December 28, 2016) 65. Over the same period, FDA warning letters and Form 483 observations have cited cybersecurity failures under design verification, design validation and risk analysis, CAPA, corrections and removals, and 510(k) requirements.

Key takeaways

  • Since March 29, 2023, sponsors of a 510(k), PMA, PDP, De Novo or HDE for a cyber device must include information showing that the device meets section 524B(b) 135136137. Under FDA's transition policy, refuse-to-accept decisions based solely on missing 524B information began on October 1, 2023.
  • Section 524B(b) has three paragraphs that together set four obligations: a postmarket vulnerability plan (b)(1), secure-by-design processes and patches and updates (b)(2), and a software bill of materials (SBOM) covering commercial, open-source and off-the-shelf components (b)(3) 139140141143.
  • For cyber devices, failing to comply with the section 524B(b)(2) requirements on device cybersecurity is a prohibited act under section 301(q) 267.
  • FDA expects the SBOM to be machine-readable and to give, for each component, its level of support and end-of-support date 164166.
  • Enforcement findings most often cite 21 CFR 820.30(g) (risk analysis within design validation) and 820.100(a) (CAPA). FDA has also cited 820.30(f), Part 806 and 807.81(a)(3)(i) in cybersecurity cases 180182186187265269281.

Which devices are "cyber devices"

A device is a cyber device only if it meets all three statutory criteria 138:

  1. It includes software that the sponsor validated, installed or authorized, as a device or in a device.
  2. It has the ability to connect to the internet.
  3. It contains technological characteristics, validated, installed or authorized by the sponsor, that could be vulnerable to cybersecurity threats.

FDA reads "ability to connect to the internet" broadly. The final guidance lists network, server or cloud connections; radio-frequency links such as Wi-Fi, cellular, Bluetooth and Bluetooth Low Energy; magnetic inductive communications; and hardware connectors capable of reaching the internet, such as USB, Ethernet or serial ports 258. The connection does not have to be continuous. FDA's example is a USB port used only briefly for servicing, which still counts as the ability to connect 259.

The obligation falls on whoever submits the application, including the manufacturer. It also applies to a modification of a cyber device when that modification requires a 510(k), PMA, PDP, De Novo or HDE submission 135136137.

The four statutory requirements of section 524B(b)

RequirementWhat the statute and FDA guidance call for
Postmarket planA plan to monitor, identify and address cybersecurity vulnerabilities and exploits within a reasonable time, as appropriate, including coordinated vulnerability disclosure and related procedures 139
Secure processesProcesses and procedures, designed, developed and maintained, that provide reasonable assurance the device and related systems are cybersecure 140
Updates and patchesUpdates and patches made available for the device and related systems: on a reasonably justified regular cycle for known unacceptable vulnerabilities, and as soon as possible out of cycle for critical vulnerabilities that could cause uncontrolled risks 141142
SBOMA software bill of materials that includes commercial, open-source and off-the-shelf software components 143

FDA states that failing to comply with any section 524B(b)(2) requirement on ensuring device cybersecurity is a prohibited act under section 301(q) of the FD&C Act 267.

Premarket documentation FDA expects

Security risk management

FDA asks for a set of connected, traceable analyses rather than a standalone security section. The security risk management report should cover threat modeling, the cybersecurity risk assessment, the SBOM, component support information, vulnerability assessments and assessments of unresolved anomalies. It should be traceable to the threat model, the risk assessment, the SBOM and testing 31.

  • Threat modeling. Threat modeling should run throughout design and cover all elements of the medical device system. The submission should document pre- and post-mitigation risks, assumptions about the system and use environment, and risks arising from the supply chain and other lifecycle activities. It should also explain why the chosen methodology was used 32.
  • Cybersecurity risk assessment. The assessment should state how risk is scored before and after mitigation, the acceptance criteria, and how security risks are transferred into the safety risk process. It should cover residual risk from both intentional and unintentional cybersecurity failures 3334.
  • Exploitability, not probability. FDA separates cybersecurity risk assessment from probabilistic safety risk management. The security assessment asks whether a vulnerability can be exploited, rather than estimating the likelihood of an incident from historical data. FDA treats the two processes as distinct but connected 36.
  • Security architecture views. At a minimum, FDA recommends a Global System View, a Multi-Patient Harm View, an Updateability/Patchability View and one or more Security Use Case Views. Each should trace architecture elements to security requirements. A view that is omitted should be explained 3511.

Cybersecurity testing

Submissions should include security testing documentation with reports or assessments. This covers fuzz testing, attack surface analysis and closed-box scanning for known vulnerabilities 110111. Penetration test reports should state the testers' independence from the developers, their technical expertise, the scope, duration, methods and findings 112. For all testing, FDA wants the manufacturer's assessment of the findings, including the rationale for any finding that was not fixed or was deferred to a later release. Where a third party performed the testing, the original third-party report should be provided 113114.

Quality system, SPDF and labeling

FDA describes a secure product development framework (SPDF) as a set of lifecycle processes, running from design through decommissioning, that reduce vulnerabilities. It recommends using QMSR/ISO 13485 design processes to implement an SPDF 206207. Security risk management should be integrated into the quality management system alongside safety risk management but kept distinct from it 208. Software validation and risk management may need cybersecurity-specific processes, and quality-system records can support the premarket demonstration 209.

Labeling should tell users about relevant cybersecurity controls and risks. This includes communication interfaces, third-party software, known vulnerabilities, and how to configure and update the device securely 21050.

SBOM expectations

For cyber devices the SBOM is a statutory requirement. For other devices FDA still recommends one 162143. FDA also notes that a robust SBOM covers manufacturer-developed components and upstream dependencies 163. The guidance recommends:

  • Format. A machine-readable SBOM consistent with the NTIA minimum elements ("baseline attributes"), in an industry-accepted format 164165.
  • Support status. For each component, its level of support (actively maintained, no longer maintained or abandoned) and its end-of-support date. This can be in the SBOM itself or in an addendum 166.
  • Known vulnerabilities. All known vulnerabilities in the device and its components, including entries in CISA's Known Exploited Vulnerabilities Catalog. For each, the submission should say how it was discovered and give a safety and security risk assessment of the device and system impacts 51.
  • Omissions. If SBOM information cannot be provided, the submission should justify why 167.

Unsupported and end-of-support software components

FDA expects manufacturers to plan for components losing support rather than assume they will stay secure without maintenance.

  • Premarket disclosure. Support level and end-of-support date should be documented for every SBOM component 166. Known component vulnerabilities and support information belong in the security risk management materials 5152.
  • Lifecycle planning. Submissions should describe plans to update or replace third-party components if support ends. Where supplier support may end before the device's intended support life, FDA suggests considering access to the supplier's source code 5049.
  • Communicating end of support. Users should be told about known or anticipated end of support or end of life for the device and its components 53.
  • Risk transfer. For devices that stay in service after support ends, there should be a pre-established, communicated process for transferring risk. That process should make clear that users' cybersecurity risk can be expected to increase over time. Risk should be transferred only after the relevant risks have been assessed and communicated 5354.

Postmarket management and patch timelines

Manufacturers are expected to monitor for vulnerabilities and threats, reassess risk, mitigate before exploitation where possible, validate remediations and communicate residual risk 146147148149. A coordinated vulnerability disclosure policy should, at a minimum, acknowledge receipt of a vulnerability report to the person who submitted it 150151152.

FDA classifies risk of patient harm as controlled or uncontrolled, based on exploitability and severity after taking existing controls into account 153154. Changes made only to address controlled-risk vulnerabilities, and routine cybersecurity patches, are generally treated as device enhancements rather than recalls 155156. The cybersecurity management plan should set a patch release timeline, and the plan should generally justify the length of the regular update cycle. The guidance does not set a fixed number of days for either regular-cycle or out-of-cycle patches 157158159.

Device modifications

For changes that may affect cybersecurity, a modification submission for a cyber device should include the full required and recommended documentation. For changes unlikely to affect cybersecurity, FDA recommends a narrower package. That package includes a 524B(b)(1) plan if none was filed before, or a reference to the prior submission and a summary of changes to the plan if one was 190. FDA's review generally focuses on changes to cybersecurity controls, but FDA will consider known cybersecurity concerns regardless of the proposed change 191.

What FDA inspections and warning letters have cited

Firm and dateDeviceCybersecurity findingCitation
St. Jude Medical, Form 483 dated February 17, 2017 3940Merlin@home and high-voltage devicesPort-security verification checked only authorized interfaces. Risk assessments did not account for all known security risks and reported vulnerabilities 4142Design verification; risk analysis
Abbott (St. Jude Medical), warning letter, April 12, 2017 175Merlin@home and high-voltage devices 180181No verification that network ports would stay closed to unauthorized interfaces. Risk ratings did not reflect a third-party assessment. A hardcoded universal unlock code was not evaluated as a hazard 18018221 CFR 820.30(f); 820.30(g)
Draeger Medical Systems (posted by FDA as "Draegar"), January 17, 2020 178Multiple product lines (not named in the letter text) 183A complaint identifying cybersecurity vulnerabilities across product lines was cited within a CAPA finding 18421 CFR 820.100(a)
Medtronic, December 9, 2021 169Remote controllers for MiniMed 508 and MiniMed Paradigm insulin pumpsCorrective actions for a vulnerability exploitable by unauthorized individuals did not cover all devices in the field or notify all affected customers 26226321 CFR 820.100(a); FD&C Act 501(h)
Becton, Dickinson / CareFusion 303, November 22, 2024 179Pyxis MedStation ES, ES Anesthesia, MedBank; Pyxis CIISAFE ES 185271544 open software defect tickets, four with cybersecurity vulnerabilities. No process to assess exploitability and severity of patient harm. Security deliverables for corrective work not shown as complete 18618718827221 CFR 820.100(a); 820.30(g)
Contec Medical Systems, October 2, 2025 174B-Ultrasound and Color Doppler Ultrasound systemsA port-shutdown patch was not documented as a deviation from the test plan. Network segmentation, zone boundary and communication restriction testing were missing from the security test report. A CAPA was closed without completing the Cybersecurity Management Plan 26526621 CFR 820.30(g); 820.100(a)
Beta Bionics, January 28, 2026 173iLet ACE PumpLockscreen access flaw could allow insulin delivery to be stopped or inappropriate boluses given. A software fix shipped without the required correction report 26921 CFR 806.10(a), (b)
O.N. Diagnostics, September 1, 2026 172VirtuOst VFA 3.0.0Migration from desktop to web architecture introduced network-exposed attack surfaces and authentication risks. FDA had not reviewed an SBOM or the affected third-party components 27827928021 CFR 807.81(a)(3)(i); FD&C Act 502(o), 510(k) 281282

Patterns across the enforcement record

  • Risk analysis that ignores known vulnerabilities is the most common design-control finding. Three of the warning letters above cite 820.30(g) for cybersecurity risk analysis or security testing gaps: St. Jude/Abbott, BD/CareFusion and Contec 182187265. The St. Jude 483 shows the same theme 42. In the BD letter, FDA specifically faulted the lack of an exploitability and patient-harm severity process 187. That is the same exploitability-based approach the 2026 guidance describes 36.
  • Verification must test negative cases. In both the St. Jude 483 and the Abbott letter, FDA objected that testing confirmed authorized access but not that unauthorized access was blocked 41180.
  • CAPA must reach the whole installed base. The Medtronic, Draeger, BD and Contec letters show that open vulnerability tickets, incomplete field actions and unexecuted cybersecurity management plans are cited as CAPA failures 262184186266.
  • Cybersecurity fixes can trigger reporting obligations. The Beta Bionics letter shows that a security patch for a vulnerability with patient-harm potential can require a Part 806 report 269. This is consistent with FDA's guidance position that only controlled-risk fixes are generally treated as enhancements 155156.
  • Architecture changes can require a new 510(k). The O.N. Diagnostics letter treats a move to a web-based, network-exposed architecture as a change that required a new premarket notification. FDA specifically noted the absence of a reviewed SBOM 280281.

These letters cite the Part 820 section numbers in force at the time (for example 820.30 and 820.100). The 2026 guidance frames its quality-system expectations in terms of the QMSR and ISO 13485 design and development requirements 206209.

Practical implications for submissions and quality systems

  • Decide early whether the device is a cyber device using FDA's broad connectivity reading. Hardware ports used only for servicing still count 258259.
  • Build the SBOM as a controlled, machine-readable record with support status and end-of-support dates. Map it to the known-vulnerability analysis, including CISA's Known Exploited Vulnerabilities Catalog 16416651.
  • Keep security risk management traceable from the threat model through the risk assessment, the SBOM and testing, and score risk on exploitability 3136.
  • Include negative and boundary security tests, such as unauthorized interfaces, segmentation and zone boundaries. These are the gaps FDA cited at St. Jude/Abbott and Contec 180265.
  • Set and justify patch cycles in the postmarket plan 157158. Assess each security fix for controlled versus uncontrolled risk to decide whether Part 806 reporting applies 153155269.
  • Document end-of-support planning and the process for transferring risk to users for components that will outlive supplier support 505354.