For a novel medical device with no legally marketed predicate, the regulatory class decides what route it takes to market, what evidence it needs and how long review is likely to take. Regulatory and clinical teams planning a De Novo request need to understand how FDA assigns devices to Class I, II or III. They also benefit from data on how FDA has recently handled comparable requests, so they can set realistic timelines and anticipate the special controls FDA may impose.

The analysis below sets out the regulatory definitions of the three device classes under 21 CFR 860.3 and explains how the De Novo pathway moves a device out of automatic Class III designation. It then reviews FDA's De Novo grants from January 2025 through November 2026, covering the classes assigned, the review panels the devices came from, the special controls FDA established and the time each request took from receipt to decision.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.

How FDA classifies new medical devices, and what the 2025-2026 De Novo grants show

FDA places every medical device in one of three regulatory classes based on the controls needed to provide reasonable assurance of safety and effectiveness. A truly novel device with no predicate is Class III by default. The De Novo pathway lets FDA move it into Class I or II when general controls, or general plus special controls, are enough. FDA granted 56 De Novo requests between January 2025 and November 2026. Most were Class II devices, and the median time from receipt to decision was about 10.5 months, compared with the 120-day period the regulation describes for FDA's review. The elapsed time includes periods when the review clock was stopped for sponsor responses.

The three device classes under 21 CFR 860.3

21 CFR 860.3 defines a class as "one of the three categories of regulatory control for medical devices" 178. The class reflects the level of control needed, not risk in the abstract.

ClassRegulatory definitionPrimary controls
Class I"General controls are sufficient to provide reasonable assurance of the safety and effectiveness of the device" 180. It also covers devices where information is insufficient but the device is not life-supporting or life-sustaining, not of substantial importance in preventing impairment of health, and presents no potential unreasonable risk 181.General controls only 179
Class II"General controls alone are insufficient to provide reasonable assurance of its safety and effectiveness and there is sufficient information to establish special controls," such as performance standards, postmarket surveillance, patient registries and guidelines 163.General plus special controls
Class IIIInformation is insufficient to show that general or special controls would provide assurance, and the device is life-supporting or life-sustaining, of substantial importance in preventing impairment of human health, or presents a potential unreasonable risk of illness or injury 164.Premarket approval under section 515 164

What FDA weighs in a classification decision

Under 21 CFR 860.7, the Commissioner and classification panels consider the persons for whom the device is intended, its conditions of use (including those in labeling or advertising), probable health benefit weighed against probable injury or illness, and device reliability 165. These same rules govern classification, Class II special controls and Class III premarket approval 166.

FDA relies only on valid scientific evidence, assessed as a whole for the device's conditions of use 167. That includes well-controlled and partially controlled studies, objective trials without matched controls, well-documented case histories and significant human experience with a marketed device. Isolated case reports, random experience and unsubstantiated opinions do not qualify 168. Reasonable assurance of safety means probable benefits outweigh probable risks when the device is used with adequate directions and warnings 169. The manufacturer must make sure adequate evidence exists, and failure to show reasonable assurance under general or special controls may support Class III 170.

Classification panels are advisory committees that recommend classification and reclassification. The Commissioner makes the final determination 171172. For preamendments devices, panels review the evidence against 860.7 and recommend a class with reasons, identified risks and recommended special controls 173174. Implants and life-supporting or life-sustaining devices default to Class III unless the panel and Commissioner find that unnecessary. A Class II recommendation for such a device must explain the special controls needed and how they provide assurance 175.

Why novel devices start in Class III, and how De Novo works

Under FDA's De Novo Classification Process (Evaluation of Automatic Class III Designation) guidance, a new device type that has not been classified is automatically Class III under section 513(f)(1), whatever its actual risk 132. De Novo, under section 513(f)(2), is the route to Class I or II for a device with no legally marketed predicate when general controls, or general and special controls, provide reasonable assurance of safety and effectiveness 131. FDA's benefit-risk guidance describes De Novo devices as low-to-moderate risk 133. The pathway is not available for device types FDA has affirmatively placed in Class III 132.

Two entry routes. A sponsor can file after a 510(k) comes back not substantially equivalent (NSE), or directly ("Direct De Novo") after determining no predicate exists 72134135136. An NSE that rests only on inadequate performance data generally does not open the pathway 134.

Content. 21 CFR 860.220 requires, among other items, a classification summary showing that no legally marketed device of the same type exists, known risks with proposed mitigations, proposed special controls for Class II, a recommended class with justification, study summaries and a benefit-risk discussion, and labeling 77798083.

Acceptance and review. FDA ordinarily decides within 15 days whether to accept a request 86137. Under 860.240, substantive review begins after acceptance, and the regulation calls for a grant or decline order within 120 days of receiving the request, or of receiving the additional information that led to acceptance 101. The guidance states the same 120-day figure 140. An additional-information request stops the review clock, and a complete response restarts it 138139.

Grounds to decline. Under 860.260, FDA may decline a request for reasons that include: the device does not meet Class I or II criteria; the device type is already classified or approved under a PMA; inspection findings; deficient clinical or nonclinical data; or significant unsolicited changes after acceptance 9094959799.

After grant. FDA issues a written order placing the device in Class I or II and identifying the special controls for Class II. The device may be marketed immediately and may then serve as a predicate for future 510(k)s 142. FDA creates a classification regulation for the new device type 143. It also publishes a Federal Register notice of the classification order, including any special controls, within 30 days 104144. The identification, class and requirements are then codified in 21 CFR parts 862 to 892. Non-exempt information in the De Novo file becomes publicly disclosable once the request is granted 105.

De Novo grants, January 2025 to November 2026: headline findings

FDA's De Novo database lists 56 granted requests with decision dates from January 1, 2025 through November 2026: 27 decided in 2025 and 29 in 2026. Review panel and class come from FDA's product classification database for each new product code.

  • Class: 49 product codes are Class II, 3 are Class I (Modius Lean, Sana Device and Avulux), and 4 grants decided from September to November 2026 (Queen of Hearts, Elecsys, Virtuoso and Sonovein) were not yet in the classification database, so their class was not yet posted.
  • Panels: Neurology led with 11 grants. Cardiovascular and General Hospital had 5 each. General and Plastic Surgery, Orthopedic, and Gastroenterology/Urology had 4 each. Microbiology and Radiology had 3 each. Clinical Chemistry, ENT and Pathology had 2 each. Anesthesiology, Dental, Hematology, Immunology, Obstetrics/Gynecology, Ophthalmic and Physical Medicine had 1 each.
  • Time from receipt to decision (calendar days): median 323, mean 316, range 140 to 1,080. The 2025 median was 299 days (n = 27) and the 2026 median was 328 days (n = 29). Only 4 of 56 grants were decided within 150 calendar days. 9 took 151 to 250 days, 36 took 251 to 365 days, and 7 took more than a year.

The day counts below are calendar days from the date FDA received the request to the decision date. They include time the sponsor spent answering additional-information requests, when FDA's review clock is stopped 138139. They therefore measure total elapsed time, not FDA review days, and should not be compared directly with the 120-day regulatory period 101 or with MDUFA performance goals.

Grant-level detail

De NovoDevice (sponsor)Product code / regulationReview panelReceivedDecidedDays
DEN230051 55Myocene (Myocene)SDX / 882.1871Neurology 2122023-07-272025-01-17540
DEN240018 54CORIS System (Nanosonics)SEW / 880.6994General Hospital 2142024-05-012025-03-19322
DEN240020 53Visby Medical Women's Sexual Health Test (Visby Medical)SEA / 866.3386Microbiology 1872024-05-132025-03-28319
DEN240031 51Prapela SVS hospital bassinet pad (Prapela)QVY / 880.5151General Hospital 2132024-06-172025-04-04291
DEN240021 52Tango Belt (Active Protective Technologies)SEC / 890.3780Physical Medicine 1862024-05-162025-04-09328
DEN240064 43CT-132 (Click Therapeutics)SEE / 882.5806Neurology 1822024-11-122025-04-11150
DEN240062 44Minder System (Epi-Minder)SEM / 882.1360Neurology 1942024-10-182025-04-17181
DEN240045 47Teal Wand (Teal Health)SEP / 866.2920Microbiology 1982024-08-272025-05-09255
DEN240035 49ConcizuTrace ELISA (Randox)SES / 864.7298Hematology 2302024-07-012025-05-22325
DEN240048 45Spur Peripheral Retrievable Stent System (Reflow Medical)SEU / 870.5110Cardiovascular 1992024-09-202025-05-29251
DEN240047 46Allix5 (Clairity)SEZ / 892.8500Radiology 2032024-09-062025-05-30266
DEN240032 50APO-Easy Genotyping kit (Firalis)SFC / 866.5850Immunology 2092024-06-212025-06-12356
DEN240066 42COAPTIUM Connect with TISSIUM LIGHT (Tissium)SFD / 882.5270Neurology 2052024-11-212025-06-17208
DEN240038 48Oxiplex (FzioMed)QVL / 888.3047Orthopedic 2332024-07-222025-06-17330
DEN240068 40ArteraAI Prostate (Artera)SFH / 864.3755Pathology 2162024-11-252025-07-31248
DEN250002 32Delphi-MD System (Quantalx Neuroscience)SFN / 882.1860Neurology 2222025-01-032025-08-20229
DEN240067 41KMT2A Breakapart FISH Probe Kit PDx (Cytocell)SFS / 864.1885Pathology 2272024-11-222025-09-19301
DEN240080 33Biolinq Shine Microsensor (Biolinq)SFU / 862.1359Clinical Chemistry 2262024-12-272025-09-23270
DEN250016 22Essilor Stellest (Essilor of America)QUR / 886.5845Ophthalmic 2322025-04-282025-09-25150
DEN220077 56ProSense System (IceCure Medical)QXW / 878.4355General, Plastic Surgery 1932022-10-192025-10-031,080
DEN240073 38PainChek Adult (PainChek)SGB / 882.1472Neurology 1832024-12-112025-10-06299
DEN240076 35Modius Lean (Neurovalens), Class ISFW / 876.5984Gastroenterology, Urology 2082024-12-182025-10-17303
DEN240074 37Pulsenmore ES (Pulsenmore)SGJ / 892.1590Radiology 2152024-12-112025-10-31324
DEN250015 23Neurolyser XR (Fusmobile)SGN / 882.4420Neurology 2212025-04-252025-11-14203
DEN240077 34MuReva OM (MuReva Phototherapy)SGQ / 872.5595Dental 2292024-12-232025-11-26338
DEN250032 17LIAISON XL MUREX Anti-HDV (DiaSorin)SGW / 866.3176Microbiology 2062025-07-302025-12-23146
DEN250004 31QIKCAP System (Hai Solutions)SGX / 880.6512General Hospital 1852025-01-302025-12-23327
DEN250009 27VitaSmart Hypothermic Oxygenated Perfusion System (Bridge to Life)SGY / 876.5881Gastroenterology, Urology 2022025-03-312026-01-15290
DEN250005 30Sana Device (Sana Health), Class IQYN / 890.5775Neurology 1842025-03-032026-01-16319
DEN250006 29Amferia Wound Dressing (Amferia)SHA / 878.4013General, Plastic Surgery 2282025-03-032026-02-06340
DEN250007 28Delivery Date AI (Ultrasound AI)SHE / 892.8200Radiology 2072025-03-172026-02-11331
DEN250012 26LifeVac (LifeVac)QXN / 874.5400Ear, Nose, Throat 2102025-04-112026-03-04327
DEN250014 24Tyto Insights for Eardrum Bulging Detection (Tyto Care)SHL / 874.4775Ear, Nose, Throat 2312025-04-212026-03-17330
DEN240071 39neuropacs (Automated Imaging Diagnostics)SHO / 882.2000Neurology 2042024-12-042026-04-03485
DEN250033 16XplantR Explant Tool (Hjarta Care)SHR / 870.4520Cardiovascular 1882025-08-072026-04-17253
DEN250023 21Aurie System (CathBuddy)SHV / 876.5120Gastroenterology, Urology 2242025-06-062026-04-30328
DEN240075 36Prolystica Pass Thru Disinfectant Wipes (Steris)SHM / 880.6891General Hospital 2202024-12-122026-05-01505
DEN250013 25Modius Spero (Neurovalens)SHX / 882.5807Neurology 2002025-04-162026-05-08387
DEN250028 18DeepView AI System (SpectralMD)SHY / 878.1840General, Plastic Surgery 1972025-06-272026-05-21328
DEN250057 6PCWP Analysis Software (Cardiosense)SIF / 870.1150Cardiovascular 1952025-11-172026-05-22186
DEN250025 20Syncron-E (Autonomous Healthcare)QVC / 868.5896Anesthesiology 2172025-06-232026-06-18360
DEN250034 15Claria System (Claria Medical)SIM / 884.4075Obstetrics/Gynecology 1962025-08-072026-07-02329
DEN250037 14TESSA Spatial Surgery System (Smith & Nephew)SIQ / 888.4561Orthopedic 2252025-08-222026-07-06318
DEN250040 13SportSuite Vision Software (Stryker Endoscopy)SIT / 888.1150Orthopedic 2012025-09-022026-07-17318
DEN250068 3OTTAVA Robotic Surgical System (Auris Health / J&J MedTech)SIR / 878.4966General, Plastic Surgery 2192025-12-182026-07-21215
DEN250042 12CaRi-Heart (Caristo Diagnostics)SIU / 870.2215Cardiovascular 1912025-09-042026-07-28327
DEN250046 9Aletta Autonomous Robotic Phlebotomy Device (Vitestro)SIW / 880.6975General Hospital 1922025-09-262026-08-19327
DEN250043 11Qoramatic Automated Stool Management (CM Technologies)SJB / 876.5230Gastroenterology, Urology 2232025-09-152026-08-20339
DEN250067 4ReneuRx (Neural Ice) (Brixton Biosciences)SJA / 888.3048Orthopedic 2112025-12-152026-08-21249
DEN250026 19Avulux (Matap II), Class IQEO / 882.5811Neurology 1892025-06-232026-08-21424
DEN250049 8Libre Duo 10 Day Continuous Dual Glucose Ketone Monitoring System (Abbott)SJD / 862.1354Clinical Chemistry 1902025-09-292026-08-25330
DEN250044 10Queen of Hearts (ACS AI ECG Model) (Powerful Medical)SHS / not yet listedNot yet in classification database2025-09-222026-09-03346
DEN260007 2SmartRead Medical Device Identifier (BridgeSource Medical)SJJ / 870.3645Cardiovascular 2182026-03-242026-09-29189
DEN260027 1Elecsys HBsAg II quant II RxDx (Roche Diagnostics)SJM / 866.3177 284Not yet in classification database2026-06-022026-10-20140
DEN250054 7Virtuoso Endoscopy System (Virtuoso Surgical)SJS / 878.4967 415Not yet in classification database2025-10-012026-10-30394
DEN250063 5Sonovein System (Theraclion)SJR / not yet listedNot yet in classification database2025-12-082026-11-04331

"Review panel" is the panel FDA assigns to the new product code in its product classification database. For a few codes it differs from the panel of the CFR part the regulation sits in. Sana Device, for example, is assigned to the Neurology panel under a Part 890 (physical medicine) regulation 184. Four grants decided between September and November 2026 (Queen of Hearts, Elecsys, Virtuoso and Sonovein) were not yet in the classification database at the time of review, and FDA had not yet posted their classification orders and decision summaries 291441283414300.

Outliers in elapsed time

The longest elapsed times reflect requests that stayed open across multiple fiscal years. ProSense (cryoablation for low-risk breast cancer) took 1,080 days 56, Myocene took 540 days 55, and Prolystica Pass Thru wipes took 505 days 36. At the other end, Elecsys HBsAg II quant II RxDx was decided in 140 days 1, LIAISON XL MUREX Anti-HDV in 146 days 17, and CT-132 and Essilor Stellest in 150 days each 4322. Because the counts include time on hold, a long elapsed time does not by itself mean FDA review was slow.

Special controls FDA set in these orders

Classification orders and decision summaries were available for 52 of the 56 grants. For 47 of those, the orders yielded a list of special controls that could be summarized. Across those 47, the most common control categories were:

Special control categoryDe Novo orders including it (of 47)
Clinical performance data or validation43
Labeling content (performance summaries, warnings, limitations)41
Software verification, validation and hazard analysis26
Non-clinical/bench performance testing24
Biocompatibility of patient-contacting components21
Human factors/usability testing20
EMC and electrical safety19
Sterility, package integrity and shelf life18
Postmarket surveillance or performance monitoring6

These tallies are approximate. They come from keyword classification of summarized controls, and for a few orders only part of the special-controls list was legible in the posted text (for example Spur 688, Allix5 298 and Oxiplex 289).

The standard Class II template

Many hardware orders use the same set of controls. The Myocene order is typical: non-clinical performance testing, EMC and electrical, thermal and mechanical safety, software verification and validation, biocompatibility, human factors with a use-related risk analysis, and labeling of stimulation output and interpretation of results 670672673674675676. The Neurolyser XR HIFU order follows the same pattern and adds in vivo testing of ablative energy delivery and injury to target and non-target tissues, plus acoustic and targeting characterization 274275.

Software and AI devices: postmarket performance management plans

Several 2026 software orders go beyond premarket validation. For Delivery Date AI, FDA required independent test datasets and prespecified, clinically justified performance goals 812. It also required a postmarket performance management plan that regularly assesses real-world generalizability and performance, detects performance changes and their sources, and communicates current performance to users 816. The PCWP Analysis Software order includes a similar plan 427. Its clinical-performance control calls for 95% confidence intervals, demographic subgroup analysis and reporting of indeterminate-output rates 422. ArteraAI Prostate's controls specify representative training cases, independent validation datasets with documented provenance, and subgroup analyses 357358. PainChek's controls require accuracy and test-retest reliability against a clinical reference standard and labeling that the device is not a stand-alone diagnostic tool 791794.

High-complexity systems: training, distribution limits and reporting

The OTTAVA Robotic Surgical System order has 16 special controls. They include postmarket surveillance unless FDA finds premarket data sufficient 634, device-specific training 636, distribution restricted to facilities that maintain that training 637, and annual reports of cumulative complaints and adverse events until FDA ends the requirement 648. The Aletta autonomous phlebotomy order requires labeling of first-puncture success and the expected rate of referral to manual phlebotomy 527.

Clinical endpoints and postmarket surveillance written into the controls

Some orders set specific clinical expectations. For Essilor Stellest myopia-control lenses, FDA required premarket clinical validation and postmarket surveillance. The order specifies a minimum lower 95% confidence bound of -0.50 D or an alternative clinical justification, and evaluation of rebound after treatment stops 837. For ProSense, the order names 5-year ipsilateral breast tumor recurrence, or another clinically justified endpoint, together with postmarket surveillance 499. Patient labeling must also explain the limits of imaging in confirming complete tumor destruction 510.

IVDs and home-use diagnostics

The IVD orders focus on intended-use limits and analytical and clinical validation. The Visby lay-user STI test order limits intended use to specimens suitable for self-collection and testing. It excludes uses such as aiding diagnosis of vector-borne illness or viral hepatitis, and it requires a prospective study in the home-use setting 407408410. The Teal Wand order restricts the device to specimen types FDA finds appropriate for lay collection and bars use with tests where lay-collected specimens could cause unacceptable harm 257258. The KMT2A FISH probe kit is classified as a revumenib eligibility detection system 434. Its validation controls cover cutoffs, reproducibility across readers, lots and sites, and analytical sensitivity and specificity 437. For both Biolinq Shine and Libre Duo, the controls require interference testing, measures that prevent sensor use beyond the claimed wear period, and usability studies 860861862345346347.

Class I De Novos

Three grants were Class I under general controls only, and all three are exempt from 510(k), subject to the limitations on exemptions. They are Modius Lean, a cranial electrotherapy stimulator for weight management 248249; Sana Device, an audio-visual stimulator for pain relief 610; and Avulux, light-attenuation lenses for migraine-associated light sensitivity 398397. Nearly all the Class II orders reviewed state that the device type is not exempt from premarket notification, for example CORIS 806 and Libre Duo 349. Later devices of these types can therefore come to market through 510(k)s, and the De Novo device may serve as a predicate 142.

Practical takeaways for De Novo sponsors

  • Plan for the full elapsed time, not the regulatory clock. The regulation describes a 120-day decision period 101, but the median elapsed time for 2025-2026 grants was 323 calendar days, including time spent answering additional-information requests 138139.
  • Propose special controls that match FDA's current templates. 21 CFR 860.220 requires the request to include proposed special controls and risk mitigations 79. Recent orders show recurring categories, each tied to the risks of the device type: clinical performance, labeling, software, biocompatibility, EMC, sterility and human factors controls.
  • Recent AI-enabled software orders required postmarket performance monitoring. The Delivery Date AI and PCWP Analysis Software orders both require a postmarket performance management plan 816427.
  • Your special controls become the bar for followers. The granting order creates the classification regulation, and the device can serve as a 510(k) predicate 142143. Later entrants will be held to the controls FDA writes into the order.
Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.