For a novel medical device with no legally marketed predicate, the regulatory class decides what route it takes to market, what evidence it needs and how long review is likely to take. Regulatory and clinical teams planning a De Novo request need to understand how FDA assigns devices to Class I, II or III. They also benefit from data on how FDA has recently handled comparable requests, so they can set realistic timelines and anticipate the special controls FDA may impose.
The analysis below sets out the regulatory definitions of the three device classes under 21 CFR 860.3 and explains how the De Novo pathway moves a device out of automatic Class III designation. It then reviews FDA's De Novo grants from January 2025 through November 2026, covering the classes assigned, the review panels the devices came from, the special controls FDA established and the time each request took from receipt to decision.
How FDA classifies new medical devices, and what the 2025-2026 De Novo grants show
FDA places every medical device in one of three regulatory classes based on the controls needed to provide reasonable assurance of safety and effectiveness. A truly novel device with no predicate is Class III by default. The De Novo pathway lets FDA move it into Class I or II when general controls, or general plus special controls, are enough. FDA granted 56 De Novo requests between January 2025 and November 2026. Most were Class II devices, and the median time from receipt to decision was about 10.5 months, compared with the 120-day period the regulation describes for FDA's review. The elapsed time includes periods when the review clock was stopped for sponsor responses.
The three device classes under 21 CFR 860.3
21 CFR 860.3 defines a class as "one of the three categories of regulatory control for medical devices" 178. The class reflects the level of control needed, not risk in the abstract.
| Class | Regulatory definition | Primary controls |
|---|---|---|
| Class I | "General controls are sufficient to provide reasonable assurance of the safety and effectiveness of the device" 180. It also covers devices where information is insufficient but the device is not life-supporting or life-sustaining, not of substantial importance in preventing impairment of health, and presents no potential unreasonable risk 181. | General controls only 179 |
| Class II | "General controls alone are insufficient to provide reasonable assurance of its safety and effectiveness and there is sufficient information to establish special controls," such as performance standards, postmarket surveillance, patient registries and guidelines 163. | General plus special controls |
| Class III | Information is insufficient to show that general or special controls would provide assurance, and the device is life-supporting or life-sustaining, of substantial importance in preventing impairment of human health, or presents a potential unreasonable risk of illness or injury 164. | Premarket approval under section 515 164 |
What FDA weighs in a classification decision
Under 21 CFR 860.7, the Commissioner and classification panels consider the persons for whom the device is intended, its conditions of use (including those in labeling or advertising), probable health benefit weighed against probable injury or illness, and device reliability 165. These same rules govern classification, Class II special controls and Class III premarket approval 166.
FDA relies only on valid scientific evidence, assessed as a whole for the device's conditions of use 167. That includes well-controlled and partially controlled studies, objective trials without matched controls, well-documented case histories and significant human experience with a marketed device. Isolated case reports, random experience and unsubstantiated opinions do not qualify 168. Reasonable assurance of safety means probable benefits outweigh probable risks when the device is used with adequate directions and warnings 169. The manufacturer must make sure adequate evidence exists, and failure to show reasonable assurance under general or special controls may support Class III 170.
Classification panels are advisory committees that recommend classification and reclassification. The Commissioner makes the final determination 171172. For preamendments devices, panels review the evidence against 860.7 and recommend a class with reasons, identified risks and recommended special controls 173174. Implants and life-supporting or life-sustaining devices default to Class III unless the panel and Commissioner find that unnecessary. A Class II recommendation for such a device must explain the special controls needed and how they provide assurance 175.
Why novel devices start in Class III, and how De Novo works
Under FDA's De Novo Classification Process (Evaluation of Automatic Class III Designation) guidance, a new device type that has not been classified is automatically Class III under section 513(f)(1), whatever its actual risk 132. De Novo, under section 513(f)(2), is the route to Class I or II for a device with no legally marketed predicate when general controls, or general and special controls, provide reasonable assurance of safety and effectiveness 131. FDA's benefit-risk guidance describes De Novo devices as low-to-moderate risk 133. The pathway is not available for device types FDA has affirmatively placed in Class III 132.
Two entry routes. A sponsor can file after a 510(k) comes back not substantially equivalent (NSE), or directly ("Direct De Novo") after determining no predicate exists 72134135136. An NSE that rests only on inadequate performance data generally does not open the pathway 134.
Content. 21 CFR 860.220 requires, among other items, a classification summary showing that no legally marketed device of the same type exists, known risks with proposed mitigations, proposed special controls for Class II, a recommended class with justification, study summaries and a benefit-risk discussion, and labeling 77798083.
Acceptance and review. FDA ordinarily decides within 15 days whether to accept a request 86137. Under 860.240, substantive review begins after acceptance, and the regulation calls for a grant or decline order within 120 days of receiving the request, or of receiving the additional information that led to acceptance 101. The guidance states the same 120-day figure 140. An additional-information request stops the review clock, and a complete response restarts it 138139.
Grounds to decline. Under 860.260, FDA may decline a request for reasons that include: the device does not meet Class I or II criteria; the device type is already classified or approved under a PMA; inspection findings; deficient clinical or nonclinical data; or significant unsolicited changes after acceptance 9094959799.
After grant. FDA issues a written order placing the device in Class I or II and identifying the special controls for Class II. The device may be marketed immediately and may then serve as a predicate for future 510(k)s 142. FDA creates a classification regulation for the new device type 143. It also publishes a Federal Register notice of the classification order, including any special controls, within 30 days 104144. The identification, class and requirements are then codified in 21 CFR parts 862 to 892. Non-exempt information in the De Novo file becomes publicly disclosable once the request is granted 105.
De Novo grants, January 2025 to November 2026: headline findings
FDA's De Novo database lists 56 granted requests with decision dates from January 1, 2025 through November 2026: 27 decided in 2025 and 29 in 2026. Review panel and class come from FDA's product classification database for each new product code.
- Class: 49 product codes are Class II, 3 are Class I (Modius Lean, Sana Device and Avulux), and 4 grants decided from September to November 2026 (Queen of Hearts, Elecsys, Virtuoso and Sonovein) were not yet in the classification database, so their class was not yet posted.
- Panels: Neurology led with 11 grants. Cardiovascular and General Hospital had 5 each. General and Plastic Surgery, Orthopedic, and Gastroenterology/Urology had 4 each. Microbiology and Radiology had 3 each. Clinical Chemistry, ENT and Pathology had 2 each. Anesthesiology, Dental, Hematology, Immunology, Obstetrics/Gynecology, Ophthalmic and Physical Medicine had 1 each.
- Time from receipt to decision (calendar days): median 323, mean 316, range 140 to 1,080. The 2025 median was 299 days (n = 27) and the 2026 median was 328 days (n = 29). Only 4 of 56 grants were decided within 150 calendar days. 9 took 151 to 250 days, 36 took 251 to 365 days, and 7 took more than a year.
The day counts below are calendar days from the date FDA received the request to the decision date. They include time the sponsor spent answering additional-information requests, when FDA's review clock is stopped 138139. They therefore measure total elapsed time, not FDA review days, and should not be compared directly with the 120-day regulatory period 101 or with MDUFA performance goals.
Grant-level detail
| De Novo | Device (sponsor) | Product code / regulation | Review panel | Received | Decided | Days |
|---|---|---|---|---|---|---|
| DEN230051 55 | Myocene (Myocene) | SDX / 882.1871 | Neurology 212 | 2023-07-27 | 2025-01-17 | 540 |
| DEN240018 54 | CORIS System (Nanosonics) | SEW / 880.6994 | General Hospital 214 | 2024-05-01 | 2025-03-19 | 322 |
| DEN240020 53 | Visby Medical Women's Sexual Health Test (Visby Medical) | SEA / 866.3386 | Microbiology 187 | 2024-05-13 | 2025-03-28 | 319 |
| DEN240031 51 | Prapela SVS hospital bassinet pad (Prapela) | QVY / 880.5151 | General Hospital 213 | 2024-06-17 | 2025-04-04 | 291 |
| DEN240021 52 | Tango Belt (Active Protective Technologies) | SEC / 890.3780 | Physical Medicine 186 | 2024-05-16 | 2025-04-09 | 328 |
| DEN240064 43 | CT-132 (Click Therapeutics) | SEE / 882.5806 | Neurology 182 | 2024-11-12 | 2025-04-11 | 150 |
| DEN240062 44 | Minder System (Epi-Minder) | SEM / 882.1360 | Neurology 194 | 2024-10-18 | 2025-04-17 | 181 |
| DEN240045 47 | Teal Wand (Teal Health) | SEP / 866.2920 | Microbiology 198 | 2024-08-27 | 2025-05-09 | 255 |
| DEN240035 49 | ConcizuTrace ELISA (Randox) | SES / 864.7298 | Hematology 230 | 2024-07-01 | 2025-05-22 | 325 |
| DEN240048 45 | Spur Peripheral Retrievable Stent System (Reflow Medical) | SEU / 870.5110 | Cardiovascular 199 | 2024-09-20 | 2025-05-29 | 251 |
| DEN240047 46 | Allix5 (Clairity) | SEZ / 892.8500 | Radiology 203 | 2024-09-06 | 2025-05-30 | 266 |
| DEN240032 50 | APO-Easy Genotyping kit (Firalis) | SFC / 866.5850 | Immunology 209 | 2024-06-21 | 2025-06-12 | 356 |
| DEN240066 42 | COAPTIUM Connect with TISSIUM LIGHT (Tissium) | SFD / 882.5270 | Neurology 205 | 2024-11-21 | 2025-06-17 | 208 |
| DEN240038 48 | Oxiplex (FzioMed) | QVL / 888.3047 | Orthopedic 233 | 2024-07-22 | 2025-06-17 | 330 |
| DEN240068 40 | ArteraAI Prostate (Artera) | SFH / 864.3755 | Pathology 216 | 2024-11-25 | 2025-07-31 | 248 |
| DEN250002 32 | Delphi-MD System (Quantalx Neuroscience) | SFN / 882.1860 | Neurology 222 | 2025-01-03 | 2025-08-20 | 229 |
| DEN240067 41 | KMT2A Breakapart FISH Probe Kit PDx (Cytocell) | SFS / 864.1885 | Pathology 227 | 2024-11-22 | 2025-09-19 | 301 |
| DEN240080 33 | Biolinq Shine Microsensor (Biolinq) | SFU / 862.1359 | Clinical Chemistry 226 | 2024-12-27 | 2025-09-23 | 270 |
| DEN250016 22 | Essilor Stellest (Essilor of America) | QUR / 886.5845 | Ophthalmic 232 | 2025-04-28 | 2025-09-25 | 150 |
| DEN220077 56 | ProSense System (IceCure Medical) | QXW / 878.4355 | General, Plastic Surgery 193 | 2022-10-19 | 2025-10-03 | 1,080 |
| DEN240073 38 | PainChek Adult (PainChek) | SGB / 882.1472 | Neurology 183 | 2024-12-11 | 2025-10-06 | 299 |
| DEN240076 35 | Modius Lean (Neurovalens), Class I | SFW / 876.5984 | Gastroenterology, Urology 208 | 2024-12-18 | 2025-10-17 | 303 |
| DEN240074 37 | Pulsenmore ES (Pulsenmore) | SGJ / 892.1590 | Radiology 215 | 2024-12-11 | 2025-10-31 | 324 |
| DEN250015 23 | Neurolyser XR (Fusmobile) | SGN / 882.4420 | Neurology 221 | 2025-04-25 | 2025-11-14 | 203 |
| DEN240077 34 | MuReva OM (MuReva Phototherapy) | SGQ / 872.5595 | Dental 229 | 2024-12-23 | 2025-11-26 | 338 |
| DEN250032 17 | LIAISON XL MUREX Anti-HDV (DiaSorin) | SGW / 866.3176 | Microbiology 206 | 2025-07-30 | 2025-12-23 | 146 |
| DEN250004 31 | QIKCAP System (Hai Solutions) | SGX / 880.6512 | General Hospital 185 | 2025-01-30 | 2025-12-23 | 327 |
| DEN250009 27 | VitaSmart Hypothermic Oxygenated Perfusion System (Bridge to Life) | SGY / 876.5881 | Gastroenterology, Urology 202 | 2025-03-31 | 2026-01-15 | 290 |
| DEN250005 30 | Sana Device (Sana Health), Class I | QYN / 890.5775 | Neurology 184 | 2025-03-03 | 2026-01-16 | 319 |
| DEN250006 29 | Amferia Wound Dressing (Amferia) | SHA / 878.4013 | General, Plastic Surgery 228 | 2025-03-03 | 2026-02-06 | 340 |
| DEN250007 28 | Delivery Date AI (Ultrasound AI) | SHE / 892.8200 | Radiology 207 | 2025-03-17 | 2026-02-11 | 331 |
| DEN250012 26 | LifeVac (LifeVac) | QXN / 874.5400 | Ear, Nose, Throat 210 | 2025-04-11 | 2026-03-04 | 327 |
| DEN250014 24 | Tyto Insights for Eardrum Bulging Detection (Tyto Care) | SHL / 874.4775 | Ear, Nose, Throat 231 | 2025-04-21 | 2026-03-17 | 330 |
| DEN240071 39 | neuropacs (Automated Imaging Diagnostics) | SHO / 882.2000 | Neurology 204 | 2024-12-04 | 2026-04-03 | 485 |
| DEN250033 16 | XplantR Explant Tool (Hjarta Care) | SHR / 870.4520 | Cardiovascular 188 | 2025-08-07 | 2026-04-17 | 253 |
| DEN250023 21 | Aurie System (CathBuddy) | SHV / 876.5120 | Gastroenterology, Urology 224 | 2025-06-06 | 2026-04-30 | 328 |
| DEN240075 36 | Prolystica Pass Thru Disinfectant Wipes (Steris) | SHM / 880.6891 | General Hospital 220 | 2024-12-12 | 2026-05-01 | 505 |
| DEN250013 25 | Modius Spero (Neurovalens) | SHX / 882.5807 | Neurology 200 | 2025-04-16 | 2026-05-08 | 387 |
| DEN250028 18 | DeepView AI System (SpectralMD) | SHY / 878.1840 | General, Plastic Surgery 197 | 2025-06-27 | 2026-05-21 | 328 |
| DEN250057 6 | PCWP Analysis Software (Cardiosense) | SIF / 870.1150 | Cardiovascular 195 | 2025-11-17 | 2026-05-22 | 186 |
| DEN250025 20 | Syncron-E (Autonomous Healthcare) | QVC / 868.5896 | Anesthesiology 217 | 2025-06-23 | 2026-06-18 | 360 |
| DEN250034 15 | Claria System (Claria Medical) | SIM / 884.4075 | Obstetrics/Gynecology 196 | 2025-08-07 | 2026-07-02 | 329 |
| DEN250037 14 | TESSA Spatial Surgery System (Smith & Nephew) | SIQ / 888.4561 | Orthopedic 225 | 2025-08-22 | 2026-07-06 | 318 |
| DEN250040 13 | SportSuite Vision Software (Stryker Endoscopy) | SIT / 888.1150 | Orthopedic 201 | 2025-09-02 | 2026-07-17 | 318 |
| DEN250068 3 | OTTAVA Robotic Surgical System (Auris Health / J&J MedTech) | SIR / 878.4966 | General, Plastic Surgery 219 | 2025-12-18 | 2026-07-21 | 215 |
| DEN250042 12 | CaRi-Heart (Caristo Diagnostics) | SIU / 870.2215 | Cardiovascular 191 | 2025-09-04 | 2026-07-28 | 327 |
| DEN250046 9 | Aletta Autonomous Robotic Phlebotomy Device (Vitestro) | SIW / 880.6975 | General Hospital 192 | 2025-09-26 | 2026-08-19 | 327 |
| DEN250043 11 | Qoramatic Automated Stool Management (CM Technologies) | SJB / 876.5230 | Gastroenterology, Urology 223 | 2025-09-15 | 2026-08-20 | 339 |
| DEN250067 4 | ReneuRx (Neural Ice) (Brixton Biosciences) | SJA / 888.3048 | Orthopedic 211 | 2025-12-15 | 2026-08-21 | 249 |
| DEN250026 19 | Avulux (Matap II), Class I | QEO / 882.5811 | Neurology 189 | 2025-06-23 | 2026-08-21 | 424 |
| DEN250049 8 | Libre Duo 10 Day Continuous Dual Glucose Ketone Monitoring System (Abbott) | SJD / 862.1354 | Clinical Chemistry 190 | 2025-09-29 | 2026-08-25 | 330 |
| DEN250044 10 | Queen of Hearts (ACS AI ECG Model) (Powerful Medical) | SHS / not yet listed | Not yet in classification database | 2025-09-22 | 2026-09-03 | 346 |
| DEN260007 2 | SmartRead Medical Device Identifier (BridgeSource Medical) | SJJ / 870.3645 | Cardiovascular 218 | 2026-03-24 | 2026-09-29 | 189 |
| DEN260027 1 | Elecsys HBsAg II quant II RxDx (Roche Diagnostics) | SJM / 866.3177 284 | Not yet in classification database | 2026-06-02 | 2026-10-20 | 140 |
| DEN250054 7 | Virtuoso Endoscopy System (Virtuoso Surgical) | SJS / 878.4967 415 | Not yet in classification database | 2025-10-01 | 2026-10-30 | 394 |
| DEN250063 5 | Sonovein System (Theraclion) | SJR / not yet listed | Not yet in classification database | 2025-12-08 | 2026-11-04 | 331 |
"Review panel" is the panel FDA assigns to the new product code in its product classification database. For a few codes it differs from the panel of the CFR part the regulation sits in. Sana Device, for example, is assigned to the Neurology panel under a Part 890 (physical medicine) regulation 184. Four grants decided between September and November 2026 (Queen of Hearts, Elecsys, Virtuoso and Sonovein) were not yet in the classification database at the time of review, and FDA had not yet posted their classification orders and decision summaries 291441283414300.
Outliers in elapsed time
The longest elapsed times reflect requests that stayed open across multiple fiscal years. ProSense (cryoablation for low-risk breast cancer) took 1,080 days 56, Myocene took 540 days 55, and Prolystica Pass Thru wipes took 505 days 36. At the other end, Elecsys HBsAg II quant II RxDx was decided in 140 days 1, LIAISON XL MUREX Anti-HDV in 146 days 17, and CT-132 and Essilor Stellest in 150 days each 4322. Because the counts include time on hold, a long elapsed time does not by itself mean FDA review was slow.
Special controls FDA set in these orders
Classification orders and decision summaries were available for 52 of the 56 grants. For 47 of those, the orders yielded a list of special controls that could be summarized. Across those 47, the most common control categories were:
| Special control category | De Novo orders including it (of 47) |
|---|---|
| Clinical performance data or validation | 43 |
| Labeling content (performance summaries, warnings, limitations) | 41 |
| Software verification, validation and hazard analysis | 26 |
| Non-clinical/bench performance testing | 24 |
| Biocompatibility of patient-contacting components | 21 |
| Human factors/usability testing | 20 |
| EMC and electrical safety | 19 |
| Sterility, package integrity and shelf life | 18 |
| Postmarket surveillance or performance monitoring | 6 |
These tallies are approximate. They come from keyword classification of summarized controls, and for a few orders only part of the special-controls list was legible in the posted text (for example Spur 688, Allix5 298 and Oxiplex 289).
The standard Class II template
Many hardware orders use the same set of controls. The Myocene order is typical: non-clinical performance testing, EMC and electrical, thermal and mechanical safety, software verification and validation, biocompatibility, human factors with a use-related risk analysis, and labeling of stimulation output and interpretation of results 670672673674675676. The Neurolyser XR HIFU order follows the same pattern and adds in vivo testing of ablative energy delivery and injury to target and non-target tissues, plus acoustic and targeting characterization 274275.
Software and AI devices: postmarket performance management plans
Several 2026 software orders go beyond premarket validation. For Delivery Date AI, FDA required independent test datasets and prespecified, clinically justified performance goals 812. It also required a postmarket performance management plan that regularly assesses real-world generalizability and performance, detects performance changes and their sources, and communicates current performance to users 816. The PCWP Analysis Software order includes a similar plan 427. Its clinical-performance control calls for 95% confidence intervals, demographic subgroup analysis and reporting of indeterminate-output rates 422. ArteraAI Prostate's controls specify representative training cases, independent validation datasets with documented provenance, and subgroup analyses 357358. PainChek's controls require accuracy and test-retest reliability against a clinical reference standard and labeling that the device is not a stand-alone diagnostic tool 791794.
High-complexity systems: training, distribution limits and reporting
The OTTAVA Robotic Surgical System order has 16 special controls. They include postmarket surveillance unless FDA finds premarket data sufficient 634, device-specific training 636, distribution restricted to facilities that maintain that training 637, and annual reports of cumulative complaints and adverse events until FDA ends the requirement 648. The Aletta autonomous phlebotomy order requires labeling of first-puncture success and the expected rate of referral to manual phlebotomy 527.
Clinical endpoints and postmarket surveillance written into the controls
Some orders set specific clinical expectations. For Essilor Stellest myopia-control lenses, FDA required premarket clinical validation and postmarket surveillance. The order specifies a minimum lower 95% confidence bound of -0.50 D or an alternative clinical justification, and evaluation of rebound after treatment stops 837. For ProSense, the order names 5-year ipsilateral breast tumor recurrence, or another clinically justified endpoint, together with postmarket surveillance 499. Patient labeling must also explain the limits of imaging in confirming complete tumor destruction 510.
IVDs and home-use diagnostics
The IVD orders focus on intended-use limits and analytical and clinical validation. The Visby lay-user STI test order limits intended use to specimens suitable for self-collection and testing. It excludes uses such as aiding diagnosis of vector-borne illness or viral hepatitis, and it requires a prospective study in the home-use setting 407408410. The Teal Wand order restricts the device to specimen types FDA finds appropriate for lay collection and bars use with tests where lay-collected specimens could cause unacceptable harm 257258. The KMT2A FISH probe kit is classified as a revumenib eligibility detection system 434. Its validation controls cover cutoffs, reproducibility across readers, lots and sites, and analytical sensitivity and specificity 437. For both Biolinq Shine and Libre Duo, the controls require interference testing, measures that prevent sensor use beyond the claimed wear period, and usability studies 860861862345346347.
Class I De Novos
Three grants were Class I under general controls only, and all three are exempt from 510(k), subject to the limitations on exemptions. They are Modius Lean, a cranial electrotherapy stimulator for weight management 248249; Sana Device, an audio-visual stimulator for pain relief 610; and Avulux, light-attenuation lenses for migraine-associated light sensitivity 398397. Nearly all the Class II orders reviewed state that the device type is not exempt from premarket notification, for example CORIS 806 and Libre Duo 349. Later devices of these types can therefore come to market through 510(k)s, and the De Novo device may serve as a predicate 142.
Practical takeaways for De Novo sponsors
- Plan for the full elapsed time, not the regulatory clock. The regulation describes a 120-day decision period 101, but the median elapsed time for 2025-2026 grants was 323 calendar days, including time spent answering additional-information requests 138139.
- Propose special controls that match FDA's current templates. 21 CFR 860.220 requires the request to include proposed special controls and risk mitigations 79. Recent orders show recurring categories, each tied to the risks of the device type: clinical performance, labeling, software, biocompatibility, EMC, sterility and human factors controls.
- Recent AI-enabled software orders required postmarket performance monitoring. The Delivery Date AI and PCWP Analysis Software orders both require a postmarket performance management plan 816427.
- Your special controls become the bar for followers. The granting order creates the classification regulation, and the device can serve as a 510(k) predicate 142143. Later entrants will be held to the controls FDA writes into the order.