FDA-Authorized Non-Invasive Neuromodulation Devices for Migraine and Cluster Headache: Pathways, Indications and Clinical Evidence

Chetan Mishra
Chetan Mishra
Nov 1, 2026

Non-invasive neuromodulation has become an established device category for primary headache disorders, and the regulatory record behind it is a useful reference for sponsors of new stimulation technologies. Knowing how earlier devices reached the market, which classification regulations and special controls apply, and how much clinical evidence FDA accepted helps regulatory and clinical teams choose a pathway, pick predicates and design pivotal studies.

The analysis below covers the FDA-authorized device types for migraine and cluster headache, including transcutaneous supraorbital stimulation, single-pulse transcranial magnetic stimulation, non-invasive vagus nerve stimulation, remote electrical neuromodulation and thermal vestibular stimulation. For each one it traces the founding De Novo classification and the later 510(k) clearances, the cleared indications and patient populations, prescription versus over-the-counter status, and the design and outcomes of the trials FDA relied on.

Rhizome did this research. Let it help you with your most critical decisions by creating a free account.

FDA-authorized non-invasive neuromodulation devices for migraine and cluster headache: pathways, indications and clinical evidence

FDA has authorized five distinct types of non-invasive neuromodulation device for primary headache disorders. Each type entered the market through its own De Novo classification, which created a Class II regulation with special controls. Later 510(k) clearances then broadened indications, added new populations, moved some devices to over-the-counter (OTC) use and brought in competing devices. The five modalities are:

  • transcutaneous supraorbital (trigeminal) electrical stimulation
  • single-pulse transcranial magnetic stimulation (sTMS)
  • non-invasive vagus nerve stimulation (nVNS)
  • remote electrical neuromodulation of the upper arm (REN)
  • thermal vestibular stimulation

Key takeaways

  • Five De Novo classifications anchor the field. These are Cefaly (DEN120019), Cerena TMS (DEN130022), gammaCore (DEN150048), the ThermoNeuroModulation (TNM) Device (DEN170023) and Nerivio Migra (DEN180059). Each was classified into Class II with device-specific special controls 357380268331307.
  • The De Novo evidence came from randomized, sham-controlled trials, and the results were not always clean. FDA granted Cefaly even though one primary comparison (reduction in migraine days) was not statistically significant 363. The Cerena effectiveness result came from a post-hoc analysis of subjects who treated an attack while in pain 381. The gammaCore pivotal ACT1 study missed its overall endpoint but showed benefit in episodic cluster headache 274.
  • Most indication expansions went through the 510(k) pathway. For gammaCore, these included acute migraine, migraine prevention, adjunctive cluster prevention, adolescents and hemicrania syndromes 473453668569. For Nerivio, they included chronic migraine, adolescents, prevention and children aged 8 and older 705778681754. For SpringTMS, they included prevention and adolescents 731743. Some of these expansions had new clinical data behind them. Others relied on extrapolation or real-world data.
  • The documented Cefaly Rx-to-OTC switch relied on human factors data, not a new clinical trial. Cefaly Dual's OTC clearance (K201895) stated that no clinical performance testing was done and that a human-factors validation study supported the switch 499500.

Regulatory framework: the five headache neuromodulation classifications

Device type (generic name)RegulationProduct codeFounding De NovoDecision dateOriginal indication
Transcutaneous electrical nerve stimulator to treat headache21 CFR 882.5891 357PCCCefaly, STX-Med (DEN120019)2014-03-11 245Prophylactic treatment of episodic migraine, age 18 and older 356
Transcranial magnetic stimulator for headache21 CFR 882.5808 380OKPCerena TMS, eNeura (DEN130022)2013-12-13 9Acute treatment of pain associated with migraine with aura 379
External vagal nerve stimulator for headache21 CFR 882.5892 (corrected; see note) 269PKRgammaCore, electroCore (DEN150048)2017-04-14 246Acute treatment of pain associated with episodic cluster headache in adults 265
Thermal vestibular stimulator for headache21 CFR 882.5893 331QARTNM Device, Scion Neurostim (DEN170023)2018-03-26 247Prophylactic treatment of episodic migraine, age 12 and older 330
Trunk and limb electrical stimulator to treat headache21 CFR 882.5899 307QGTNerivio Migra, Theranica (DEN180059)2019-05-20 242Acute treatment of migraine with or without aura, age 18 and older, without chronic migraine 306

Note on the gammaCore regulation citation. The gammaCore De Novo decision summary lists 21 CFR 882.8592 268. A later FDA letter states that it corrects the April 14, 2017 classification order and gives 21 CFR 882.5892 269. Use the corrected number when citing the classification.

The special controls follow a common pattern across all five types:

  • biocompatibility of patient-contacting components
  • electromagnetic compatibility (EMC) and electrical, mechanical and thermal safety testing
  • software verification, validation and hazard analysis
  • characterization of device output
  • labeling requirements

Each type also has its own modality-specific controls. The TMS controls require magnetic-output, field-map and sound-level characterization 386. The thermal vestibular controls require waveform and temperature limits and validation of earpiece cleaning 338. The REN controls require impedance monitoring and electrode performance testing 315. The Cefaly and Cerena controls also call for clinical performance data 346386.

Transcutaneous supraorbital stimulation (product code PCC)

Cefaly De Novo (DEN120019)

Cefaly is a TENS device worn on the forehead. Each session lasts 20 minutes, and intensity ramps up over the first 14 minutes 354355. FDA granted it for prophylactic treatment of episodic migraine in adults, as a prescription device that is contraindicated in patients with pacemakers or implanted or wearable defibrillators 356.

Evidence. The pivotal PREMICE study was a prospective, multicenter, double-blind, randomized, sham-controlled trial run at five Belgian headache clinics. A one-month run-in was followed by a 90-day treatment period 358359. The trial randomized 67 subjects (34 active, 33 sham), and 59 completed per protocol 360361. There were two primary outcomes: change in monthly migraine days, and the proportion of patients with at least a 50% reduction in migraine days 362. FDA's benefit-risk discussion says that the responder finding supported benefit. It also says the between-group comparison for reduction in migraine days was not statistically significant 363.

There were no serious adverse events in the trial 364. In a European post-market survey of 2,313 users, 4.3% reported complaints, and all were non-serious and reversible 365. FDA concluded that probable benefits outweighed risks while acknowledging the mixed primary results 367368.

510(k) progression for Cefaly

510(k)DeviceChangeClinical evidence in the 510(k) summary
K160237CefalyChanges to power supply, dimensions and electrode fixation; intended use unchanged 842None summarized
K171446Cefaly AcuteNew indication: acute treatment of migraine with or without aura, age 18 and older 540. Uses 100 Hz/60-minute sessions instead of 60 Hz/20 minutes 544Not provided in the summary 539
K173006Cefaly DualCombines the acute and prophylactic programs in one device 807810Not provided 806
K201895Cefaly Dual (OTC)Switch from Rx to OTC 496498No clinical testing. Human-factors validation supported the switch 499500
K212071Cefaly Dual Connected / Enhanced with RFIDBluetooth app, RFID electrode detection and docking station. OTC labeling uses "preventative" in place of "prophylactic" 556557Indications and population unchanged 558
K234029Cefaly ConnectedApp-based two-way control of treatment 421None summarized 422

Relivion and other PCC devices

Neurolief's Relivion (K203419) was cleared for acute treatment of migraine in adults, with Cefaly Acute as the predicate. Unlike Cefaly, it stimulates at both the forehead and the occiput 633635636. This is one of the few PCC 510(k)s with substantial new clinical data:

  • RIME study. A sham-controlled trial in 131 patients. Pain relief at 2 hours was 60% with active treatment versus 37% with sham in the modified ITT population (p=0.0180) 643644.
  • SP-301 study. A randomized, double-blind, placebo-controlled study in 55 patients. Pain-intensity improvement favored active treatment at 1, 2 and 24 hours 639640.
  • OS-TNS study. Thirty adults. Mean VAS pain change was −79.2% with active treatment versus +14.9% with sham (P=0.0002) 638.

A later Relivion 510(k) (K212106) added a remote physician interface and included no new clinical study 722723.

Several other 510(k) devices use Cefaly as the predicate and include no new clinical data:

  • ALLIVE (K192773, prevention, Rx) relied on a clinical evaluation report based on the predicate's clinical data 796802.
  • HeadaTerm (K172450, prevention, Rx) listed clinical testing as "Not Applicable" 616622.
  • HeadaTerm 2 (K230782) was cleared OTC on the basis of a 24-participant usability verification 868872.
  • Elexir (K211380) relied on the predicate's clinical data 594599. Elexir 2.0 (K242719) is a prescription device for acute and preventive treatment 831832.
  • Milieve (K233751, OTC) 581582 and DOOPANG (K232749, OTC) 518519 cite Cefaly Dual OTC as the predicate.
  • Migraine TENS Digital Pain Reliever (K210364, OTC) also cites Cefaly Dual OTC 858860.
  • Enso for Migraine (Hinge Health, K254216, cleared 2026-04-16) is an OTC device for acute and preventive treatment. No clinical testing was done. Effectiveness was supported by published literature on the predicate, and a human factors study supported OTC use 3651656657.

Single-pulse transcranial magnetic stimulation (product code OKP)

Cerena De Novo (DEN130022)

Cerena delivers two brief magnetic pulses. Its device type is defined as directing pulsed magnetic fields at discrete brain regions to induce electrical currents 377378. The original indication was limited to acute treatment of pain associated with migraine with aura 379.

Evidence. The trial enrolled 201 subjects, randomized 1:1 to device or sham. The effectiveness analysis used 113 subjects who treated an attack while in pain (53 active, 60 sham). FDA's summary describes this as a post-hoc analysis 381382. In that population:

  • pain-free at 2 hours: 37.74% with Cerena versus 16.67% with sham (p=0.0181)
  • pain-free at 24 hours: 33.96% versus 10% (p=0.0025) 383384

Adverse event rates were similar between arms, and no seizures were reported 385. FDA noted that effectiveness on associated migraine symptoms was not demonstrated and that long-term data were not available 388.

SpringTMS lineage

  • K140094 (SpringTMS). Same indication as Cerena, with hardware and use-authorization changes. The summary cites the Cerena predicate as K130556 606608609.
  • K161663 (sTMS mini). A smaller device with an internal battery. Supported by non-clinical testing only 430434438.
  • K162797 (SpringTMS). Expanded the indication to "acute and prophylactic treatment of migraine headache," dropping the restriction to migraine with aura 731734. Support came from a prospective, single-arm, multicenter observational study measured against a performance goal. In the full analysis set (n=132), mean monthly migraine headache days fell by 2.8 from a baseline of 9.1 (P<0.0001) 735736. Of 217 subjects, 28.57% reported at least one adverse event, and none were serious 737.
  • K182976 (SpringTMS). Added adolescents aged 12 and older. No new clinical testing was done, and the summary refers back to the predicate's single-arm ESPOUSE study 743746747.
  • K230358 (SAVI Dual). Same indications as SpringTMS, adding a cellular use-authorization option. Supported by non-clinical testing 818822823811.

The prevention claim for sTMS is the clearest example in this field of a new indication cleared through a 510(k) on single-arm data compared against a performance goal, without a sham control.

Non-invasive vagus nerve stimulation (product code PKR)

gammaCore De Novo (DEN150048)

gammaCore stimulates the cervical vagus nerve through the skin on the side of the neck. It was granted for acute treatment of pain associated with episodic cluster headache in adults 265. The labeling states that safety and effectiveness in chronic cluster headache have not been established 267. The decision summary cites animal data on reduced trigeminal nucleus caudalis glutamate as a potential mechanism 266.

Evidence. Two randomized, double-blind, sham-controlled trials supported the grant 271.

  • ACT1 (pivotal). The endpoint was response at 15 minutes for the first attack. Overall, 26.7% of active subjects responded versus 15.1% with sham (P=0.10). In the episodic cluster headache subgroup, the rates were 34.2% versus 10.6% (P<0.01) 274.
  • ACT2. The endpoint was the proportion of attacks pain-free at 15 minutes. Overall, the rates were 13.5% versus 11.5% (P=0.71). In episodic cluster headache, they were 47.5% versus 6.2% (P<0.01) 276277.

There were no device-related serious adverse events 278. FDA noted uncertainty in the size of clinical improvement and no demonstrated benefit in chronic cluster headache 286.

Indication expansion through 510(k)

510(k)Indication changeEvidence basis
K171306, K172270, K180538None. User-interface, hardware and dose-programming changes 534770698Clinical studies not required 536771699
K173442 (gammaCore-S)Added acute treatment of migraine in adults 473476PRESTO, a sham-controlled randomized trial at 10 Italian centers. nVNS beat sham at 30 and 60 minutes but missed the 120-minute primary endpoint. Repeated-measures analysis supported superiority through 120 minutes 479
K182369 (gammaCore Sapphire)Added adjunctive preventive treatment of cluster headache 398402PREVA, a randomized sham-controlled trial of adjunctive preventive nVNS in chronic cluster headache; acute migraine data from K173442 were also carried over 404
K191830Added preventive treatment of migraine in adults 453456PREMIUM (ITT n=332). Benefit was shown only in a post-hoc mITT analysis of subjects with at least 67% adherence, with a gain of 0.74 migraine days per month (P=0.043) 459460461. EVENT (n=59) did not reach significance (P=0.56) 464
K203546Extended acute and preventive migraine indications to adolescents aged 12 and older 668671Extrapolation from adult data, plus a 9-patient adolescent study 671674
K211856Added hemicrania continua and paroxysmal hemicrania in adults 569572Extrapolation from cluster headache RCTs plus audits and case series. Benefit was reported in 15/19 and 11/14 patients 575

gammaCore shows how far a single De Novo classification can be extended through 510(k)s. Several of the added indications rested on trials that missed their prespecified primary endpoints, on post-hoc analyses, or on case series.

Remote electrical neuromodulation (product code QGT)

Nerivio Migra De Novo (DEN180059)

Nerivio is a smartphone-controlled stimulator worn on the upper arm. It treats headache by stimulating at a site away from the head and neck 302305. It was granted for acute treatment of migraine in adults without chronic migraine 306.

Evidence. The pivotal trial was a 252-subject, multicenter, randomized, double-blind, sham-controlled study 308. In the modified ITT population:

  • pain relief at 2 hours: 66.7% with active treatment versus 38.8% with sham, a gain of 27.9 percentage points (p<0.0001)
  • pain freedom at 2 hours: 37.4% versus 18.4% (p=0.0036)
  • relief of the most bothersome symptom: 46.3% versus 22.2% 311312

Device-related adverse events occurred in 4.8% of active and 2.4% of sham subjects. All were mild, and none were serious 313.

Nerivio 510(k) expansions

  • K201824. Removed the chronic migraine exclusion 705708. In the supporting data, 57.1% of participants had pain relief in at least half of their treated attacks 709.
  • K203181. Lowered the minimum age to 12 778781. In the adolescent study (final analysis set n=39), pain relief at 2 hours was 71.8% and pain freedom was 35.9% 784.
  • K223169. Added preventive treatment, with the device used every other day 681684. In a double-blind active-versus-sham trial, monthly migraine days fell by 3.97 with active treatment versus 1.28 with sham, a gain of −2.69 days (p<0.001) 685. Adolescent real-world app data were also submitted 688.
  • K232152 (Nerivio Infinity). Rechargeable version. No new clinical trials 487488.
  • K241756. Lowered the minimum age to 8 754757. Support came from a retrospective analysis of real-world data on 293 children aged 6 to 11, which reported no adverse events 759760.
  • K250405. Labeling updates on long-term use and use in pregnancy, supported by post-market real-world data and a published survey study 510511.

Nerivio is the clearest example in this field of real-world data supporting pediatric expansion through 510(k).

Thermal vestibular stimulation (product code QAR)

The TNM Device (DEN170023) delivers controlled thermal waveforms through earpieces placed in the ear canals to stimulate the vestibular system 328329. It was granted for prophylactic treatment of episodic migraine in patients aged 12 and older 330.

Evidence. The supporting study was a multicenter, triple-blinded, randomized, placebo-controlled study of adjunctive treatment in adults aged 18 to 65. It randomized 100 subjects: 50 active, 30 placebo and 20 partially active placebo 332333. The primary endpoint was the within-group reduction in monthly migraine days in the active group from baseline to month 3 847. Excluding subjects with more than 15 headaches per month, the results were:

  • per protocol: −3.6 days with active treatment versus −0.9 with placebo (p=0.0142)
  • ITT: −3.3 versus −0.9 (p=0.0246) 849850851

The ≥50% responder rates (57.1% versus 33.3% per protocol) were not statistically significant 852853. Use in patients aged 12 to 17 was supported by extrapolation, since the trial enrolled only adults 340. No serious adverse events occurred 336.

Cross-cutting regulatory observations

  1. De Novo with sham-controlled RCTs set the bar for each new mechanism. Every founding classification relied on a randomized sham or placebo comparison. FDA accepted mixed or post-hoc primary results where risks were low and could be mitigated by special controls 363381286.
  2. 510(k)s have carried new indications, not only design changes. Prevention claims for sTMS, nVNS and Nerivio 734456684, and new disorders for gammaCore 473572, were cleared by 510(k) under the same regulation, with FDA treating the added claim as the same intended use even when the predicate indication was narrower.
  3. The evidence behind pediatric expansions varies. It ranged from no new clinical data (SpringTMS K182976) 747, to adult extrapolation (gammaCore K203546, TNM) 671340, to prospective adolescent studies and real-world data (Nerivio K203181, K241756) 784759.
  4. OTC conversions in the PCC class depend on usability evidence. Cefaly Dual OTC, HeadaTerm 2 and Enso were supported by human-factors or usability work, not new effectiveness trials 500872657.
  5. Follow-on PCC devices increasingly cite literature or the predicate's clinical data. Examples include ALLIVE, Elexir and Enso 802599656. This keeps the bar to entry for supraorbital TENS competitors low.

Open questions worth pursuing

  • How CMS coverage policy and payer evidence reviews treat these devices compared with their FDA indications.
  • Postmarket safety signals for each product code.
  • Whether any newer De Novo requests for headache (for example, non-neuromodulation digital or optical therapies) are building on these special controls.
  • How EU MDR notified-body certification compares with the US pathway for the same devices.