Participant selection is a central part of human factors validation for combination products. Sponsors of autoinjectors, prefilled syringes and pens must decide which user groups to test as distinct populations, whether to include patients, lay caregivers, healthcare providers and pediatric users, and how to treat injection-naive versus experienced participants. If these choices are wrong, the agency may ask for additional studies, which delays the program.
The analysis below reviews the guidance framework FDA applies to participant selection, including how distinct user groups are defined and how the 15-participant minimum is applied. It then examines review precedents in which FDA reviewers judged sponsors' user-group choices, covering role, age, prior device or injection experience, and training status, along with the justifications reviewers accepted or rejected.
How FDA assesses user-group selection in human factors validation studies for drug-device combination products
FDA's general minimum of 15 participants applies to each distinct user population, not to a study as a whole. In its human factors (HF) reviews of autoinjectors, prefilled syringes, pens and other drug-device combination products, it asks whether each distinct user group is represented by enough participants who resemble the real intended users. Distinct groups can be separated by role (patient, lay caregiver, healthcare provider), by age (adult, adolescent, child), by prior injection or device experience, and by training status. FDA reviewers (mainly DMEPA, with CDRH human factors staff in earlier reviews) have flagged missing groups, undersized subgroups and unrepresentative substitutes. They have also accepted justified omissions when the use-related risk analysis or the intended-use setting supported them. This article summarizes the guidance framework and the review precedents behind it.
The guidance framework
Distinct user populations and the 15-participant minimum
The CDRH guidance Applying Human Factors and Usability Engineering to Medical Devices is the main reference for participant selection, and combination-product HF reviews apply the same logic. Sponsors should identify user groups through preliminary analyses and evaluations. Users should be treated as separate groups when differences in their tasks, knowledge, experience or expertise could affect how they interact with the user interface and create a different potential for use error 83. The guidance recommends at least 15 participants per distinct user population, not 15 across the whole study. Some device types may warrant a higher minimum 84.
The guidance names two pairings that are distinct when both are intended users:
- Healthcare providers and lay users are different professional categories. Each needs its own group of at least 15 87.
- Pediatric and adult users are different age categories. They are likewise separate populations when both are intended users 87.
Participants within each group should reflect the actual intended population as closely as possible. Relevant characteristics include age, education or literacy, sensory or physical impairments and occupational specialty. When the patients' medical condition can cause functional limitations, participants should have representative limitations too 8586. The guidance lists general knowledge of similar devices, and experience with the particular device, as user characteristics that affect ability to operate a device 303304.
Combination-product-specific guidance
The guidance Application of Human Factors Engineering Principles for Combination Products: Questions and Answers tells sponsors to consider intended-user characteristics, concomitant diseases and conditions, and limitations that may affect use. Its examples include decreased vision or hearing, varying literacy and cognitive decline in geriatric patients 897576. On training, it distinguishes three situations 306307308:
| Training scenario | Expectation |
|---|---|
| Training is part of the user-interface design requirements | Validate the training program itself, including how consistent training is delivered and how trainers are trained |
| Training is not part of the design requirements | The user interface should support safe and effective use without training |
| Training is optional or cannot be ensured | May test both trained and untrained users. Testing only untrained users may be acceptable when they represent the higher-risk scenario |
The draft guidance Contents of a Complete Submission for Threshold Analyses and Human Factors Submissions to Drug and Biologic Applications asks sponsors to report distinct user groups by participant number and type. It also asks them to explain the characteristics that distinguish groups and can affect interaction, such as limited hand dexterity or cognitive deficit 9091.
The FDA guidances cited here do not set an explicit rule for recruiting injection-naive versus injection-experienced participants 7375298. That expectation shows up most clearly in review documents, covered next.
Prior injection experience: generally a distinct user group
FDA's stated position in reviews
Review records show that FDA usually treats injection-naive and injection-experienced users as separate groups:
- SKYTROFA (lonapegsomatropin-tcgd), BLA 761177. In protocol feedback, the sponsor argued that every participant would be new to the novel growth hormone autoinjector. Reviewers rejected the idea that device novelty was enough to merge experience groups. They said injection-naive and injection-experienced users are generally separate, distinct user groups and asked for at least 15 of each 143.
- WAINUA, NDA 217388. FDA restated that injection-experienced and injection-naive participants are generally distinct groups, each needing at least 15. It accepted certain deviations in how participants were classified on a case-specific basis because those participants had extensive prior injection experience 33.
Undersized naive subgroups draw findings
- SYMJEPI (epinephrine), NDA 207534. DMEPA identified methodology flaws. The study had only eight naive lay users against the usual minimum of 15 per distinct group. Most participants had been trained. Participants may have been excluded for vision, hand-strength or literacy issues, and the sponsor did not adequately justify that the training reflected real-world US use 161163.
- SEMGLEE (insulin glargine), application 210605. The study called its participants representative of intended users. Even so, reviewers found only eight untrained pen-naive pediatric patients, below the previously advised minimum of 15 per distinct group, and planned to request more pediatric data. The pediatric protocol had not been submitted for advance review 64.
- BESREMI, BLA 761166. The caregiver group had only 10 injection-experienced and 5 injection-naive caregivers, against FDA's general expectation of 15 per group. In this case FDA found no usability-relevant differences between patients and caregivers and accepted their combined coverage of lay users 198.
Studies designed around experience strata
Many accepted studies split each role group by experience, and often by training status as well:
- DUPIXENT (dupilumab), BLA 761055. The standard prefilled syringe (PFS) study enrolled 23 injection-naive and 15 syringe-experienced patients, 15 injection-naive and 21 syringe-experienced lay caregivers, 15 syringe-experienced nurses and 15 pharmacists. The pharmacists performed product differentiation only 122123. The PFS with safety shield used a similar structure 124125.
- CYLTEZO (adalimumab-adbm), BLA 761058. The study had at least 15 trained and 15 untrained participants in each of three distinct groups (RA patients, prospective caregivers and HCPs), which matched FDA's prior protocol feedback. It also had at least 15 naive and 15 experienced participants in each patient and caregiver group 135278. Reviewers found no new or unique risks after IFU revisions 136.
- ZEMBRACE SYMTOUCH (sumatriptan), NDA 208223. The study enrolled 15 injection-naive trained, 15 injection-naive untrained and 15 injection-experienced untrained migraine patients. These groups were tied to realistic scenarios: moving from oral therapy to a first injection, untrained self-education as a worst case, and switching from another injectable 167168.
- ZURNAI (nalmefene) autoinjector, NDA 218590. The study used five groups of 15: first responders, plus injection-experienced and injection-naive lay adults and lay adolescents aged 10 to 17. DMEPA recorded no methodology concerns about user groups 151152.
"Naive" must be defined precisely
Several reviews show that "syringe-naive" or "pen-naive" is not the same as "injection-naive." In the KEVZARA (sarilumab) study, syringe-naive participants could have used a pen or autoinjector before 120. ADMELOG used pen-naive and pen-experienced strata 138. SIMPONI used SmartJect-naive and SmartJect-experienced categories, which describe device experience rather than general injection experience 146147. Sponsors should say which experience variable defines each stratum and why it is the one that matters for use-error risk.
Patients: representativeness, functional limitations and surrogates
FDA weighs whether participants share the physical, cognitive and perceptual characteristics of the intended patient population. For rare diseases, its decisions on surrogate or proxy patients have depended on the facts of each case:
- Accepted: DAWNZERA (donidalorsen), NDA 219407. Ionis used proxy patients and caregivers, citing the rarity of HAE and the absence of disease-related dexterity, visual or cognitive impairments that would affect usability. DMEPA consulted the clinical division, which agreed, and found the proxies reasonable 130. During review, FDA requested the number of actual HAE patients versus surrogates and subgroup detail on injection experience 200.
- Accepted: REDEMPLO (plozasiran), NDA 219947. Only 2 of 15 patient participants had FCS. The rest were adult surrogates with high triglycerides. After a consult confirmed that the surrogates represented FCS patients physically, cognitively, visually and anthropometrically, DMEPA found the approach reasonable 140141142.
- Rejected: HULIO (adalimumab-fkjp). FDA rejected the proposed use of surrogate participants for adolescents with JIA, offered because recruitment was slow. Those participants would not reproduce JIA-related joint swelling, stiffness and pain that limits finger, hand and wrist movement, which are the main considerations for autoinjector use 234235.
- Reviewable, not endorsed: REPATHA (evolocumab). The injection-experienced patient group consisted entirely of surrogates. DMEPA noted that the protocol had not been submitted in advance and stated that the approach did not preclude review 237238.
Patient-specific risks can also rule out substitute users. In an ANDA comparative use HF study for an Apotex teriparatide device, FDA acknowledged that caregivers are intended users of Forteo. It still rejected recruiting caregivers in place of patients, because the key question was whether elderly patients with osteoporosis could grasp and control a slimmer device with different shape and tactile features 2027.
Caregivers
FDA expects lay caregivers to be tested as a group when they are intended users, unless the sponsor justifies leaving them out:
- ADMELOG / ADMELOG SOLOSTAR, application 209196. A CDRH HF deficiency found the HF data inadequate because type 1 diabetes patients and lay caregivers, both intended users, were missing. FDA asked for HF data that included them or a justification for excluding them 189190. The sponsor had argued that patients without pen experience adequately represented caregivers 139.
- BENLYSTA. FDA accepted a study with no caregiver group because caregiver interactions with the device did not differ significantly from those of patients and HCPs 69.
- ZARXIO (filgrastim-sndz). Reviewers questioned how representative the caregivers were. Half of the 20 caregiver participants were 65 or older, even though partial dosing would mostly apply to pediatric patients dosed by their parents, and only five participants fit that parental-age profile 65.
- DUPIXENT. Caregiver recruitment was widened beyond AD caregivers to caregivers of people with other chronic illnesses in order to fill quotas. The review records this as a protocol deviation 126127.
Healthcare providers
FDA does not treat HCP participation as always required, and it does not exempt HCPs because of their professional skills. The test is whether the professionals who will actually perform each task are represented:
- Missing HCP group flagged: WAINUA, NDA 217388. HCPs were intended users but were not planned as a distinct group. FDA recommended adding 15 representative HCPs or justifying the omission 191.
- Untrained HCPs required: YCANTH, NDA 212905. Every participant had been trained, and the sponsor had not shown that training would reliably reach all real-world users. FDA recommended a supplemental study with 15 untrained HCPs as a distinct group 193. The later study enrolled 16 untrained providers, and DMEPA found the HF study acceptable 277.
- Setting-specific HCP groups: Tresiba FlexTouch 200 units/mL. FDA requested 15 trained and 15 untrained inpatient nurses, a user group not assessed in earlier studies 224.
- Omission accepted: EXDENSUR (depemokimab), BLA 761458. FDA agreed with leaving HCPs out of the PFS study based on its review of the use-related risk analysis, while the autoinjector study included 15 nurses 221. After GSK clarified that intended US users were HCPs only, FDA limited its review to HCP data and did not review the lay-user data 70239.
- Professional subgroups: ABILIFY MAINTENA KIT. FDA accepted recruiting only psychiatric-setting nurses and nurse practitioners, and called the omission of physicians "not ideal" but acceptable. Reconstitution and dose withdrawal are mainly nursing tasks, and reviewers had no reason to think physician practice differed 222.
- Pharmacists and differentiation tasks. For TOUJEO, FDA expected differentiation testing to include all representative users, including HCPs, patients and pharmacists 197. HUMALOG KWIKPEN and DUPIXENT enrolled pharmacists for differentiation tasks only 158160122. ZUSDURI classified pharmacy professionals as a separate group from nurses, medical assistants and physician assistants 229.
Pediatric and adolescent users
When pediatric self-injection is intended, FDA expects representative pediatric participants. It also checks whether the study design matches how dosing is actually divided between children and caregivers:
- NGENLA (somatrogon). FDA advised that HF studies include representative pediatric age groups expected to self-inject, caregivers and HCPs, using the final to-be-marketed product and IFU 63.
- IMBRUVICA, NDA 217003. FDA had recommended a distinct group made up mainly of pediatric and adolescent users, with caregiver help available as in real life, rather than combined patient-caregiver dyads. The sponsor did not adopt the recommendation. FDA still reviewed the results because the testing reflected task-sharing at home and recorded which dyad member performed each problematic task 195.
- XARELTO, NDA 215859. FDA had recommended pediatric HF validation data. Studies in children aged 10 to 17 did not support self-administration, so the sponsor proposed labeling that limits administration to adults 194.
- QFITLIA (fitusiran) enrolled 15 adolescents aged 12 to 17 together with injection-naive and injection-experienced lay caregivers. FDA found that the recruitment deviations did not prevent review 66. VOXZOGO tested trained and untrained adult caregivers and untrained patients aged 11 to 18 6768.
- ZEGALOGUE enrolled adolescent lay caregivers aged 10 to 17. These are a different category from adolescent self-injectors 71. FDA accepted the sponsor's explanation for failing to recruit 10-year-olds 199.
- SIMLANDI. The comparative use HF study enrolled only Humira-experienced adult RA patients. FDA concluded that a separate HF validation study, which showed few errors in adults and adolescents regardless of prior Humira experience, supported extending the findings to adolescents 282930.
Training status as a related dimension
Training status is a separate question from user role, but FDA reviews the two together. It expects study conditions to match the training users will actually receive:
- AFREZZA. FDA rejected the assumption that testing only untrained users covers intended use when some users will be trained and others will not. It asked for a justified decay interval between training and testing 273274279.
- INBRIJA, NDA 209184. "Untrained" participants had received an IFU overview, which FDA counted as training. FDA recommended a supplemental study in untrained patients 192.
- TYVASO DPI. The study left out the recommended untrained group. The sponsor argued that specialty-pharmacy nurses would train and certify every user 275276.
- ACTEMRA. Reviewers concluded that a 30-to-60-minute decay interval did not reflect a product dosed weekly to every other week. They did not require repeat testing, because the observed failures were known for this device type 285286.
Practical implications for study design
Taken together, the guidance and review record point to several design expectations:
- Define groups from the use-related risk analysis. Separate groups wherever role, age, experience or training could change use-error risk. Enroll at least 15 per group 8384.
- Treat injection-naive and injection-experienced users as separate groups by default. Device novelty alone has not been accepted as a reason to merge them 14333. Define precisely what "naive" means.
- Justify every omitted group in writing. Caregiver or HCP omissions have been accepted when interactions were equivalent or the risk analysis supported them 69221. They have been flagged when no justification was given 189191.
- Choose surrogates for the functional limitations that matter. A recruitment shortfall alone does not justify substitutes 235. Clinical-division concurrence has helped surrogate approaches get accepted 130141.
- Submit the protocol in advance. Several deficiencies and caveats involved protocols that FDA had not reviewed before the study ran 64142238.