Chetan Mishra
Chetan Mishra
Published Nov 17, 2026 · Updated Nov 23, 2026

Sponsors of drug-device and biologic-device combination products have to plan for two pediatric frameworks at once. The Pediatric Research Equity Act attaches to the drug or biologic application. Device-oriented expectations cover pediatric information, device design and human factors for young patients and caregivers. Misjudging how the two interact can lead to unexpected pediatric study obligations, rejected waiver or deferral requests, or postmarketing requirements that surface late in review.

The analysis below sets out which application changes trigger a PREA assessment for a combination product and how FDA has treated device constituents such as autoinjectors and prefilled syringes. It also covers how device capabilities affect deferrals, waivers and pediatric human factors programs. It draws on FDA statutory provisions, guidance and specific review precedents to show how the agency has applied these principles in practice.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.

PREA and pediatric device considerations for drug-device combination products: requirements and FDA precedents

A drug-device or biologic-device combination product faces two separate pediatric frameworks. The first is the Pediatric Research Equity Act (PREA), which attaches to the drug or biologic application. The second is a set of device-oriented expectations: section 515A pediatric information, pediatric device design and extrapolation policy, and human factors expectations for pediatric users and caregivers. FDA's precedents show that the device constituent rarely triggers PREA by itself. It does often determine which pediatric ages a product can serve, so it shapes deferrals, waivers, postmarketing requirements (PMRs) and human factors programs.

Key takeaways

  • PREA triggers are drug-centric. An application or supplement under FD&C Act section 505 or PHS Act section 351 needs a pediatric assessment if it is for a new active ingredient, indication, dosage form, dosing regimen or route of administration, unless FDA waives or defers it or the indication has orphan designation (orphan indications are exempt, except for the molecularly targeted oncology requirement) 44032466.
  • Adding an autoinjector is generally not a new dosage form. In the methotrexate autoinjector reviews (Otrexup, Rasuvo), FDA said adding an autoinjector did not itself trigger PREA. The new subcutaneous route for certain indications did 500507.
  • Device limits drive deferrals and waivers. Erelzi's available presentations were fixed-dose (a 25 mg syringe and a 50 mg autoinjector) and could not accurately dose children weighing under 63 kg. FDA therefore deferred the pediatric study for children under 63 kg and required a dose-adjustable presentation as PREA PMR 3110-1 637638.
  • Device limits can support a partial waiver, though usability is not a separate statutory ground. For Primatene Mist, FDA waived studies in children under 4 because young children lack the dexterity and coordination to use the device and because asthma is hard to diagnose at that age 131.
  • Pediatric human factors is a separate workstream from PREA. FDA treats pediatric and adult users as distinct populations when both are intended users, and recommends at least 15 participants per distinct user population in validation testing 582578.

Framework 1: PREA requirements that attach to the drug or biologic constituent

Triggers, deferrals and waivers

The PREA regulations are 21 CFR 314.55 for NDAs and 21 CFR 601.27 for BLAs. Under both, covered applications must contain data to assess safety and effectiveness in all relevant pediatric subpopulations and to support dosing and administration 32466. Effectiveness may be extrapolated from adult studies when disease course and drug effects are sufficiently similar, usually with supplemental pediatric information 33167.

FDA may defer pediatric assessments, for example when the adult product is ready for approval before pediatric studies are complete. The applicant's request must certify the grounds for the delay, describe the planned studies and show that they will be conducted with due diligence 33268.

FDA may grant full or partial waivers on these grounds 3343353367071:

Waiver groundFull waiverPartial waiver (specific age group)
No meaningful therapeutic benefit over existing pediatric therapies and unlikely to be used in substantial numbers of pediatric patientsYesYes
Necessary studies are impossible or highly impracticableYesYes
Evidence strongly suggests the product would be ineffective or unsafeYesYes
Reasonable attempts to produce a necessary pediatric formulation have failedNoYes

A formulation-based waiver covers only the age groups that need that formulation. A waiver based on evidence of ineffectiveness or lack of safety requires that information in labeling 33772. FDA's PREA compliance guidance says formulation-based waivers generally involve unusually difficult technological problems 446. The newer draft regulatory-considerations guidance also treats them as grounds for a partial waiver only 447. Indications with orphan designation are exempt from PREA 33974, though not from the molecularly targeted oncology investigation requirement 430444.

Where the device fits in pediatric study planning

The Pediatric Study Plans guidance covers the delivery device under its "Age-Appropriate Formulation Development" section. Sponsors should describe how they will ensure appropriate formulation design, including, to the extent practicable, the design of delivery systems, "such as capsules, tablets, infusions, and devices," to be used in pediatric studies 13. The initial pediatric study plan (iPSP) is due no later than 60 days after the end-of-phase 2 meeting unless FDA agrees to another date 467.

FDA's guidance on pen, jet and related injectors states that changes in dose, rate, route or method of injection may trigger PREA. For an injector proposed with an unapproved drug, or a new use of an approved one, the guidance recommends contacting the Office of Combination Products (OCP) to determine the lead center 468469.

Framework 2: Device-side pediatric expectations

Section 515A pediatric information

Under 21 CFR 814.20(b)(13), a PMA must include readily available information on any pediatric subpopulations (neonates, infants, children, adolescents) that have the disease or condition the device is intended to treat, diagnose or cure, and the number of affected pediatric patients 321. Similar provisions apply to humanitarian use devices 322. "Pediatric patients" means patients 21 years of age or younger at the time of diagnosis or treatment 323. FDA's guidance applies section 515A to PMAs, PMA supplements, HDEs and PDPs 209. It asks for a separate section titled "515A Pediatric Device Use Information" 221. It also states that providing this information does not by itself establish safety and effectiveness for a pediatric indication 225.

FDA's device guidances use these pediatric subgroups 212213214215:

SubgroupAge range
NeonatesBirth through the first 28 days
Infants29 days to less than 2 years
Children2 years to less than 12 years
Adolescents12 through 21 years (up to the 22nd birthday)

Pediatric device design and extrapolation

Premarket Assessment of Pediatric Medical Devices says clinical data are not needed for every pediatric device. Bench and animal testing may be enough in some cases 176. Under "Human Factors," the guidance lists considerations for each targeted pediatric subgroup 472:

  • device size
  • manual dexterity and strength
  • labeling clarity and ease of use
  • age-appropriate interfaces
  • the maturity needed for safe operation

For implantable devices, the guidance names combination-product issues such as possible drug/device interactions 473.

The pediatric extrapolation guidance allows existing adult data to support pediatric indications in PMA, De Novo and HDE submissions. Its policies do not apply to 510(k)s 254255. It also asks whether pediatric use needs changes in device size, materials, energy delivery or human factors, including self-administration versus administration by a guardian. If such differences could affect safety or effectiveness in a clinically meaningful way, full extrapolation is not feasible 606607.

Which framework applies to a given combination product

FDA assigns the lead center (CDER, CBER or CDRH) based on the product's primary mode of action (PMOA) 362363342. A single marketing application is generally appropriate 365. The non-lead constituents still need supporting data. For a drug-led product, that means device engineering, biocompatibility, performance and design-validation data 368. Postapproval changes to the device constituent of an NDA-approved product are submitted to that NDA 371372.

FDA guidances do not state an explicit rule mapping lead-center assignment to PREA applicability. The pen/injector guidance likewise does not say that center assignment itself decides whether PREA applies 469. PREA's triggers are written for section 505 and 351 applications 440. In practice, this means PREA attaches through the NDA or BLA for drug-led and biologic-led combination products. PMA, PMA supplement, HDE and PDP submissions instead carry the section 515A information requirement 209. 510(k) and De Novo requests do not. Sponsors with unclear jurisdiction can use an RFD or Pre-RFD with OCP 367.

Precedents: when a device change does or does not trigger PREA

Product / applicationFDA position on PREABasis
Otrexup (methotrexate autoinjector), NDA 204824Adding an autoinjector to injectable methotrexate "does not trigger PREA as this change is not considered a new dosage form." The new SC route for RA and psoriasis did trigger PREA.500
Rasuvo (methotrexate autoinjector), NDA 205776Autoinjector not a new dosage form and did not itself trigger PREA. The new route for RA and severe psoriasis was the trigger.507
Rasuvo, earlier filing checklistAfter Otrexup's approval, the reviewer noted that the SC route no longer triggered PREA, but a different fixed-dose regimen "might" still do so. This was not confirmed at filing.512
Brekiya (dihydroergotamine autoinjector), NDA 215400, 2025"This application does not trigger PREA." There was no new active ingredient, indication, dosage form, dosing regimen or route.510
Auvi-Q (epinephrine autoinjector), NDA 201739, 2009 pre-NDA adviceNone of the five PREA triggers applied, but FDA encouraged study in all appropriate age groups and noted a possible BPCA Written Request.502
Symjepi (epinephrine prefilled syringe), NDA 207534, 2017The application did not trigger PREA, so FDA did not require lower doses.417418
Auvi-Q 0.1 mg strength supplementThe new 0.1 mg strength triggered PREA as a new dosing regimen. The existing 0.3 mg and 0.15 mg strengths had not.409411412

In this set, a new device presentation of an already-approved drug did not, by itself, count as a new dosage form. PREA was triggered when the presentation came with a new route, indication or dosing regimen, such as a new strength for smaller children. The Rasuvo checklist shows that fixed-dose device presentations can raise a dosing-regimen question, and sponsors should settle it with the review division early.

Precedents: device limitations shaping deferrals, waivers and PMRs

Erelzi (etanercept-szzs), BLA 761042

As a non-interchangeable biosimilar, Erelzi was treated as having a new active ingredient for PREA purposes 492442. At approval on August 30, 2016, FDA took these actions 634635636637638:

  • Waived studies in some indications, including pJIA in children under 2 years.
  • Deferred the pJIA study for children weighing less than 63 kg. The available fixed-dose presentations (a 25 mg syringe and a 50 mg autoinjector) could not accurately dose children weighing under 63 kg.
  • Imposed PMR 3110-1: "Develop a presentation that can be used to accurately administer etanercept-szzs to pediatric patients who weigh less than 63 kg."

Sandoz submitted the final report in May 2021, and FDA found PMR 3110-1 fulfilled on November 10, 2022 640. The device limitation led to a deferral and a requirement to develop a new presentation. It did not lead to a waiver for that population.

Otrexup and Rasuvo (methotrexate autoinjectors)

For Otrexup, FDA considered the PJIA requirement for ages 2 to 17 fulfilled because methotrexate was already labeled for subcutaneous use in that population. Studies under age 2 were waived as impossible or highly impractical 504535. The applicant had asked to waive studies through age 6 because the minimum 10 mg dose could not support small weight-based increments. The clinical team did not accept that request 536. Labeling instead described the device's dosing limitations 504. For Rasuvo, the CDTL memorandum accepted a similar waiver for the youngest PJIA patients and a pediatric psoriasis waiver. PeRC concurred on April 2, 2014 508.

Auvi-Q (epinephrine autoinjector), NDA 201739, 0.1 mg strength

FDA found that literature-supported PK/PD modeling plus a sonographic needle-length study fulfilled PREA for patients weighing 7.5 kg or more. No new clinical studies were required 413. The sonographic study supported an exposed needle length that would not reach bone in the mid-anterolateral thigh 412. For patients under 7.5 kg, the Division recommended a waiver because a fixed dose was inappropriate and vial formulations allowed weight-based dosing. PeRC agreed 413383. FDA also noted dosing errors with vial-and-syringe preparation and the risks of excess dose and bone contact when devices meant for larger children are used in smaller ones 414. In an earlier case, the Twinject 0.15 mg strength supplement (NDA 20-800, S-001, 2004) triggered PREA. FDA deferred the assessment in patients under 15 kg 513514.

Primatene Mist (epinephrine inhalation aerosol), NDA 205920

FDA waived studies in children under 4. The reasons were that young children lacked the dexterity and coordination to use the device and that asthma is difficult to diagnose at that age. FDA deferred studies in ages 4 to 11 as PMR 3514-1 131132. The clinical reviewer had argued for a different waiver ground: no meaningful therapeutic benefit over existing therapies, with unlikely substantial use 540.

Baqsimi (nasal glucagon), NDA 210134

FDA accepted the pivotal pediatric study in ages 4 to under 17 after discussion with PeRC 669673. It waived studies under 1 year because of the small number of patients 674. It deferred ages 1 to under 4 as PMR 3621-1 675676. A 2025 supplement fulfilled that PMR with Study IGBO and expanded the indication to age 1 and older 672677.

Autoinjector and pen BLAs and NDAs with PREA PMRs or orphan exemptions

Product / presentationOutcomeSource
Nucala (mepolizumab) autoinjector and safety syringePREA PMRs for a PK/PD study in ages 6 to 11 and a 12-month extension. Studies under 6 were waived as impossible or highly impracticable.8689919293
Benlysta (belimumab) SC autoinjector, BLA 761043/S-034, June 2025Approved for patients 5 and older with lupus nephritis as the response to a PREA PMR.979899
Ngenla (somatrogon) prefilled pen, BLA 761184Exempt from PREA because of orphan designation. The device PMC covered shipping studies.100101
Dawnzera autoinjector, NDA 219407Exempt from PREA because of orphan designation.94
Pyzchiva (ustekinumab) autoinjector supplements, 2025FDA stated no pediatric studies would be required under PREA for these sBLAs. The letter excerpt does not state the reason.83

Pediatric human factors for combination products

Guidance expectations

FDA's human factors guidance treats pediatric and adult users as distinct populations when both are intended users. Testing should include representative pediatric users 582. FDA recommends at least 15 participants from each distinct user population 578. User groups are defined by characteristics or tasks that could affect use, such as age, literacy and physical impairments 580581. The combination-product human factors guidance names patients and lay caregivers, including in homes and schools, as intended users. Where training cannot be ensured, the HF validation study may test both trained and untrained users 584586. One FDA example says adding pediatric users introduces new risks even without an interface change and calls for a new HF validation study 587.

FDA's draft guidance on bridging for drug-device and biologic-device combination products gives a worked example. Adult user-interface data could be bridged, but a pediatric human factors assessment was still needed because children had not been studied 449. When the intended population changes, the draft also calls for these steps 471:

  • reassess device design inputs and outputs
  • run additional verification testing
  • conduct clinical studies with the to-be-marketed device
  • submit a pediatric HF validation report

Review precedents

Product / applicationPediatric HF finding
Semglee (insulin glargine), NDA 210605FDA found too few untrained, injection-naïve pediatric users in the original studies. It recommended a new validation study with 15 such pediatric patients, and the follow-up study tested children aged 6 to 11 and 12 to 17. 424344
Simlandi (adalimumab-ryvk), BLA 761299HF validation included adolescents with JIA. They made no more device-use errors than adults. FDA concluded a separate adolescent comparative-use study was unlikely to add information for the interchangeability assessment. 45464748
RiVive (naloxone nasal spray), NDA 217722The HF validation study enrolled ages 12 to 17. FDA said results could not be generalized to younger untested ages. 5253
Narcan (naloxone nasal spray), NDA 208411/S-006The HF validation study enrolled ages 15 to 17. FDA said the data could not be generalized to adolescents younger than 15. 565758
TOBI Podhaler, NDA 201688Most pediatric participants needed caregiver help. The reviewer recommended "not approvable" pending device and instruction improvements, and recommended that caregivers or providers set up the device. 60616263
Exdensur (depemokimab), BLA 761458Pediatric and caregiver groups were tested, but FDA focused on healthcare-professional data because the sponsor confirmed HCP-only administration. 4950
Vistaseal (CBER), BLA 125640FDA tied human factors validation deficiencies to the pediatric PREA PMR, and the applicant committed to a new HF study within the pediatric trial. FDA later accepted dropping that study after a 2019 change to the device and IFU. 296298

Inhalation products: spacers and valved holding chambers

Inhaled products show the delivery device shaping pediatric evidence. For Ventolin HFA, studies in children from birth to under 4 years used facemask spacers or valved holding chambers and did not show efficacy 138139. FDA said efficacy could not be extrapolated from older children, whose studies had not used a holding chamber and facemask 140. The resulting labeling states that safety and effectiveness below age 4, with or without a spacer, had not been demonstrated 143. For Spiriva Respimat, children aged 1 to 4 used a holding chamber with a facemask in trials. Bench testing used pediatric flow rates and found that fine-particle dose fell with longer holding times 144145. For Dulera, FDA deferred pediatric studies for ages 5 to under 12 and waived them below age 5 135. Xopenex HFA's deferred PREA program included MDI studies in children from birth to about 48 months 136137.

Practical implications for sponsors

  • Map PREA triggers at the constituent level. A new device presentation of an approved drug has generally not been treated as a new dosage form 500507510. Confirm whether a new route, strength or fixed-dose regimen is involved, because those changes have triggered PREA 411507512.
  • Address device capability in the iPSP. FDA expects formulation and delivery-system design for pediatric studies to be described 13. Where a fixed-dose device cannot cover low-weight patients, the precedents point to a deferral plus a PMR for an age-appropriate presentation (Erelzi), or to a partial waiver where alternative dosage forms allow weight-based dosing (Auvi-Q under 7.5 kg) 637638413.
  • Do not assume a device dosing limit justifies a broad waiver. In Otrexup, FDA did not accept a request to waive through age 6 based on the minimum 10 mg dose 536.
  • Plan pediatric and caregiver HF validation with the pediatric program. FDA has declined to generalize human factors results to younger untested ages (RiVive, Narcan) and has required more pediatric validation when representation was inadequate (Semglee) 535744.
  • For device-led products, prepare the 515A section and use the device extrapolation framework when seeking pediatric indications 221254.

Questions a reader may want to pursue next include how PeRC has handled PREA for co-packaged versus single-entity combination products, and how pediatric HF requirements have been written into specific PREA PMRs.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.