Break-Loose, Glide and Injection Force Testing for Prefilled Syringes and Autoinjectors in FDA NDA and BLA Reviews

Chetan Mishra
Chetan Mishra
Published Nov 22, 2026

Break-loose and glide force decide whether a prefilled syringe or autoinjector delivers its full dose reliably over its shelf life. FDA reviewers often question these attributes in combination product submissions. When the verification data, conditioning or control strategy behind them is weak, the result can be information requests, review-cycle delays or post-marketing commitments, so device, CMC and regulatory teams need to know what reviewers expect before they file.

The analysis below draws on published Drugs@FDA and CBER review documents. It covers how reviewers treated break-loose, glide and injection force as essential performance requirements, and what they expected for testing on the final finished product, real-time aging and post-shipping conditioning, statistical acceptance approaches and lot-release controls. It also covers how sponsors applied the ISO 11040-4, ISO 11040-8 and ISO 11608 test methods at the syringe and autoinjector levels.

Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.

Break-loose, glide and injection force testing for prefilled syringes and autoinjectors: what FDA reviewers ask for in NDA and BLA reviews

Break-loose force (the force to start the plunger stopper moving) and glide or extrusion force (the force to keep it moving) are device attributes that FDA reviewers repeatedly question in combination product reviews. Published Drugs@FDA and CBER review documents show a recurring pattern. FDA reviewers often expect these forces to be treated as essential performance requirements (EPRs) for a prefilled syringe (PFS). They often ask for them to be verified on the final finished combination product, maintained through real-time shelf life and after shipping, and controlled at lot release, although some reviews accepted alternatives. For autoinjectors, FDA shifts the focus to system-level outputs such as injection time and activation force, which are tested under the ISO 11608 framework.

Most numeric force limits are redacted in the public reviews. This article therefore covers the test design, conditioning, statistics and control strategy that reviewers accepted or challenged, plus the observed results that are disclosed.

Key takeaways

  • FDA's draft guidance on essential drug delivery performance characteristics (EDDCs) names break-loose force and glide force as EDDCs for a prefilled syringe used as the delivery device. When the PFS is a subassembly inside an autoinjector, injection time is the system-level EDDC, and break-loose and glide force are component attributes reflected in it 147148.
  • Reviewers repeatedly asked sponsors to test break-loose and glide force on the drug-filled combination product with the marketed needle or needle safety device, not on the bare syringe 233443444.
  • For Entyvio, shelf-life support had to come from real-time-aged samples, and FDA did not accept pooling earlier time points to verify later ones 225. It also requested force testing after simulated shipping 231475.
  • FDA often pushed break-loose and glide force into lot-release specifications, though it accepted alternatives in some cases (see Sustol below; Cabenuva was accepted without designating these forces as EPRs). Where they were not in place at approval, it was handled through information requests, resubmission deficiencies or post-marketing commitments 228229268270272.
  • ISO 11040-8:2016 (with the ISO 11040-4 test setup) is the standard reviewers pointed to for PFS force methods 47850. It was replaced by ISO 11040-8:2026 in June 2026, so new filings should check the current edition. Reviews applied ISO 11608-1 to autoinjector dose accuracy and preconditioning, and ISO 11608-5 to automated functions and needle cover lock-out 532543.

How FDA guidance frames the forces

Essential drug delivery performance characteristics

The draft guidance Essential Drug Delivery Performance Characteristics for Devices Intended to Deliver Drugs and Biological Products describes EDDCs as system-level functions that depend on device design rather than user actions. FDA recommends verifying, validating, controlling and maintaining them over the product lifecycle, and generally including specifications and acceptance criteria in the submission 149135150. The guidance lists break-loose force and glide force as EDDCs for a prefilled syringe 147.

The guidance's hypothetical autoinjector example draws the distinction clearly. Dose accuracy, extended needle length, activation force and injection time are the EDDCs. Break-loose force, glide force and needle inner diameter are component attributes that show up in injection time. That classification does not mean the component forces are unimportant to drug quality or manufacturing 151152153154155. The guidance also says the "primary functions" in ISO 11608-1:2022 do not necessarily align with FDA's EDDCs. Sponsors can use data from the standard to support an EDDC when the test method fits that characteristic 167168.

ISO 11040-4 conformance alone does not show interoperability

FDA's 2013 draft guidance Glass Syringes for Delivering Drug and Biological Products: Technical Information to Supplement ISO Standard 11040-4 explains that conformance to ISO 11040-4 alone does not demonstrate interoperability, even with an ISO 594-2-conforming connector 169170. (ISO 594-2 has since been replaced by ISO 80369-7.) For the connected system, it recommends supplemental data on seal integrity throughout shelf life, glide force, break force and separation force 171172173.

Pen and autoinjector performance testing

The final guidance Technical Considerations for Pen, Jet, and Related Injectors Intended for Use with Drugs and Biological Products recommends using the most current FDA-recognized versions of applicable standards, including ISO 11608-1 for pen-injector requirements and test methods. Testing should use the final marketed injector with the intended drug product, or the sponsor must justify representative products. Submissions should report methods, endpoints, predefined acceptance criteria, results, analysis and conclusions 163164165166.

For emergency-use injectors, FDA's reliability draft guidance makes failure to inject the primary endpoint. It generally includes dose accuracy, extended needle length, activation force and injection time in the reliability analysis. Acceptance criteria should reflect clinical performance, not manufacturing capability 158126159.

Prefilled syringe break-loose and glide force: what reviewers accepted

The table summarizes disclosed test conditions and results from approved BLAs. Observed values are results, not acceptance limits, which are mostly redacted.

Product (application)Test design and disclosed results
Dupixent (BLA 761055)Break-loose and mean glide force measured on bulk PFS at release and on the finished safety-system PFS on stability. In one assembled lot, break-loose rose from 4 to 6 N and glide from 9 to 11 N between baseline and 1 month accelerated, staying below specification. FDA requested updated long-term and accelerated data, then required no further shelf-life information 373375376377240.
Skyrizi (risankizumab)Instron plunger-force method with a 29 G needle. Across seven batches, break-out was about 4 N (batch maxima 4 to 5 N) and gliding 7 to 9 N (batch maxima 8 to 10 N). The reviewer found requirements met; limits redacted 380381382.
Hyrimoz (BLA 761071)Routine testing at PFS level was considered sufficient because the needle safety device activates after full plunger depression. Development testing of the assembled product (20 samples per test) gave break-loose 5.63 N mean and glide 4.11 N mean; both passed. The design-verification limit came from formative human factors work 385386.
Tremfya (BLA 761061)For the 1.0 mL PFS, piston release force averaged 3.31 N (2.82 to 3.75 N) and piston travel force 5.08 N (3.98 to 6.64 N) across two trials of 10 390391392393.
Praluent (BLA 125559)Final assembled, drug-filled syringes compression-tested at 240 mm/min after shield removal. Break-loose defined as the maximum force to initiate stopper movement and glide as the maximum to complete movement. All results passed; the reviewer cited FDA's glass-syringe guidance supplementing ISO 11040 412413414.
Tezspire (BLA 761224)Clinical accessorized PFS stored at 5°C tested at 36 months with validated design-verification methods (n=10 per parameter). Peak break-loose 12.6 N and peak extrusion 16.2 N, both within specification. FDA requested a 95% confidence / 95% reliability analysis using the ISO 11608-1:2014 Table B.1 k-factor 398399.
Ziextenzo (BLA 761045)At 5 ± 3°C, break-loose values were 2.37 to 2.79 N at 0 months and 3.11 to 3.14 N at 36 months 404.
Fulphila (BLA 761075)After transport vibration, 20 samples met force requirements: break-loose 3 to 6 N (mean 4 N) and glide from under 1 N to 3 N (mean 2 N) 407.
Hulio (BLA 761154)Long-term glide testing used 5 samples at 205 mm/min, which the reviewer considered reasonable. Break-loose was not tested because the sponsor said the coated stopper needed only an extrusion specification; the reviewer found this provisionally acceptable and requested raw data 456457458459.

Formulation and component selection also feature in reviews. For Palynziq (BLA 761079), component-selection studies showed glide force increased with longer needles, smaller internal radius, higher viscosity and lower temperature 401.

Recurring FDA requests and deficiencies

Test the final finished combination product

For Cligavyx (NDA 210063), prior Complete Response deficiencies asked for PFS design specifications including break-loose and glide force, tested on the drug-filled combination product with the supplied safety needle of the same gauge and length 233441. For a fulvestrant 505(b)(2) (NDA 210326), FDA confirmed aged, drug-filled samples were tested with the safety glide needle 446.

FDA also questioned whether syringe-only testing captures the finished device. For Hadlima (BLA 761059), FDA found syringe-only release testing adequate for deliverable volume and break-loose force but not for glide force, because it missed the needle-safety activation peak seen with the finished device 115. For Camcevi (NDA 211488), a Refusal to File cited the need for protocols and reports verifying dose accuracy, break-loose and glide force on the final finished product with a sample size sufficient for a 95% confidence interval. FDA rejected substituting stability data for design verification because the stability sample size was inadequate 265.

Justify acceptance criteria against clinical function

For Trulicity (BLA 125469), FDA distinguished process capability from design verification. It asked for functional limits tied to safe and effective delivery and for justified lower limits on initiating and maintaining injection, since only upper bounds were listed 424425. For Amvuttra (NDA 215515), FDA asked the sponsor to set a two-sided break-loose/glide specification or justify a one-sided limit, and to reconcile different force criteria used in design verification and shelf-life testing 244245. For Ajovy (BLA 761089), FDA requested a traceability matrix linking specifications to performance data 462.

Demonstrate performance through shelf life and after shipping

For Entyvio (BLA 761133), FDA asked for PFS EPR testing through the proposed shelf life using real-time-aged samples and did not accept pooling earlier time points to verify later ones. The sponsor supplied 18-month real-time data 225. FDA also reiterated a request for stability injection-force curves to detect secondary peaks from clogging 468. For Andembry (BLA 761367), FDA requested stability data on at least three lots 471472.

Shipping is a standard preconditioning step. For Fensolvi (NDA 213150), FDA asked for average break-loose force through shelf life and post-shipping dose accuracy, break-loose and glide force for each syringe and the mixed final syringe 231. For Gvoke PFS (NDA 212097), FDA asked for break-loose and glide verification as essential functions and recommended justifying worst-case preconditioning (expiry-equivalent accelerated aging, vibration, temperature cycling and drop) applied sequentially 476.

Control forces at lot release and on stability

FDA consistently pushed force testing into the commercial control strategy:

  • Idacio (BLA 761255): FDA asked why existing design specifications could not immediately serve as release criteria for break-loose/glide and activation force, and asked for these EPRs to be added to the post-approval stability protocol. The sponsor implemented both 228229.
  • Eticovo (BLA 761066): a resubmission deficiency required break-loose and glide force in lot-release specifications 268. The recommended post-market commitment covered adding them after method validation and commercial-batch data generation 270.
  • Erelzi (BLA 761042): the sponsor committed to implementing PFS break-loose/glide release specifications based on commercial-batch data 272.
  • Sustol (NDA 022445): PMC 3094-7 added break force, glide force and peak injection force to the stability protocol, with real-time data supplementing accelerated data. FDA accepted stability testing in lieu of lot-release testing 267.
  • Udenyca (BLA 761039) and Perseris (NDA 210655): FDA asked for break-loose/glide specifications in section 3.2.P.5.1, with supporting verification and shelf-life evidence 253254260261.

FDA does not always require these as EPRs. For the Cabenuva co-pack, the sponsor explained the syringes are filled at point of use and that break-loose/glide and injection force were not designated EPRs. FDA marked the response adequate 250251252.

How ISO 11040-4 and ISO 11040-8 are applied

Reviews separate standards cited for force methods from standards cited for syringe component conformance.

ISO 11040-8 for force methods. For lanreotide (NDA 215395), FDA asked the sponsor to test the PFS under ISO 11040-8, which references the ISO 11040-4 test setup, in the final finished form. The original test speed and data acquisition rate differed from the recommendations, and the test device lacked the external needle safety feature. The repeated comparative injectability study against Somatuline Depot used 100 mm/min and 100 Hz for break-loose, maximum and dynamic glide force, with a percent-difference acceptance criterion. FDA also required justification of acceptance criteria and sample size where the standard does not specify them 5051525354.

Other reviews show similar method assignments:

Product (application)Standard appliedAttribute
Zegalogue (NDA 214231)ISO 11040-8 §6.2 (extrusion force)Break-loose and glide force, upper limits in N; deliverable volume also under ISO 11040-8 with gravimetric testing 7980
Amvuttra (NDA 215515)ISO 11040-8:2016 §6.2 at 100 mm/minBreak-loose and glide force, and needle safety activation force 85
Rykindo (NDA 212849)ISO 11040-8 §6.2 and ISO 11040-4 Annex EBreak-free and glide force, before and after reconstitution 88
Xtrenbo / Enoby (BLA 761439)ISO 11040-8 for break-loose; ISO 11040-4 Annex E for glide; ISO 11040-4 Annex G.6 for rigid needle shield removalUpper limits for both forces at release and end of shelf life; range for shield removal 606162
Avtozma (BLA 761420)ISO 11040-4 and 11040-8PFS break-loose, glide and activation force at release and on stability; autoinjector cap removal under ISO 11608-1:2022 5657

ISO 11040-4 for syringe component attributes. For Dupixent, FDA accepted ISO 11040-4 prefilled-syringe tests including rigid needle shield removal (observed 7 to 18 N across three tests of 32 syringes) and glass flange breakage resistance (observed minima of 121 to 193 N) 3866. Andembry cited ISO 11040-4:2015 for rigid needle shield removal and ISO 7864:2016 for needle pull-out, not for plunger forces 471474. Abrilada cited ISO 11040-4 for syringe conformance rather than as the force method 438. Fulphila used ISO 11040-4:2015 for barrel dimensions, glass type and hydrolytic resistance 93.

Applicability limits. For Fensolvi, the sponsor treated ISO 11040-8 as guidance rather than directly applicable because the user mixes the product before injection, and the reviewer accepted leveraging prior shipping validation 116117. Safety-system activation and override forces were typically assessed under ISO 23908 rather than ISO 11040 6285.

Autoinjectors: applying ISO 11608-1 and ISO 11608-5

For autoinjectors, reviewers used ISO 11608-1 for dose accuracy and for preconditioning before performance testing, and ISO 11608-5 for automated functions including needle cover lock-out. Injection time, activation force, needle extension and cap removal usually carried device-specific limits 532533534535.

Product (application)ISO 11608 application and findings
Nucala (BLA 761122)Delivered volume met ISO 11608-1:2014 at standard, cool and warm atmospheres, after dry heat and cold storage (60 each), free fall (30) and vibration (20). Under ISO 11608-5 §§4.3.6 to 4.3.7, the guard shielded and locked automatically. Mean results: cap removal 10.05 N, actuation 7.28 N, needle extension 6.05 mm, delivery time 8.97 s, cover override 222.38 N 532536537539540541542.
Hyrimoz (BLA 761071)Injection time into air, delivered volume (ISO 11608-1:2012, n=60, one-sided tolerance bound at 95% confidence and 97.5% probability content), exposed needle length, activation force and activation travel passed. ISO 11608-5:2012 needle-cover displacement under load passed 543533.
Erelzi (BLA 761042)Parts 1 and 5 informed design verification of cap removal torque, activation force range, exposed needle length, maximum injection time and post-use cover displacement. FDA asked for system-level EPR verification on the final product in the BLA rather than reliance on a master file, and recommended adding injection depth and audible/visual feedback to lot release 550551555559.
Hadlima PushTouch (BLA 761059)Deliverable volume per ISO 11608-1 and 11608-5 (cold storage row ≥0.8 mL, n=60). FDA found n=60 treated as attribute testing did not give adequate reliability; the sponsor reanalyzed with a k-value for 95% confidence and 97.5% reliability 560562564565.
Qfitlia pen (NDA 219019)All ISO 11608-1 functional criteria met. Needle-cover lock override cited the ISO 11608-5 requirement to withstand at least twice the maximum activation force 566567.
Bimzelx (BLA 761151)Dose and injection time tested under ISO 11608-1:2014 preconditions (30 to 60 devices). Maximum observed injection time was about 11.8 s after cool conditioning, against a labeled 15 s hold 535.
Mounjaro (NDA 215866)Applicable ISO 11608-5:2012 sections met. Injection time measured from button press to completion of the timed function; prior platform data accepted for unlock torque 568569570.
Zurnai (NDA 218590)ISO 11608-1:2022 D1 delivered volume ≥0.50 mL. The reviewer noted a high activation-force specification but found no triggering difficulty in human factors testing, including pediatric users 576577.
Rasuvo (NDA 205776)FDA asked the sponsor to compare its testing to the 2000 edition against ISO 11608-1:2012, test changed or new requirements and justify deviations. The 2012 edition introduced a one-sided dose tolerance interval for D1 devices 589590591.

The link between PFS forces and autoinjector performance appears explicitly in Tezspire, where FDA requested activation-force testing on the final combination product because glide force affects activation force 276.

Statistical expectations

Reviewers routinely examined sample sizes and confidence/reliability claims for force and dose tests:

  • 95% confidence / 95% reliability with the ISO 11608-1:2014 Table B.1 k-factor for Tezspire break-loose and extrusion force 399.
  • 95% confidence / 99% reliability for Amvuttra activation force through shelf life and for Nyvepria break-loose and extrusion force (50 samples each for real-time and accelerated aging) 86464465.
  • 95% confidence / 97.5% reliability or probability content for autoinjector dose testing (Hyrimoz, Hadlima, Wainua) 543565582.
  • For Abrilada, FDA asked for data distributions, statistical criteria and a justification of sample sizes below 30 for individual strengths 436437.
  • For Avtozma, FDA questioned varying preconditioned sample sizes (n=10 to 60) and accepted the ISO 11608-1:2022 justification with supporting k-values 583584.

Practical implications for submissions

Based on these reviews, a PFS or autoinjector dossier that anticipates FDA's questions will typically:

  1. Classify break-loose and glide force as EPRs for a standalone PFS, or explain why injection time and activation force capture them for an autoinjector 147148.
  2. Define the ISO 11040-8 / ISO 11040-4 method parameters (test speed, acquisition rate, needle, marketed safety device; reviews cite the 2016 edition of ISO 11040-8, so check the 2026 edition for the current test setup) and justify acceptance criteria and sample sizes the standard leaves open 5054.
  3. Tie limits to clinical function and user capability, including lower limits where relevant, rather than process capability alone 424425159.
  4. Verify on the final finished combination product after worst-case preconditioning and shipping, with real-time-aged data covering the full proposed shelf life 233225476.
  5. Place the forces in 3.2.P.5.1 release specifications and the stability protocol, or justify why incoming controls or stability testing are sufficient 253228267.
  6. State confidence and reliability levels and the statistical basis for each test 399564465.
Rhizome is an AI regulatory coworker that researches 143 health authority databases, and cites a specific page for each claim. Create a free account to ask it questions or have your agents search its databases.Connect it to your Claude or ChatGPT for free to ask it questions or have your agents search its databases.