# FDA Regulation of Generative AI and LLM-Enabled Clinical Software

How FDA's existing device framework, 520(o) software exclusions, and PCCPs apply to LLM-enabled clinical software, and whether any generative AI device has been authorized.

Published by Rhizome AI on 2026-10-16. Web version: https://rhizomeai.com/articles/fda-regulates-generative-ai-and-llm-enabled

## How FDA regulates generative AI and LLM-enabled clinical software

### Short answer

FDA has no generative AI-specific statute, regulation, or final guidance. LLM-enabled clinical software goes through the existing device framework. Whether a product is a device depends on its intended use and function, under the device definition and the section 520(o) software exclusions. A product that is a device follows the usual risk-based pathways (510(k), De Novo, PMA). Those pathways are layered with FDA's AI-specific lifecycle recommendations (still in draft, not binding requirements) and, where relevant, a predetermined change control plan (PCCP) [124][219][438].

On authorizations: at the November 2024 Digital Health Advisory Committee (DHAC) meeting, FDA stated, "we have not authorized any generative AI-enabled devices, nor have we authorized any adaptive AI-enabled devices" [352]. At the November 2025 DHAC meeting, FDA made a narrower statement: "to date, there are no generative AI-enabled medical devices that have been authorized for use in mental health conditions" [353]. That statement does not address other clinical areas. A targeted review of 510(k) and De Novo records found no cleared or granted device whose public documents describe an LLM or generative-AI component as part of its function [382][383][384][285].

### Is LLM-enabled software a device?

FDA's DHAC briefing materials use the term "GenAI-enabled device" for a medical device in which generative AI is integral to the primary function or directly enables the output. FDA also says that not every healthcare use of generative AI is a device [438][439]. For example, FDA noted that some uses, such as administrative documentation, fall outside device regulation. By contrast, it described diagnosis-generating tools and mental-health chatbots as potential device use cases [450].

The device boundary is set by existing guidance.

- **Clinical decision support (CDS).** FDA's final *Clinical Decision Support Software* guidance was issued in January 2026 [8][18]. Under it, CDS falls outside the device definition only if it meets all four section 520(o)(1)(E) criteria [19]:
  - it does not acquire, process, or analyze medical images, IVD signals, or physiological signals [20]
  - it displays, analyzes, or prints medical information [21]
  - it is intended to support or provide recommendations to a health care professional, which FDA reads as supporting rather than replacing or directing the HCP's judgment [22][23]
  - it lets the HCP independently review the basis for its recommendations [24]

  Functions that give recommendations to patients or caregivers rather than HCPs meet the device definition [25][26][27][28]. FDA also does not consider critical, time-sensitive decisions to satisfy criterion 4 [30]. The guidance does not set a separate rule for software just because it generates text. Function and intended use control the outcome [29].
- **General wellness and enforcement discretion.** The general wellness guidance was reissued January 6, 2026, replacing the 2019 edition. CDRH does not intend to review low-risk products intended only for general wellness, but that policy does not cover outputs used for clinical treatment decisions [314][315][320]. The September 28, 2022 device software functions and mobile medical applications guidance separates non-device functions from lower-risk device functions under enforcement discretion. Supplemental coaching for patients managing identified conditions is one example of the second group [321][322].

For an LLM product, the practical point is this. A clinician-facing summarizer or recommender that is transparent and reviewable may fit the non-device CDS criteria. A patient-facing chatbot that addresses a disease, or a tool that generates a differential diagnosis, is likely to be a device function.

### How FDA frames generative AI risk

FDA's DHAC briefing materials give working definitions:

- **Generative AI:** models that generate derived synthetic content (text, images, audio, video) resembling their training data [434].
- **Foundation models:** models applicable to multiple tasks, built on broad datasets and later adapted for particular uses [435][436].
- **Large language models:** models trained on large text corpora to generate natural-language responses for tasks such as summarization or question answering. A chatbot provides a conversational interface to an LLM [437].

FDA identified these regulatory challenges specific to generative AI:

- **Intended-use boundaries.** Models that handle many tasks and inputs make the limits of a device's intended use hard to confirm. That in turn complicates risk-based classification [438].
- **Output variability and hallucination.** FDA describes hallucinations as erroneous or false content that may look authentic. This is especially consequential where accurate clinical information is critical [440][441]. Free-text and multimodal inputs can also produce combinations never seen during development [442].
- **Foundation-model dependency.** Third-party foundation models are generally not built for a specific device. Monitoring may need to cover both the device model and the underlying foundation model, and site-specific implementations add complexity [443][444][445]. The committee also raised concerns about limited transparency into the training and tuning data of third-party models [446][447].

FDA's hypothetical use cases included:

- a foundation model supporting both tumor identification and a chatbot that discusses the image and the pathology report
- a chatbot that combines laboratory results, vital signs, and patient data to generate a differential diagnosis [444][442]

### DHAC recommendations on evidence and monitoring

The inaugural DHAC meeting (November 20 to 21, 2024) covered generative AI-enabled devices broadly. It applied a total product lifecycle (TPLC) lens to premarket performance evaluation, risk management, and postmarket monitoring [124][125][126][157].

- **Premarket:** the committee pointed to:
  - clear characterization of intended use, setting, users, and human-AI interaction
  - testing tailored to the intended population, including hallucination and error rates, repeatability, and stress testing
  - transparency and usability measures to limit overreliance [129][449]
- **Postmarket:** the committee pointed to:
  - scalable surveillance of real-world performance, drift, hallucinations, and adverse events
  - auditing across sites and populations
  - sharing performance findings with providers and patients [131]

The November 6, 2025 meeting focused on generative AI-enabled digital mental-health devices intended to diagnose or treat psychiatric conditions, not general-purpose chatbots [127][128]. FDA identified these risks [127][128]:

- confabulation
- biased or inappropriate content
- omitted medical information
- accuracy decline and data drift
- patient misinterpretation of outputs
- uncertainty about clinician oversight

FDA asked whether interactive, adaptive devices need different clinical evidence and trial designs [130]. These were discussion points, not new requirements [133].

### Applicable guidance for AI-enabled device submissions

| Document | Status and date | Relevance to LLM-enabled devices |
|---|---|---|
| *Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendations* | Draft, January 7, 2025 [240][224] | TPLC-based submission content (summarized below the table). It has no LLM-specific recommendations; transformers are mentioned as one model class to describe [251]. FDA specifically requested comment on whether the guidance adequately addresses generative AI [158][159]. |
| *Predetermined Change Control Plans for AI-Enabled Device Software Functions* | Final, August 18, 2025 [201][216] | A PCCP has three parts: Description of Modifications, Modification Protocol, and Impact Assessment [218]. FDA must establish it through a 510(k), De Novo, or PMA before changes are made under it [219]. This is the main mechanism for planned model updates without a new submission for each change [217]. |
| *Clinical Decision Support Software* | Final, January 2026 [8][18] | Sets the device/non-device boundary for clinician-facing recommendation software [19]. |
| *Considerations for the Use of AI to Support Regulatory Decision-Making for Drug and Biological Products* | Draft, January 7, 2025 [408][409] | This is not device guidance. It sets a seven-step, risk-based credibility framework tied to context of use for AI that produces data supporting drug safety, effectiveness, or quality decisions [416][417][419][421][389]. It excludes drug discovery and operational uses, such as drafting submissions, that do not affect patient safety, drug quality, or study reliability [414][415]. |

The draft device guidance asks for the following submission content:

- a device description that states AI use, inputs and outputs, and clinical workflow [241][221]
- model description and development methods [242][243]
- dataset provenance, representativeness, and test-data independence [244][245][246]
- subgroup performance validation and human factors evaluation [223][247][248]
- transparency information, with an optional model card [228][226]
- cybersecurity documentation and postmarket performance monitoring [249][250][222]

A search of FDA's guidance library found no current guidance that specifically addresses generative AI, LLMs, or foundation models in the AI sense [167][168][193][194].

### The August 2026 discussion paper

On August 18, 2026, FDA issued a discussion paper seeking feedback on generative AI-enabled medical devices. FDA described it as not a final regulatory framework [165][166]. The paper covers [288]:

- a possible two-axis risk-assessment framework
- a premarket "competency assessment" approach, inspired at a high level by physician training and evaluation, that combines non-clinical benchmarking with clinical confirmation
- several risk-proportionate postmarket monitoring approaches
- targeted questions on foundation models and agentic AI

Comments are due October 19, 2026 under docket FDA-2026-N-7874 [289]. The announcement does not say whether FDA has authorized any generative AI-enabled device [136]. Sponsors developing LLM-based clinical products should treat this paper as the clearest signal of where premarket expectations are heading.

### Authorization status: what the record shows

- **Generative AI overall:** FDA said in November 2024 that it had not authorized any generative AI-enabled device [352].
- **Mental health:** FDA said in November 2025 that no generative AI-enabled device had been authorized for mental-health conditions. Across all AI-enabled devices, FDA counted more than 1,200 authorizations [134][135][353].
- **AI-enabled devices generally:** FDA's public AI-enabled device list is large but not exhaustive. FDA reported 950 AI/ML-enabled authorizations as of August 2024 [155][156]. Most entries are imaging, signal-analysis, and planning tools.
- **510(k) and De Novo records:** searches for "large language model," "generative AI," "natural language generation," and chatbot terms found no device whose public summary describes an LLM or generative component as part of its function [382][383][384]. One near miss is the GI Genius module 510(k), K241887, cleared July 25, 2024. Its summary says the updated hardware can accelerate generative-AI inference models. It does not say the cleared lesion-detection function uses one [285][286][287].

This is evidence of absence in FDA's published statements and in 510(k) and De Novo summary text. It does not prove that no authorization exists, because summary documents may not describe the model architecture. As of the most recent FDA statements found, FDA has not publicly identified any LLM-based medical device it has authorized.

### FDA's internal use of LLMs

FDA uses LLMs internally, separately from regulating them. In June 2025 it launched Elsa, an agency-wide LLM-powered tool for reading, writing, summarizing, and review support [161][162]. FDA added optional agentic-AI capabilities in December 2025 [163] and announced Elsa 4.0 in May 2026 [164]. These are agency operational tools, not authorized medical devices.

## Sources

Numbered to match the [N] markers in the text.

- [8] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 6
- [18] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 4
- [19] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 7
- [20] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 6
- [21] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 6
- [22] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 6
- [23] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 13
- [24] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 6
- [25] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 4
- [26] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 5
- [27] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 7
- [28] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 13
- [29] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 13
- [30] [Clinical Decision Support Software: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/109618/download) — FDA Guidance, p. 14
- [124] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 6
- [125] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 5
- [126] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 3
- [127] [DHAC November 6, 2025 Summary Report](https://www.fda.gov/media/190450/download) — FDA Advisory Committee Meetings, p. 1
- [128] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 6
- [129] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 14
- [130] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 15
- [131] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 15
- [133] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 2
- [134] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 5
- [135] [DHAC November 6, 2025 Summary Report](https://www.fda.gov/media/190450/download) — FDA Advisory Committee Meetings, p. 1
- [136] [FDA Seeks Public Feedback to Inform Regulatory Approach for Generative AI-Enabled Medical Devices](https://www.fda.gov/news-events/press-announcements/fda-seeks-public-feedback-inform-regulatory-approach-generative-ai-enabled-medical-devices) — FDA Press Announcements, p. 1
- [155] [FDA Roundup: October 20, 2023](https://www.fda.gov/news-events/press-announcements/fda-roundup-october-20-2023) — FDA Press Announcements, p. 1
- [156] [FDA Roundup: August 9, 2024](https://www.fda.gov/news-events/press-announcements/fda-roundup-august-9-2024) — FDA Press Announcements, p. 1
- [157] [FDA Roundup: September 17, 2024](https://www.fda.gov/news-events/press-announcements/fda-roundup-september-17-2024) — FDA Press Announcements, p. 1
- [158] [FDA Issues Comprehensive Draft Guidance for Developers of Artificial Intelligence-Enabled Medical Devices](https://www.fda.gov/news-events/press-announcements/fda-issues-comprehensive-draft-guidance-developers-artificial-intelligence-enabled-medical-devices) — FDA Press Announcements, p. 1
- [159] [FDA Issues Comprehensive Draft Guidance for Developers of Artificial Intelligence-Enabled Medical Devices](https://www.fda.gov/news-events/press-announcements/fda-issues-comprehensive-draft-guidance-developers-artificial-intelligence-enabled-medical-devices) — FDA Press Announcements, p. 1
- [161] [FDA Announces Completion of First AI-Assisted Scientific Review Pilot and Aggressive Agency-Wide AI Rollout Timeline](https://www.fda.gov/news-events/press-announcements/fda-announces-completion-first-ai-assisted-scientific-review-pilot-and-aggressive-agency-wide-ai) — FDA Press Announcements, p. 1
- [162] [FDA Launches Agency-Wide AI Tool to Optimize Performance for the American People](https://www.fda.gov/news-events/press-announcements/fda-launches-agency-wide-ai-tool-optimize-performance-american-people) — FDA Press Announcements, p. 1
- [163] [FDA Expands Artificial Intelligence Capabilities with Agentic AI Deployment](https://www.fda.gov/news-events/press-announcements/fda-expands-artificial-intelligence-capabilities-agentic-ai-deployment) — FDA Press Announcements, p. 1
- [164] [FDA Expands AI Capabilities and Completes Data Platform Consolidation](https://www.fda.gov/news-events/press-announcements/fda-expands-ai-capabilities-and-completes-data-platform-consolidation) — FDA Press Announcements, p. 1
- [165] [FDA Seeks Public Feedback to Inform Regulatory Approach for Generative AI-Enabled Medical Devices](https://www.fda.gov/news-events/press-announcements/fda-seeks-public-feedback-inform-regulatory-approach-generative-ai-enabled-medical-devices) — FDA Press Announcements, p. 1
- [166] [FDA Seeks Public Feedback to Inform Regulatory Approach for Generative AI-Enabled Medical Devices](https://www.fda.gov/news-events/press-announcements/fda-seeks-public-feedback-inform-regulatory-approach-generative-ai-enabled-medical-devices) — FDA Press Announcements, p. 1
- [167] [Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device…](https://www.fda.gov/media/166704/download) — FDA Guidance, p. 6
- [168] [Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device…](https://www.fda.gov/media/166704/download) — FDA Guidance, p. 13
- [193] [Real-World Data: Assessing Electronic Health Records and Medical Claims Data To Support Regulatory Decision-Making for…](https://www.fda.gov/media/152503/download) — FDA Guidance, p. 14
- [194] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 23
- [201] [Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device…](https://www.fda.gov/media/166704/download) — FDA Guidance, p. 15
- [216] [Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device…](https://www.fda.gov/media/166704/download) — FDA Guidance, p. 13
- [217] [Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device…](https://www.fda.gov/media/166704/download) — FDA Guidance, p. 14
- [218] [Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device…](https://www.fda.gov/media/166704/download) — FDA Guidance, p. 15
- [219] [Marketing Submission Recommendations for a Predetermined Change Control Plan for Artificial Intelligence-Enabled Device…](https://www.fda.gov/media/166704/download) — FDA Guidance, p. 15
- [221] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 11
- [222] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 35
- [223] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 29
- [224] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 4
- [226] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 41
- [228] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 8
- [240] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 4
- [241] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 11
- [242] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 27
- [243] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 29
- [244] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 41
- [245] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 8
- [246] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 27
- [247] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 30
- [248] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 30
- [249] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 43
- [250] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 37
- [251] [Artificial Intelligence-Enabled Device Software Functions: Lifecycle Management and Marketing Submission Recommendation…](https://www.fda.gov/media/184856/download) — FDA Guidance, p. 41
- [285] [GI Genius Module 100 (GGM100.US); GI Genius Module 200 (GGM200.US); ColonPRO 4.0 (CPRO40.US); GI Genius Module 300… · Cosmo Artificial Intelligence - Ai, Ltd. (K241887)](https://www.accessdata.fda.gov/cdrh_docs/pdf24/K241887.pdf) — FDA 510(k) or De Novo, p. 6
- [286] [GI Genius Module 100 (GGM100.US); GI Genius Module 200 (GGM200.US); ColonPRO 4.0 (CPRO40.US); GI Genius Module 300… · Cosmo Artificial Intelligence - Ai, Ltd. (K241887)](https://www.accessdata.fda.gov/cdrh_docs/pdf24/K241887.pdf) — FDA 510(k) or De Novo, p. 6
- [287] [GI Genius Module 100 (GGM100.US); GI Genius Module 200 (GGM200.US); ColonPRO 4.0 (CPRO40.US); GI Genius Module 300… · Cosmo Artificial Intelligence - Ai, Ltd. (K241887)](https://www.accessdata.fda.gov/cdrh_docs/pdf24/K241887.pdf) — FDA 510(k) or De Novo, p. 6
- [288] [FDA Seeks Public Feedback to Inform Regulatory Approach for Generative AI-Enabled Medical Devices](https://www.fda.gov/news-events/press-announcements/fda-seeks-public-feedback-inform-regulatory-approach-generative-ai-enabled-medical-devices) — FDA Press Announcements, p. 1
- [289] [FDA Seeks Public Feedback to Inform Regulatory Approach for Generative AI-Enabled Medical Devices](https://www.fda.gov/news-events/press-announcements/fda-seeks-public-feedback-inform-regulatory-approach-generative-ai-enabled-medical-devices) — FDA Press Announcements, p. 1
- [314] [General Wellness: Policy for Low Risk Devices: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/90652/download) — FDA Guidance, p. 1
- [315] [General Wellness: Policy for Low Risk Devices: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/90652/download) — FDA Guidance, p. 5
- [320] [General Wellness: Policy for Low Risk Devices: Guidance for Industry and Food and Drug Administration Staff](https://www.fda.gov/media/90652/download) — FDA Guidance, p. 9
- [321] [Policy for Device Software Functions and Mobile Medical Applications: Guidance for Industry and Food and Drug Administr…](https://www.fda.gov/media/80958/download) — FDA Guidance, p. 5
- [322] [Policy for Device Software Functions and Mobile Medical Applications: Guidance for Industry and Food and Drug Administr…](https://www.fda.gov/media/80958/download) — FDA Guidance, p. 17
- [352] [DHAC Nov. 21, 2024 Transcript](https://www.fda.gov/media/185101/download) — FDA Advisory Committee Meetings, p. 78
- [353] [DHAC November 6, 2025 Transcript](https://www.fda.gov/media/190451/download) — FDA Advisory Committee Meetings, p. 15
- [382] [AI-Rad Companion Brain MR · Siemens Healthcare GmbH (K253057)](https://www.accessdata.fda.gov/cdrh_docs/pdf25/K253057.pdf) — FDA 510(k) or De Novo, p. 12
- [383] [CDM Insights · Oxford Brain Diagnostics, Ltd. (K240680)](https://www.accessdata.fda.gov/cdrh_docs/pdf24/K240680.pdf) — FDA 510(k) or De Novo, p. 10
- [384] [CDM Insights · Oxford Brain Diagnostics, Ltd. (K240680)](https://www.accessdata.fda.gov/cdrh_docs/pdf24/K240680.pdf) — FDA 510(k) or De Novo, p. 10
- [389] [Considerations for the Use of Artificial Intelligence To Support Regulatory Decision-Making for Drug and Biological Pro…](https://www.fda.gov/media/184830/download) — FDA Guidance, p. 18
- [408] [Considerations for the Use of Artificial Intelligence To Support Regulatory Decision-Making for Drug and Biological Pro…](https://www.fda.gov/media/184830/download) — FDA Guidance, p. 1
- [409] [Considerations for the Use of Artificial Intelligence To Support Regulatory Decision-Making for Drug and Biological Pro…](https://www.fda.gov/media/184830/download) — FDA Guidance, p. 8
- [414] [Considerations for the Use of Artificial Intelligence To Support Regulatory Decision-Making for Drug and Biological Pro…](https://www.fda.gov/media/184830/download) — FDA Guidance, p. 6
- [415] [Considerations for the Use of Artificial Intelligence To Support Regulatory Decision-Making for Drug and Biological Pro…](https://www.fda.gov/media/184830/download) — FDA Guidance, p. 5
- [416] [Considerations for the Use of Artificial Intelligence To Support Regulatory Decision-Making for Drug and Biological Pro…](https://www.fda.gov/media/184830/download) — FDA Guidance, p. 8
- [417] [Considerations for the Use of Artificial Intelligence To Support Regulatory Decision-Making for Drug and Biological Pro…](https://www.fda.gov/media/184830/download) — FDA Guidance, p. 8
- [419] [Considerations for the Use of Artificial Intelligence To Support Regulatory Decision-Making for Drug and Biological Pro…](https://www.fda.gov/media/184830/download) — FDA Guidance, p. 8
- [421] [Considerations for the Use of Artificial Intelligence To Support Regulatory Decision-Making for Drug and Biological Pro…](https://www.fda.gov/media/184830/download) — FDA Guidance, p. 8
- [434] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 4
- [435] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 4
- [436] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 6
- [437] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 29
- [438] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 5
- [439] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 6
- [440] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 4
- [441] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 5
- [442] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 6
- [443] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 4
- [444] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 5
- [445] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 20
- [446] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 20
- [447] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 22
- [449] [DHAC November 6, 2025 Executive Summary](https://www.fda.gov/media/189391/download) — FDA Advisory Committee Meetings, p. 20
- [450] [DHAC Nov.20-21, 2024 FDA Executive Summary](https://www.fda.gov/media/182871/download) — FDA Advisory Committee Meetings, p. 6

---

Rhizome AI researched and wrote this article from health-authority documents. Ask it your own regulatory question at https://rhizomeai.com. More articles: https://rhizomeai.com/articles.md. Guide for agents: https://rhizomeai.com/llms.txt