# Cell Therapy Release Specifications Accepted by FDA in BLA Reviews

How FDA has handled viability, cell dose, identity and purity release specifications for licensed cell therapies, drawn from CBER BLA review documents.

Published by Rhizome AI on 2026-10-16. Web version: https://rhizomeai.com/articles/cell-therapy-release-specifications-in-fda-bla

## Cell therapy release specifications in FDA BLA reviews: viability, cell dose, identity and purity

Most licensed cell therapies have a release specification table in the public CBER review record. The numeric acceptance criteria for viability, purity and potency are usually redacted as (b)(4) in those tables. FDA does disclose two other things fairly consistently: the definition of strength (dose), with numeric ranges, and the pass/fail safety criteria (sterility, mycoplasma). Unredacted numeric viability limits appear mainly for non-genetically modified products, such as allogeneic islets and licensed cord blood. The sections below set out what the review memos and summary basis documents show, attribute by attribute.

### Key findings

- **Dose is defined by the active cell population, not total cells.** For CAR T products, strength is expressed as CAR-positive viable T cells. Kymriah calculates it as % CAR expression × viable-cell concentration × volume per dose ÷ 100 [292]. Yescarta and Tecartus use viable cell count combined with anti-CD19 CAR expression [255][76].
- **Viability limits are almost always redacted, in genetically modified and non-modified products alike.** Kymriah, Yescarta, Tecartus, Abecma, Carvykti, Aucatzyl, Amtagvi, Tecelra, Casgevy and Ryoncil all list a viability test with a (b)(4) criterion [291][254][75][16][224][43][335][473][124][483].
- **Where viability is disclosed, values sit at or above the 70% guidance floor.** Lantidra requires ≥70% viable islets [60]. The Bloodworks cord blood BLA requires >85% viable nucleated cells [541]. FDA's 2008 somatic cell therapy CMC guidance describes 70% as generally the minimum acceptable viability specification, and a lower limit must be supported with data [613][614].
- **Identity usually pairs a transgene test with a cell-type test.** Examples include CAR qPCR "positive" for Kymriah [296], CD34 expression plus on-target editing frequency for Casgevy [118][119], and a CD166/CD105/CD45 panel for Ryoncil [485]. This matches the CAR T guidance, which says HLA typing alone is insufficient [561].
- **Safety criteria are the most transparent part of the table.** Sterility "no growth" or "negative" and mycoplasma "negative" or "not detected" appear unredacted across products. Endotoxin limits are almost always redacted. Lantidra is the exception, with a disclosed ≤5 EU/kg/hour limit [68].

### Strength (cell dose) specifications

FDA treats strength as a calculated release attribute. The calculation is tied to the potency-relevant cell fraction and, for weight-based products, to patient weight.

| Product | Dose/strength definition | Dose range or release criterion |
|---|---|---|
| Kymriah (tisagenlecleucel) | Transduced viable T cells | 0.2 to 5.0 × 10⁶/kg (≤50 kg); 0.1 to 2.5 × 10⁸ total (>50 kg) [292] |
| Yescarta (axicabtagene ciloleucel) | Viable CAR-positive T cells per kg | Release range (b)(4). Labeled target dose 2.0 × 10⁶/kg, maximum 2 × 10⁸ CAR T cells [255][256][257] |
| Tecartus (brexucabtagene autoleucel) | CAR-positive viable T cells from weight, viable concentration and CAR expression | Original BLA (MCL): 2 × 10⁶ CAR T cells/kg, capped at 2 × 10⁸ for patients ≥100 kg [76][82][77]. The later B-ALL indication uses 1 × 10⁶/kg, capped at 1 × 10⁸ |
| Breyanzi (lisocabtagene maraleucel) | CAR-positive viable T cells, CD8 and CD4 components dosed 1:1 | 50 to 110 × 10⁶ per dose in the original BLA review; release table strength criterion (b)(4) [442][444][445] |
| Abecma (idecabtagene vicleucel) | CAR-positive T cells from concentration, % CAR+ and bag volume | 300 to 460 × 10⁶ CAR-positive T cells [19][20][21] |
| Carvykti (ciltacabtagene autoleucel) | CAR⁺ viable T cells, per kg (≤100 kg) or total (>100 kg) | 0.5 to 1.0 × 10⁶/kg; 0.5 to 1.0 × 10⁸ total [226][219] |
| Aucatzyl | CD19 CAR-positive viable T cells | 410 × 10⁶ ± 25% [44][45] |
| Tecelra (afamitresgene autoleucel) | MAGE-A4 TCR-positive T cells, calculated | ≥2,680.0 × 10⁶ cells; bags released are controlled to stay within the dose range [470][471] |
| Amtagvi (lifileucel) | Total viable cells | 7.5 × 10⁹ to 72 × 10⁹ viable cells per dose [332][333][334] |
| Lantidra (donislecel) | Islet equivalents (EIN) per kg | ≥5,000 EIN/kg initial; ≥4,500 EIN/kg subsequent transplant [61] |
| Ryoncil (remestemcel-L) | Cell concentration | ≥6.68 × 10⁶ cells/mL [484] |
| Skysona | Total cell concentration | 4 to 30 × 10⁶ total cells/mL [462] |
| MACI | Cell density on collagen membrane | 500,000 to 1,000,000 cells/cm² [97] |
| Cord blood (Hemacord, Bloodworks) | Total nucleated cells and viable CD34+ per unit | TNC >5.0 × 10⁸/unit; viable CD34+ ≥1.25 × 10⁶/unit (Hemacord) [535][537]; >1.25 × 10⁶ viable CD34+ (Bloodworks) [541][542] |

Two points matter when you benchmark against these values:

1. **Labeled dose and release criterion are not always the same thing.** For Casgevy, the administered dose is at least 3 × 10⁶ viable CD34+ cells/kg, and one dose can be supplied from multiple lots. The release criterion for viable cell count itself is redacted [128][129][123]. For Lyfgenia, the approval letter gives 1.7 to 20 × 10⁶ total cells/mL, including 1.4 to 20 × 10⁶ CD34+ cells/mL, and a minimum dose of 3.0 × 10⁶ CD34+ cells/kg. These values cannot be mapped to the redacted release-table rows [384][380].
2. **Total cell count is often reported without a specification.** Kymriah's total cell count is "report cells/mL" with no limit. The calculated viable-cell number does carry a (redacted) limit [293][294][295].

### Viability

Every BLA reviewed includes a viability test. The disclosed criteria are limited:

- **Lantidra:** ≥70% viable islets by SYTO 13 Green/ethidium bromide stain with microscopic evaluation [60].
- **Bloodworks HPC, Cord Blood:** >85% viable nucleated cells [541]. Hemacord lists CD45+ viability with a redacted criterion [536]. MD Anderson's cord blood BLA redacts its nucleated-cell and CD34+ viability criteria [546].
- **CAR T and gene-modified HSC products:** Viability is listed, sometimes as a potency-related attribute (Yescarta, Tecartus), but the criterion is (b)(4) [254][259][75]. For Breyanzi, FDA describes a finalized viability criterion but does not state the cutoff [440].
- **Tissue-engineered products:** Stratagraft has a viability assay with a redacted method and criterion [506][507].

The guidance baseline is consistent. The 1998 somatic cell and gene therapy guidance asks for viability to be quantified with an established lower acceptable limit [612]. Both the 2008 somatic cell CMC guidance and the 2020 gene therapy CMC guidance (for ex vivo modified cells) recommend at least 70% unless data support a lower limit [613][615].

### Identity

Identity tests in the approved tables fall into three patterns:

- **Transgene detection:** Kymriah uses CAR qPCR with a "positive" criterion [296]. Yescarta and Tecartus detect scFv heavy-chain variable region, linker and CD28 sequences, with redacted criteria [258][78]. Carvykti is "positive for detection of CAR transgene" in the summary basis [227]. Abecma requires anti-BCMA02 CAR-positive T cells to be detected [14]. Aucatzyl requires CD19 CAR expression "detected" [46].
- **Cell phenotype or edit:** Casgevy requires "positive" for CD34 expression by flow cytometry and for on-target editing frequency by TIDE. The positivity thresholds are redacted [118][119]. Ryoncil tests CD166⁺, CD105⁺ and CD45⁺ cell content with redacted criteria [485]; MSCs are conventionally CD166- and CD105-positive and CD45-negative. MACI requires "positive for chondrocyte cells" [92]. For Zynteglo, βA-T87Q-globin protein expression serves as an identity test [411].
- **Morphological or donor-matching identity:** Lantidra combines packed tissue volume (≤10 cc) with dithizone-stained islet morphology [62][63]. Cord blood products rely on confirmatory HLA typing of the attached segment, ABO/Rh, and label matching [539][543].

FDA's January 2024 CAR T guidance recommends testing at every development phase for both the transgene and the intended cellular composition, and states that HLA typing alone is insufficient [553][561].

### Purity and impurities

Purity attributes are listed but almost always redacted:

- **Kymriah:** % viable T cells and % viable CD19+ B cells (flow cytometry, CD45/CD19), plus residual CD3/CD28 beads by microscopy, all (b)(4) [297][298][299].
- **Abecma:** T-cell percentage, (b)(4) [15]. **Carvykti:** phenotype measures including % NK cells, (b)(4) [229].
- **Casgevy:** CD34 purity by flow cytometry, (b)(4). The review treats nonviable and non-CD34+ cells as product-related impurities and Cas9 residuals as process-related impurities evaluated in characterization, not as listed release tests [120][104].
- **Ryoncil:** Residual bovine serum albumin and residual trypsin, (b)(4) [487].
- **Yescarta:** Gentamicin as a process residual, (b)(4) [260].
- **Amtagvi:** The review states that residual viable melanoma cells are controlled by release testing. Neither the method nor the criterion is disclosed [346].

Vector copy number appears as a release test in the lentiviral HSC products, classified under potency/strength rather than purity. Zynteglo, Lyfgenia and Skysona each list VCN by qPCR with redacted criteria [408][375][462]. The CAR T guidance asks sponsors to justify the VCN release criterion through a risk assessment [567]. Several CAR T release tables (Yescarta, Tecartus, Abecma, Carvykti) do not list a separately named VCN specification [240][70][3][201].

### Potency

Potency assays are product-specific, and their criteria are typically redacted:

- **Kymriah:** CAR expression by flow cytometry and IFN-γ release against CD19-expressing targets [301][302][303].
- **Tecelra:** Cytotoxic activity by a flow cytometry-based assay [475].
- **Lantidra:** One of the few fully disclosed potency criteria: glucose-stimulated insulin secretion ratio (high:low glucose) ≥1 by ELISA [64].
- **Ryoncil:** IL-2Rα inhibition, (b)(4) [486].
- **Zynteglo:** VCN, % LVV+ cells, colony-forming cells and βA-T87Q-globin expression, all (b)(4) [408][409][410].
- **Casgevy:** On-target editing frequency by TIDE plus two redacted assays [121][122].

FDA's 2011 potency guidance does not prescribe a particular assay or acceptance criterion [616]. The CAR T guidance suggests orthogonal measures such as target-cell killing, transduction efficiency and cytokine secretion [565].

### Safety: sterility, mycoplasma, endotoxin and replication-competent virus

| Product | Sterility | Mycoplasma | Endotoxin | RCR/RCL |
|---|---|---|---|---|
| Kymriah | Negative [306] | Negative [306] | (b)(4) [306] | No detection of RCL [304][305] |
| Yescarta | No growth [264] | Negative [262] | (b)(4) [261] | Negative (two tests) [262][265][266] |
| Tecartus | Negative [82] | Negative [81] | (b)(4) [83] | Not listed separately [70] |
| Breyanzi | No growth [450] | Not detected [451] | (b)(4) [452] | Not identified in release table [453][454] |
| Abecma | No growth [24] | None detected [23] | (b)(4) [20] | Excluded from DP release testing [25][26] |
| Carvykti | No growth [232] | Not detected [234] | (b)(4) [233] | (b)(4) [235] |
| Aucatzyl | No growth [50] | Negative [52] | (b)(4) [51] | (b)(4) [53] |
| Amtagvi | No growth [342] | Not detected [341] | (b)(4) [340] | Not applicable/not listed [307] |
| Casgevy | No growth [125] | Negative [126] | (b)(4) [127] | Not listed [104] |
| Lyfgenia | No growth [381] | None detected [383] | (b)(4) [382] | Not identified [349] |
| Lantidra | No growth in 14 days; fungal no growth in 28 days [65][66] | Not stated [55] | ≤5 EU/kg/hour [68] | Not applicable |
| Ryoncil | Negative (in-process and final) [488] | Negative [488] | (b)(4) [489] | Not applicable |

The Abecma review is notable because RCL testing was excluded from drug product release instead of being assigned a criterion [25][26]. Kymriah sets its RCL qPCR limit from the assay LOQ [304][305].

### Products released before final sterility results

Several short-shelf-life products are released for infusion before the final compendial sterility result is available:

- **Provenge:** Release uses Gram stain, endotoxin content and a 2-day in-process sterility test. The 7-day sterility result is not available at infusion. Each dose contains at least 50 million autologous CD54⁺ cells activated with PAP-GM-CSF in 250 mL [199][195][196].
- **MACI:** Final sterility testing is done on conditioned medium, and the result is not available before administration [94][19].
- **Laviv:** Requires a passing bulk drug substance sterility culture and a negative final-product Gram stain before shipping. Final sterility results can take up to 14 days [161].
- **Gintuit:** Requires passing in-process microbial tests before shipping, and the final sterility result can take up to 14 days [404].

### Out-of-specification lots

In the Kymriah review, FDA describes lot release as the final quality check before commercial release [583][584]. FDA raised plans for disposing of occasional OOS lots in the commercial setting as an issue to discuss with Novartis [587]. The available review text does not state the resulting commercial policy [575]. One clinical batch that failed a color specification was released under an exception because its autologous erythrocytes were not considered a safety risk [588].

### Practical implications for sponsors

- **Define strength around the active population.** Every approved CAR T and TCR-T product ties strength to the transgene-positive viable fraction [292][255][19][470]. Agree on the calculation (inputs, weight bands, caps) early, because it becomes a release specification.
- **Expect to justify viability below 70%.** The guidance floor is 70% [613][615]. The few disclosed BLA values are at or above it [60][541].
- **Build identity around two orthogonal elements**, the transgene or edit plus the cell phenotype [561][118][119].
- **Plan numerical criteria before pivotal studies.** FDA's 2026 CMC flexibilities guidance accepts permissive early-phase criteria. It also states that release tests must have predefined acceptance criteria before Phase 2 or 3, and recommends discussing when to set numerical criteria for studies intended to provide substantial evidence of effectiveness [592].
- **Benchmarking has limits.** Because most numeric viability, purity and potency limits are redacted, public BLA records are most useful for test selection, dose definitions and safety criteria, not for copying numeric limits.

## Sources

Numbered to match the [N] markers in the text.

- [3] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20Final%20Review%20-%20ABECMA.pdf) — FDA Biologics, p. 83
- [14] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20Final%20Review%20-%20ABECMA.pdf) — FDA Biologics, p. 83
- [15] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20Final%20Review%20-%20ABECMA.pdf) — FDA Biologics, p. 83
- [16] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20Final%20Review%20-%20ABECMA.pdf) — FDA Biologics, p. 83
- [19] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20Final%20Review%20-%20ABECMA.pdf) — FDA Biologics, p. 83
- [20] [ABECMA, idecabtagene vicleucel · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147627/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [21] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20BLA%20Review%20Memo%20-%20ABECMA.pdf) — FDA Biologics, p. 44
- [23] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20Final%20Review%20-%20ABECMA.pdf) — FDA Biologics, p. 83
- [24] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20Final%20Review%20-%20ABECMA.pdf) — FDA Biologics, p. 83
- [25] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20Final%20Review%20-%20ABECMA.pdf) — FDA Biologics, p. 90
- [26] [ABECMA (idecabtagene vicleucel) · Celgene Corporation, a Bristol-Myers Squibb Company (BLA125736)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/147781/download%3Fattachment.zip/CMC%20BLA%20Review%20Memo%20-%20ABECMA.pdf) — FDA Biologics, p. 44
- [43] [AUCATZYL (obecabtagene autoleucel) · Autolus Limited (BLA125813)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/183887/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [44] [AUCATZYL (obecabtagene autoleucel) · Autolus Limited (BLA125813)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/183887/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [45] [AUCATZYL (obecabtagene autoleucel) · Autolus Limited (BLA125813)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/184205/download%3Fattachment.zip/CMC%20BLA%20Review%20Memorandum%20-%20AUCATZYL.pdf) — FDA Biologics, p. 46
- [46] [AUCATZYL (obecabtagene autoleucel) · Autolus Limited (BLA125813)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/183887/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [50] [AUCATZYL (obecabtagene autoleucel) · Autolus Limited (BLA125813)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/183887/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [51] [AUCATZYL (obecabtagene autoleucel) · Autolus Limited (BLA125813)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/183887/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [52] [AUCATZYL (obecabtagene autoleucel) · Autolus Limited (BLA125813)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/183887/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [53] [AUCATZYL (obecabtagene autoleucel) · Autolus Limited (BLA125813)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/183887/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [55] [LANTIDRA (donislecel-jujn) · CellTrans Inc. (BLA125734)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/170457/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [60] [LANTIDRA (donislecel-jujn) · CellTrans Inc. (BLA125734)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/170457/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [61] [LANTIDRA (donislecel-jujn) · CellTrans Inc. (BLA125734)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/170457/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [62] [LANTIDRA (donislecel-jujn) · CellTrans Inc. (BLA125734)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/170457/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [63] [LANTIDRA (donislecel-jujn) · CellTrans Inc. (BLA125734)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/170457/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [64] [LANTIDRA (donislecel-jujn) · CellTrans Inc. (BLA125734)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/170457/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [65] [LANTIDRA (donislecel-jujn) · CellTrans Inc. (BLA125734)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/170457/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [66] [LANTIDRA (donislecel-jujn) · CellTrans Inc. (BLA125734)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/170457/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [68] [LANTIDRA (donislecel-jujn) · CellTrans Inc. (BLA125734)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/170457/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [70] [TECARTUS (brexucabtagene autoleucel) · Kite Pharmaceuticals, Inc. (BLA125703)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/141093/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [75] [TECARTUS (brexucabtagene autoleucel) · Kite Pharmaceuticals, Inc. (BLA125703)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/141093/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [76] [TECARTUS (brexucabtagene autoleucel) · Kite Pharmaceuticals, Inc. (BLA125703)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/141093/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [77] [TECARTUS (brexucabtagene autoleucel) · Kite Pharmaceuticals, Inc. (BLA125703)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/141093/download%3Fattachment.pdf) — FDA Biologics, p. 6
- [78] [TECARTUS (brexucabtagene autoleucel) · Kite Pharmaceuticals, Inc. (BLA125703)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/141093/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [81] [TECARTUS (brexucabtagene autoleucel) · Kite Pharmaceuticals, Inc. (BLA125703)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/141093/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [82] [TECARTUS (brexucabtagene autoleucel) · Kite Pharmaceuticals, Inc. (BLA125703)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/141093/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [83] [TECARTUS (brexucabtagene autoleucel) · Kite Pharmaceuticals, Inc. (BLA125703)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/141093/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [92] [MACI, autologous cultured chondrocytes on porcine collagen membrane · Vericel Corporation (BLA125603)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/106194/download%3Fattachment.zip/CMC%20Review,%20December%2012,%202016%20-%20MACI.pdf) — FDA Biologics, p. 29
- [94] [MACI, autologous cultured chondrocytes on porcine collagen membrane · Vericel Corporation (BLA125603)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/106194/download%3Fattachment.zip/CMC%20Review,%20December%2012,%202016%20-%20MACI.pdf) — FDA Biologics, p. 29
- [97] [MACI, autologous cultured chondrocytes on porcine collagen membrane · Vericel Corporation (BLA125603)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/106194/download%3Fattachment.zip/Clinical%20Review,%20November%207,%202016%20-%20MACI.pdf) — FDA Biologics, p. 2
- [104] [CASGEVY (exagamglogene autotemcel; exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175842/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [118] [CASGEVY (exagamglogene autotemcel, exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176033/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CASGEVY.pdf) — FDA Biologics, p. 61
- [119] [CASGEVY (exagamglogene autotemcel, exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176033/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CASGEVY.pdf) — FDA Biologics, p. 61
- [120] [CASGEVY (exagamglogene autotemcel; exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175842/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [121] [CASGEVY (exagamglogene autotemcel; exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175842/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [122] [CASGEVY (exagamglogene autotemcel; exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175842/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [123] [CASGEVY (exagamglogene autotemcel; exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175842/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [124] [CASGEVY (exagamglogene autotemcel; exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175842/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [125] [CASGEVY (exagamglogene autotemcel; exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175842/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [126] [CASGEVY (exagamglogene autotemcel; exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175842/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [127] [CASGEVY (exagamglogene autotemcel; exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175842/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [128] [CASGEVY, exagamglogene autotemcel (exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175235/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CASGEVY.pdf) — FDA Biologics, p. 19
- [129] [CASGEVY, exagamglogene autotemcel (exa-cel) · Vertex Pharmaceuticals Incorporated (BLA125787, BLA125785)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175235/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CASGEVY.pdf) — FDA Biologics, p. 19
- [161] [LAVIV (azficel-T) · Fibrocell Technologies, Inc. (BLA125348)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/106221/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [195] [PROVENGE (sipuleucel-T), sipuleucel-T, PROVENGE · Dendreon Corporation (BLA125197)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/106057/download%3Fattachment.pdf) — FDA Biologics, p. 3
- [196] [PROVENGE (sipuleucel-T), sipuleucel-T, PROVENGE · Dendreon Corporation (BLA125197)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/106057/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [199] [PROVENGE (sipuleucel-T), sipuleucel-T, PROVENGE · Dendreon Corporation (BLA125197)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/106057/download%3Fattachment.pdf) — FDA Biologics, p. 6
- [201] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/157051/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CARVYKTI.pdf) — FDA Biologics, p. 30
- [219] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/157051/download%3Fattachment.zip/Clinical%20Review%20and%20Evaluation,%20October%2029,%202021%20-%20CARVYKTI.pdf) — FDA Biologics, p. 91
- [224] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/157051/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CARVYKTI.pdf) — FDA Biologics, p. 30
- [226] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/157051/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CARVYKTI.pdf) — FDA Biologics, p. 30
- [227] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/156999/download%3Fattachment.pdf) — FDA Biologics, p. 9
- [229] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/157051/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CARVYKTI.pdf) — FDA Biologics, p. 30
- [232] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/157051/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CARVYKTI.pdf) — FDA Biologics, p. 30
- [233] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/157051/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CARVYKTI.pdf) — FDA Biologics, p. 30
- [234] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/157051/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CARVYKTI.pdf) — FDA Biologics, p. 30
- [235] [CARVYKTI, ciltacabtagene autoleucel · Janssen Biotech, Inc. (BLA125746)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/157051/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20CARVYKTI.pdf) — FDA Biologics, p. 30
- [240] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [254] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [255] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [256] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 37
- [257] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 31
- [258] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [259] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [260] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [261] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [262] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/108788/download%3Fattachment.pdf) — FDA Biologics, p. 6
- [264] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [265] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [266] [YESCARTA (axicabtagene ciloleucel) · Kite Pharma Inc. (BLA125643)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/109140/download%3Fattachment.zip/CMC%20Review%20-%20YESCARTA.pdf) — FDA Biologics, p. 30
- [291] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 27
- [292] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 24
- [293] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 24
- [294] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 24
- [295] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 28
- [296] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 23
- [297] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 23
- [298] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 27
- [299] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 27
- [301] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 28
- [302] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 23
- [303] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 24
- [304] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 31
- [305] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 24
- [306] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 24
- [307] [AMTAGVI (lifileucel) · Iovance Biotherapeutics, Inc. (BLA125773)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176893/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [332] [AMTAGVI (lifileucel) · Iovance Biotherapeutics, Inc. (BLA125773)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176893/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [333] [AMTAGVI (lifileucel) · Iovance Biotherapeutics, Inc. (BLA125773)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176893/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20AMTAGVI.pdf) — FDA Biologics, p. 36
- [334] [AMTAGVI (lifileucel) · Iovance Biotherapeutics, Inc. (BLA125773)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176893/download%3Fattachment.zip/BLA%20Clinical%20Review%20and%20Evaluation%20-%20AMTAGVI.pdf) — FDA Biologics, p. 14
- [335] [AMTAGVI (lifileucel) · Iovance Biotherapeutics, Inc. (BLA125773)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176893/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [340] [AMTAGVI (lifileucel) · Iovance Biotherapeutics, Inc. (BLA125773)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176893/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [341] [AMTAGVI (lifileucel) · Iovance Biotherapeutics, Inc. (BLA125773)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176893/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [342] [AMTAGVI (lifileucel) · Iovance Biotherapeutics, Inc. (BLA125773)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176893/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [346] [AMTAGVI (lifileucel) · Iovance Biotherapeutics, Inc. (BLA125773)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/176893/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [349] [LYFGENIA, lovotibeglogene autotemcel (lovo-cel) · bluebird bio, Inc.](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175250/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [375] [LYFGENIA, lovotibeglogene autotemcel (lovo-cel) · bluebird bio, Inc.](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175250/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [380] [LYFGENIA, lovotibeglogene autotemcel (lovo-cel) · bluebird bio, Inc.](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175250/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [381] [LYFGENIA, lovotibeglogene autotemcel (lovo-cel) · bluebird bio, Inc.](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175250/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [382] [LYFGENIA, lovotibeglogene autotemcel (lovo-cel) · bluebird bio, Inc.](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175250/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [383] [LYFGENIA, lovotibeglogene autotemcel (lovo-cel) · bluebird bio, Inc.](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/175250/download%3Fattachment.pdf) — FDA Biologics, p. 8
- [384] [LYFGENIA, lovotibeglogene autotemcel (lovo-cel) · bluebird bio, Inc.](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/174617/download%3Fattachment.pdf) — FDA Biologics, p. 2
- [404] [GINTUIT, Allogeneic Cultured Keratinocytes and Fibroblasts in Bovine Collagen · Organogenesis Incorporated (BLA125400)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/106952/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [408] [ZYNTEGLO (betibeglogene autotemcel) · bluebird bio Inc. (BLA125717)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/160991/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20ZYNTEGLO.pdf) — FDA Biologics, p. 43
- [409] [ZYNTEGLO (betibeglogene autotemcel) · bluebird bio Inc. (BLA125717)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/160991/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20ZYNTEGLO.pdf) — FDA Biologics, p. 43
- [410] [ZYNTEGLO (betibeglogene autotemcel) · bluebird bio Inc. (BLA125717)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/160991/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20ZYNTEGLO.pdf) — FDA Biologics, p. 43
- [411] [ZYNTEGLO (betibeglogene autotemcel) · bluebird bio Inc. (BLA125717)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/160991/download%3Fattachment.zip/CBER%20CMC%20BLA%20Review%20Memo%20-%20ZYNTEGLO.pdf) — FDA Biologics, p. 43
- [440] [BREYANZI (lisocabtagene maraleucel) · Juno Therapeutics, Inc., a Bristol-Myers Squibb company (BLA125714)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/146425/download%3Fattachment.zip/Late-Cycle%20Meeting%20Summary,%20September%202,%202020%20-%20BREYANZI.pdf) — FDA Biologics, p. 5
- [442] [BREYANZI (lisocabtagene maraleucel) · Juno Therapeutics, Inc., a Bristol-Myers Squibb company (BLA125714)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/146425/download%3Fattachment.zip/CMC%20BLA%20Review%20Memo%20--%20BREYANZI.pdf) — FDA Biologics, p. 70
- [444] [BREYANZI (lisocabtagene maraleucel) · Juno Therapeutics, Inc., a Bristol-Myers Squibb company (BLA125714)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/146425/download%3Fattachment.zip/CMC%20BLA%20Review%20Memo%20-%20BREYANZI.pdf) — FDA Biologics, p. 108
- [445] [BREYANZI (lisocabtagene maraleucel) · Juno Therapeutics, Inc., a Bristol-Myers Squibb company (BLA125714)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/146425/download%3Fattachment.zip/CMC%20BLA%20Review%20Memo%20-%20BREYANZI.pdf) — FDA Biologics, p. 118
- [450] [BREYANZI (lisocabtagene maraleucel) · Juno Therapeutics, Inc., a Bristol-Myers Squibb company (BLA125714)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/146425/download%3Fattachment.zip/CMC%20BLA%20Review%20Memo%20--%20BREYANZI.pdf) — FDA Biologics, p. 70
- [451] [BREYANZI (lisocabtagene maraleucel) · Juno Therapeutics, Inc., a Bristol-Myers Squibb company (BLA125714)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/146425/download%3Fattachment.zip/CMC%20BLA%20Review%20Memo%20--%20BREYANZI.pdf) — FDA Biologics, p. 70
- [452] [BREYANZI (lisocabtagene maraleucel) · Juno Therapeutics, Inc., a Bristol-Myers Squibb company (BLA125714)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/146425/download%3Fattachment.zip/CMC%20BLA%20Review%20Memo%20--%20BREYANZI.pdf) — FDA Biologics, p. 70
- [453] [BREYANZI (lisocabtagene maraleucel) · Juno Therapeutics, Inc., a Bristol-Myers Squibb company (BLA125714)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/146425/download%3Fattachment.zip/CMC%20BLA%20Review%20Memo%20--%20BREYANZI.pdf) — FDA Biologics, p. 70
- [454] [BREYANZI (lisocabtagene maraleucel) · Juno Therapeutics, Inc., a Bristol-Myers Squibb company (BLA125714)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/146242/download%3Fattachment.pdf) — FDA Biologics, p. 6
- [462] [SKYSONA (elivaldogene autotemcel) · bluebird bio, Inc.](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/162098/download%3Fattachment.pdf) — FDA Biologics, p. 14
- [470] [TECELRA, afamitresgene autoleucel · Adaptimmune LLC (BLA125789)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/180981/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [471] [TECELRA, afamitresgene autoleucel · Adaptimmune LLC (BLA125789)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/180981/download%3Fattachment.pdf) — FDA Biologics, p. 11
- [473] [TECELRA, afamitresgene autoleucel · Adaptimmune LLC (BLA125789)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/180981/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [475] [TECELRA, afamitresgene autoleucel · Adaptimmune LLC (BLA125789)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/180981/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [483] [RYONCIL, remestemcel-L-rknd · Mesoblast, Inc. (BLA125706)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/185192/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [484] [RYONCIL, remestemcel-L-rknd · Mesoblast, Inc. (BLA125706)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/185192/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [485] [RYONCIL, remestemcel-L-rknd · Mesoblast, Inc. (BLA125706)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/185192/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [486] [RYONCIL, remestemcel-L-rknd · Mesoblast, Inc. (BLA125706)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/185192/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [487] [RYONCIL, remestemcel-L-rknd · Mesoblast, Inc. (BLA125706)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/185192/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [488] [RYONCIL, remestemcel-L-rknd · Mesoblast, Inc. (BLA125706)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/185192/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [489] [RYONCIL, remestemcel-L-rknd · Mesoblast, Inc. (BLA125706)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/185192/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [506] [STRATAGRAFT, allogeneic cultured keratinocytes and dermal fibroblasts in murine collagen-dsat · Stratatech Corporation (BLA125730)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/150741/download%3Fattachment.zip/CMC%20Review%20-%20STRATAGRAFT.pdf) — FDA Biologics, p. 60
- [507] [STRATAGRAFT, allogeneic cultured keratinocytes and dermal fibroblasts in murine collagen-dsat · Stratatech Corporation (BLA125730)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/150741/download%3Fattachment.zip/CMC%20Review%20-%20STRATAGRAFT.pdf) — FDA Biologics, p. 59
- [535] [HEMACORD (HPC, cord blood), Hemacord · New York Blood Center, Inc (BLA125397)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107330/download%3Fattachment.zip/CMC%20Review%20Final%20NYBC%20-%20Hemacord.pdf) — FDA Biologics, p. 38
- [536] [HEMACORD (HPC, cord blood), Hemacord · New York Blood Center, Inc (BLA125397)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107330/download%3Fattachment.zip/CMC%20Review%20Final%20NYBC%20-%20Hemacord.pdf) — FDA Biologics, p. 38
- [537] [HEMACORD (HPC, cord blood), Hemacord · New York Blood Center, Inc (BLA125397)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107330/download%3Fattachment.zip/CMC%20Review%20Final%20NYBC%20-%20Hemacord.pdf) — FDA Biologics, p. 38
- [539] [HEMACORD (HPC, cord blood), Hemacord · New York Blood Center, Inc (BLA125397)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107330/download%3Fattachment.zip/CMC%20Review%20Final%20NYBC%20-%20Hemacord.pdf) — FDA Biologics, p. 38
- [541] [HPC, Cord Blood · Bloodworks (125585)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/downloads/BiologicsBloodVaccines/CellularGeneTherapyProducts/ApprovedProducts/UCM486401.pdf) — FDA Biologics, p. 7
- [542] [HPC, Cord Blood · Bloodworks (BLA125585)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/95521/download%3Fattachment.pdf) — FDA Biologics, p. 10
- [543] [HPC, Cord Blood · Bloodworks (125585)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/downloads/BiologicsBloodVaccines/CellularGeneTherapyProducts/ApprovedProducts/UCM486401.pdf) — FDA Biologics, p. 7
- [546] [HPC, Cord Blood, HPC, Cord Blood - MD Anderson Cord Blood Bank · MD Anderson Cord Blood Bank (BLA125657)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/114643/download%3Fattachment.pdf) — FDA Biologics, p. 6
- [553] [Considerations for the Development of Chimeric Antigen Receptor (CAR) T Cell Products](https://www.fda.gov/media/156896/download) — FDA Guidance, p. 17
- [561] [Considerations for the Development of Chimeric Antigen Receptor (CAR) T Cell Products](https://www.fda.gov/media/156896/download) — FDA Guidance, p. 17
- [565] [Considerations for the Development of Chimeric Antigen Receptor (CAR) T Cell Products](https://www.fda.gov/media/156896/download) — FDA Guidance, p. 17
- [567] [Considerations for the Development of Chimeric Antigen Receptor (CAR) T Cell Products](https://www.fda.gov/media/156896/download) — FDA Guidance, p. 17
- [575] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/Late-Cycle%20Meeting%20Materials,%20June%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 3
- [583] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646, STN125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107962/download%3Fattachment.pdf) — FDA Biologics, p. 7
- [584] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 11
- [587] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/Late-Cycle%20Meeting%20Materials,%20June%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 3
- [588] [KYMRIAH (tisagenlecleucel) · Novartis Pharmaceuticals Corporation (BLA125646)](https://rhizomeai-docs.nyc3.digitaloceanspaces.com/media/107978/download%3Fattachment.zip/CMC%20Review,%20August%2029,%202017%20-%20KYMRIAH.pdf) — FDA Biologics, p. 25
- [592] [Chemistry, Manufacturing, and Controls Flexibilities for Developing Human Cellular and Gene Therapy Products for a Biol…](https://www.fda.gov/media/192321/download) — FDA Guidance, p. 6
- [612] [Guidance for Human Somatic Cell Therapy and Gene Therapy: Guidance for Industry](https://www.fda.gov/media/72402/download) — FDA Guidance, p. 15
- [613] [Content and Review of Chemistry, Manufacturing, and Control (CMC) Information for Human Somatic Cell Therapy Investigat…](https://www.fda.gov/media/73624/download) — FDA Guidance, p. 23
- [614] [Content and Review of Chemistry, Manufacturing, and Control (CMC) Information for Human Somatic Cell Therapy Investigat…](https://www.fda.gov/media/73624/download) — FDA Guidance, p. 23
- [615] [Chemistry, Manufacturing, and Control (CMC) Information for Human Gene Therapy Investigational New Drug Applications (I…](https://www.fda.gov/media/113760/download) — FDA Guidance, p. 51
- [616] [Potency Tests for Cellular and Gene Therapy Products: Final Guidance for Industry:](https://www.fda.gov/media/79856/download) — FDA Guidance, p. 3

---

Rhizome AI researched and wrote this article from health-authority documents. Ask it your own regulatory question at https://rhizomeai.com. More articles: https://rhizomeai.com/articles.md. Guide for agents: https://rhizomeai.com/llms.txt